PDMS Microcantilever-Based Flow Sensor Integration for Lab-on-a-Chip

被引:60
作者
Nezhad, Amir Sanati [1 ]
Ghanbari, Mahmood [1 ]
Agudelo, Carlos G. [1 ]
Packirisamy, Muthukumaran [1 ]
Bhat, Rama B. [2 ]
Geitmann, Anja [3 ]
机构
[1] Concordia Univ, Dept Mech Engn, Opt Biomicrosyst Lab, Montreal, PQ H3G 1M8, Canada
[2] Concordia Univ, Dept Mech Engn, Montreal, PQ H3G 1M8, Canada
[3] Univ Montreal, Inst Rech Biol Vegetale, Dept Sci Biol, Montreal, PQ H3T 1J4, Canada
关键词
Flow sensor; lab-on-a-chip; multilayer soft lithography; polydimethylsiloxane (PDMS) microcantilever; FORCE; POLYDIMETHYLSILOXANE; FABRICATION; SIMULATION;
D O I
10.1109/JSEN.2012.2223667
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In this paper, a simple practical method is presented to fabricate a high aspect ratio horizontal polydimethylsiloxane (PDMS) microcantilever-based flow sensor integrated into a microfluidic device. A multilayer soft lithography process is developed to fabricate a thin PDMS layer involving the PDMS microcantilever and the microfluidics network. A three-layer fabrication technique is explored for the integration of the microflow meter. The upper and lower PDMS layers are bonded to the thin layer to release the microcantilever for free deflection. A 3-D finite element analysis is carried out to simulate fluid-structure interaction and estimate cantilever deflection under various flow conditions. The dynamic range of flow rates that is detectable using the flow sensor is assessed by both simulation and experimental methods and compared. Limited by the accuracy of the 1.76-mu m resolution of the image acquisition method, the present setup allows for flow rates as low as 35 mu L/min to be detected. This is equal to 0.8-mu N resolution in equivalent force at the tip. This flow meter can be integrated into any type of microfluidic-based lab-on-a-chip in which flow measurement is crucial, such as flow cytometry and particle separation applications.
引用
收藏
页码:601 / 609
页数:9
相关论文
共 50 条
  • [31] Quantitative analysis of respiratory viruses based on lab-on-a-chip platform
    Zhang, Ning
    Li, Chao
    Zhan, Xiaobo
    Cheng, Zhi
    Li, Chao
    Du, Yaohua
    Tian, Feng
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2023, 415 (26) : 6561 - 6571
  • [32] Droplet-based microfluidic lab-on-a-chip for glucose detection
    Srinivasan, V
    Pamula, VK
    Fair, RB
    ANALYTICA CHIMICA ACTA, 2004, 507 (01) : 145 - 150
  • [33] Quantitative analysis of respiratory viruses based on lab-on-a-chip platform
    Ning Zhang
    Chao Yue
    Xiaobo Zhan
    Zhi Cheng
    Chao Li
    Yaohua Du
    Feng Tian
    Analytical and Bioanalytical Chemistry, 2023, 415 : 6561 - 6571
  • [34] Lab-on-a-Chip With β-Poly(Vinylidene Fluoride) Based Acoustic Microagitation
    Cardoso, V. F.
    Catarino, S. O.
    Nunes, J. Serrado
    Rebouta, L.
    Rocha, J. G.
    Lanceros-Mendez, S.
    Minas, G.
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (05) : 1184 - 1190
  • [35] Microfluidic lab-on-a-chip systems based on polymers - fabrication and application
    Guber, AE
    Heckele, M
    Herrmann, D
    Muslija, A
    Saile, V
    Eichhorn, L
    Gietzelt, T
    Hoffmann, W
    Hauser, PC
    Tanyanyiwa, J
    Gerlach, A
    Gottschlich, N
    Knebel, G
    CHEMICAL ENGINEERING JOURNAL, 2004, 101 (1-3) : 447 - 453
  • [36] Investigation of High-Speed Erythrocyte Flow and Erythrocyte-Wall Impact in a Lab-on-a-Chip
    Li, Ping
    Zheng, Lu
    Zhang, Di
    Xie, Yonghui
    Feng, Yi
    Xie, Gongnan
    ARTIFICIAL ORGANS, 2016, 40 (11) : E203 - E218
  • [37] Silica-on-silicon (SOS)-PDMS platform integrated Lab-on-a-chip (LOC) for Quantum Dot applications
    Ozhikandathil, Jayan
    Packirisamy, Muthukumaran
    PHOTONICS NORTH 2010, 2010, 7750
  • [38] TriPleX-Based Integrated Optical Ring Resonators for Lab-on-a-Chip and Environmental Detection
    Heideman, Rene
    Hoekman, Marcel
    Schreuder, Erik
    IEEE JOURNAL OF SELECTED TOPICS IN QUANTUM ELECTRONICS, 2012, 18 (05) : 1583 - 1596
  • [39] Toward integrated detection and graphene-based removal of contaminants in a lab-on-a-chip platform
    Andrzej Chałupniak
    Arben Merkoçi
    Nano Research, 2017, 10 : 2296 - 2310
  • [40] A microvalve for lab-on-a-chip applications based on electrochemically actuated SU8 cantilevers
    Ezkerra, Aitor
    Jose Fernandez, Luis
    Mayora, Kepa
    Miguel Ruano-Lopez, Jesus
    SENSORS AND ACTUATORS B-CHEMICAL, 2011, 155 (02): : 505 - 511