Wear at the titanium-titanium and the titanium-zirconia implant-abutment interface: A comparative in vitro study

被引:192
作者
Stimmelmayr, Michael [1 ]
Edelhoff, Daniel [1 ]
Gueth, Jan-Frederik [1 ]
Erdelt, Kurt [1 ]
Happe, Arndt [2 ]
Beuer, Florian [1 ]
机构
[1] Univ Munich, Dept Prosthodont, D-80336 Munich, Germany
[2] Univ Cologne, Dept Oral & Maxillofacial Plast Surg & Implantol, D-50937 Cologne, Germany
关键词
Cyclic loading; Fatigue testing; Implant abutment; Zirconia; Titanium; SOFT-TISSUE; RESTORATIONS;
D O I
10.1016/j.dental.2012.08.008
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The purpose of this study was to determine and measure the wear of the interface between titanium implants and one-piece zirconia abutments in comparison to titanium abutments. Methods. 6 implants were secured into epoxy resin blocks. The implant interface of these implants and 6 corresponding abutments (group Zr: three one-piece zirconia abutments; group Ti: three titanium abutments) were examined by a microscope and scanning electron micrograph (SEM). Also the implants and the abutments were scanned by 3D-Micro Computer Tomography (CT). The abutments were connected to the implants and cyclically loaded with 1,200,000 cycles at 100 N in a two-axis fatigue testing machine. Afterwards, all specimens were unscrewed and the implants and abutments again were scanned by microscope, SEM and CT. The microscope and SEM images were compared, the CT data were superimposed and the wear was calculated by inspection software. The statistical analysis was carried out with an unpaired t-test. Results. Abutment fracture or screw loosening was not observed during cyclical loading. Comparing the microscope and SEM images more wear was observed on the implants connected to zirconia abutments. The maximum wear on the implant shoulder calculated by the inspection software was 10.2 mu m for group Zr, and 0.7 mu m for group Ti. The influence of the abutment material on the measured wear was statistically significant (p <= 0.001; Levene-test). Significance. Titanium implants showed higher wear at the implant interface following cyclic loading when connected to one-piece zirconia implant abutments compared to titanium abutments. The clinical relevance is not clear; hence damage of the internal implant connection could result in prosthetic failures up to the need of implant removal. (C) 2012 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1215 / 1220
页数:6
相关论文
共 25 条
[1]  
ADELL R, 1990, International Journal of Oral and Maxillofacial Implants, V5, P347
[2]  
Al-Sabbagh M., 2006, DENT CLIN N AM, V50, pvi
[3]  
Al-Sabbagh Mohanad, 2006, Dent Clin North Am, V50, P391, DOI 10.1016/j.cden.2006.03.007
[4]  
Andersson B, 2003, INT J PROSTHODONT, V16, P640
[5]  
[Anonymous], COMPEND CONTIN EDUC
[6]   Fracture resistance of single-tooth implant-supported all-ceramic restorations: An in vitro study [J].
Att, W ;
Kurun, S ;
Gerds, T ;
Strub, JR .
JOURNAL OF PROSTHETIC DENTISTRY, 2006, 95 (02) :111-116
[7]   COMPARATIVE-STUDY OF ANTIROTATIONAL SINGLE TOOTH ABUTMENTS [J].
BALFOUR, A ;
OBRIEN, GR .
JOURNAL OF PROSTHETIC DENTISTRY, 1995, 73 (01) :36-43
[8]  
BRODBECK U, 2003, J ESTHET RESTOR DENT, V15, P24
[9]  
Brodbeck Urs, 2003, J Esthet Restor Dent, V15, P10, DOI 10.1111/j.1708-8240.2003.tb00278.x
[10]   The fit of cast and premachined implant abutments [J].
Byrne, D ;
Houston, F ;
Cleary, R ;
Claffey, N .
JOURNAL OF PROSTHETIC DENTISTRY, 1998, 80 (02) :184-192