A review of current local dose-area product levels for paediatric fluoroscopy in a tertiary referral centre compared with national standards. Why are they so different?

被引:28
作者
Hiorns, MP [1 ]
Saini, A [1 ]
Marsden, PJ [1 ]
机构
[1] Great Ormond St Hosp Sick Children, London WC1N 3JH, England
关键词
D O I
10.1259/bjr/36530782
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A prospective single centre study has been performed to assess dose-area product (DAP) values in children having fluoroscopic examinations and to revise local diagnostic reference levels (DRLs). DAP measurements for 2658 examinations performed in a dedicated fluoroscopy room over a period of 21 months were analysed. Data for the eight most commonly performed examinations (2215 cases) are presented. DAPs (75th centile) for upper gastrointestinal studies and micturating cystograms are substantially lower (by a factor of between x 5 and x 25) than the current national reference doses (NRDs), with some of the median values being 50 times lower. The small DAP values in all examinations demonstrate the substantial reduction in dose and consequent risk that can be achieved when both equipment performance and operator technique are optimized. Whilst we recognize that different institutions will have differing practices, it is important that practitioners are aware of the range of DAPs achievable and that NRDs do not necessarily represent best practice, and may falsely reassure.
引用
收藏
页码:326 / 330
页数:5
相关论文
共 15 条
[1]  
[Anonymous], 1997, OFFICIAL J EUROPEAN
[2]  
[Anonymous], 88 IPEM
[3]   RESULTS OF A SURVEY OF DOSES TO PEDIATRIC-PATIENTS UNDERGOING COMMON RADIOLOGICAL EXAMINATIONS [J].
CHAPPLE, CL ;
FAULKNER, K ;
LEE, REJ ;
HUNTER, EW .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (771) :225-231
[4]   RADIATION-DOSES TO PEDIATRIC-PATIENTS UNDERGOING LESS COMMON RADIOLOGICAL PROCEDURES INVOLVING FLUOROSCOPY [J].
CHAPPLE, CL ;
FAULKNER, K ;
LEE, REJ ;
HUNTER, EW .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (789) :823-827
[5]   Key factors in the optimization of paediatric X-ray practice [J].
Cook, JV ;
Kyriou, JC ;
Pettet, A ;
Fitzgerald, MC ;
Shah, K ;
Pablot, SM .
BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (887) :1032-1040
[6]  
Council of the European Communities, 1996, 16261 EUR COUNC EUR
[7]  
Dosimetry Working Party of the Institute of physical sciences in medicine, 1992, NAT PROT PAT DOS MEA
[8]   The establishment of reference doses in paediatric radiology as a function of patient size [J].
Hart, D ;
Wall, BF ;
Shrimpton, PC ;
Dance, DR .
RADIATION PROTECTION DOSIMETRY, 2000, 90 (1-2) :235-238
[9]  
HART D, 2000, NRPBR318
[10]   A STUDY OF THE RELATIONSHIP BETWEEN PATIENT DOSE AND SIZE IN PEDIATRIC RADIOLOGY [J].
MARTIN, CJ ;
FARQUHAR, B ;
STOCKDALE, E ;
MACDONALD, S .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (801) :864-871