MOQUI: an open-source GPU-based Monte Carlo code for proton dose calculation with efficient data structure

被引:17
|
作者
Lee, Hoyeon [1 ,2 ]
Shin, Jungwook [3 ]
Verburg, Joost M. [1 ,2 ]
Bobic, Mislav [1 ,2 ,4 ]
Winey, Brian [1 ,2 ]
Schuemann, Jan [1 ,2 ]
Paganetti, Harald [1 ,2 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiat Oncol, Boston, MA 02114 USA
[2] Harvard Med Sch, Boston, MA 02114 USA
[3] NCI, Div Canc Epidemiol & Genet, NIH, Rockville, MD 20850 USA
[4] ETH, Dept Phys, CH-8092 Zurich, Switzerland
关键词
Monte Carlo; graphic processing unit; proton therapy; TREATMENT PLANNING SYSTEM; THERAPY; SIMULATION; IMPLEMENTATION; OPTIMIZATION; VALIDATION; ALGORITHM; TISSUE; WATER;
D O I
10.1088/1361-6560/ac8716
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Monte Carlo (MC) codes are increasingly used for accurate radiotherapy dose calculation. In proton therapy, the accuracy of the dose calculation algorithm is expected to have a more significant impact than in photon therapy due to the depth-dose characteristics of proton beams. However, MC simulations come at a considerable computational cost to achieve statistically sufficient accuracy. There have been efforts to improve computational efficiency while maintaining sufficient accuracy. Among those, parallelizing particle transportation using graphic processing units (GPU) achieved significant improvements. Contrary to the central processing unit, a GPU has limited memory capacity and is not expandable. It is therefore challenging to score quantities with large dimensions requiring extensive memory. The objective of this study is to develop an open-source GPU-based MC package capable of scoring those quantities. Approach. We employed a hash-table, one of the key-value pair data structures, to efficiently utilize the limited memory of the GPU and score the quantities requiring a large amount of memory. With the hash table, only voxels interacting with particles will occupy memory, and we can search the data efficiently to determine their address. The hash-table was integrated with a novel GPU-based MC code, moqui. Main results. The developed code was validated against an MC code widely used in proton therapy, TOPAS, with homogeneous and heterogeneous phantoms. We also compared the dose calculation results of clinical treatment plans. The developed code agreed with TOPAS within 2%, except for the fall-off and regions, and the gamma pass rates of the results were >99% for all cases with a 2 mm/2% criteria. Significance. We can score dose-influence matrix and dose-rate on a GPU for a 3-field H&N case with 10 GB of memory using moqui, which would require more than 100 GB of memory with the conventionally used array data structure.
引用
收藏
页数:14
相关论文
共 50 条
  • [31] Evaluation of proton therapy dose in breast cancer with boron agent based on GATE Monte Carlo code
    Bijarpasi, Fereshteh Mollaee
    Sadremomtaz, Alireza
    JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2025, 18 (01)
  • [32] A photon source model based on particle transport in a parameterized accelerator structure for Monte Carlo dose calculations
    Ishizawa, Yoshiki
    Dobashi, Suguru
    Kadoya, Noriyuki
    Ito, Kengo
    Chiba, Takahito
    Takayama, Yoshiki
    Sato, Kiyokazu
    Takeda, Ken
    MEDICAL PHYSICS, 2018, 45 (07) : 2937 - 2946
  • [33] Fast GPU-based Monte Carlo code for SPECT/CT reconstructions generates improved 177Lu images
    T. Rydén
    J. Heydorn Lagerlöf
    J. Hemmingsson
    I. Marin
    J. Svensson
    M. Båth
    P. Gjertsson
    P. Bernhardt
    EJNMMI Physics, 5
  • [34] Calculation of kinetic parametersβ eff and L with modified open source Monte Carlo code OpenMC(TD)
    Romero-Barrientos, J.
    Damian, J. I. Marquez
    Molina, F.
    Zambra, M.
    Aguilera, P.
    Lopez-Usquiano, F.
    Parra, B.
    Ruiz, A.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2022, 54 (03) : 811 - 816
  • [35] A Monte Carlo based source model for dose calculation of endovaginal TARGIT brachytherapy with INTRABEAM and a cylindrical applicator
    Clausen, Sven
    Schneider, Frank
    Jahnke, Lennart
    Fleckenstein, Jens
    Hesser, Juergen
    Glatting, Gerhard
    Wenz, Frederik
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2012, 22 (03): : 197 - 204
  • [36] Technical Note: Integrating an open source Monte Carlo code "MCsquare" for clinical use in intensity-modulated proton therapy
    Deng, Wei
    Younkin, James E.
    Souris, Kevin
    Huang, Sheng
    Augustine, Kurt
    Fatyga, Mirek
    Ding, Xiaoning
    Cohilis, Marie
    Bues, Martin
    Shan, Jie
    Stoker, Joshua
    Lin, Liyong
    Shen, Jiajian
    Liu, Wei
    MEDICAL PHYSICS, 2020, 47 (06) : 2558 - 2574
  • [37] Configurable Open-source Data Structure for Distributed Conforming Unstructured Homogeneous Meshes with GPU Support
    Klinkovsky, Jakub
    Oberhuber, Tomas
    Fucik, Radek
    Zabka, Vitezslav
    ACM TRANSACTIONS ON MATHEMATICAL SOFTWARE, 2022, 48 (03):
  • [38] A general-purpose Monte Carlo particle transport code based on inverse transform sampling for radiotherapy dose calculation
    Liang, Ying
    Muhammad, Wazir
    Hart, Gregory R.
    Nartowt, Bradley J.
    Chen, Zhe J.
    Yu, James B.
    Roberts, Kenneth B.
    Duncan, James S.
    Deng, Jun
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [39] Development and clinical application of a GPU-based Monte Carlo dose verification module and software for 1.5 T MR-LINAC
    Cheng, Bo
    Xu, Yuan
    Li, Shijun
    Ren, Qiang
    Pei, Xi
    Men, Kuo
    Dai, Jianrong
    Xu, Xie George
    MEDICAL PHYSICS, 2023, 50 (05) : 3172 - 3183
  • [40] Validating a double Gaussian source model for small proton fields in a commercial Monte-Carlo dose calculation engine
    Kugel, Fabian
    Wulff, Jorg
    Baeumer, Christian
    Janson, Martin
    Kretschmer, Jana
    Brodbek, Leonie
    Behrends, Carina
    Verbeek, Nico
    Looe, Hui Khee
    Poppe, Bjorn
    Timmermann, Beate
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 529 - 541