Performance evaluation of a CyberKnife® G4 image-guided robotic stereotactic radiosurgery system

被引:101
作者
Antypas, Christos [1 ,2 ]
Pantelis, Evaggelos [1 ]
机构
[1] Clin & Diagnost Ctr, Dept Med Phys, Athens 15231, Greece
[2] Univ Athens, Aretaieion Hosp, Dept Radiol, Athens 11528, Greece
关键词
D O I
10.1088/0031-9155/53/17/016
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The aim of the current work was to present the performance evaluation procedures implemented at our department for the commissioning of a G4 CyberKnife system. This system consists of a robotic manipulator, a target-locating system and a lightweight 6-MV linac. Individual quality assurance procedures were performed for each of the CyberKnife subsystems. The system was checked for the mechanical accuracy of its robotic manipulator. The performance of the target-locating system was evaluated in terms of mechanical accuracy of both cameras' alignment and quality assurance tests of the x-ray generators and the flat-panel detectors. The traditional linac 6-MV beam characteristics and beam output parameters were also measured. Results revealed a manipulator mechanical mean accuracy of similar to 0.1 mm, with individual maximum position uncertainties less than 0.25 mm. The target-locating system mechanical accuracy was found within the acceptance limits. For the most clinically used parameters in the CyberKnife practice, e. g. 100-120 kV and 50-200 ms, kV and exposure time accuracy error were measured as less than 2%, while the precision error of the kV was determined as less than 1%. The acquired images of the ETR grid pattern revealed no geometrical distortion while the critical frequency f(50) values for cameras A and B were calculated as 1.5 lp mm(-1) and 1.4 lp mm(-1), respectively. Dose placement measurements were performed in a head and neck phantom. Results revealed sub-millimeter beam delivery precision whereas the total clinical accuracy of the system was measured equal to 0.44 +/- 0.12 mm, 0.29 +/- 0.10 mm and 0.53 +/- 0.16 mm for the skull, fiducial and Xsight spine tracking methods, respectively. The results of this work certify the G4 CyberKnife SRS system capable of delivering high dose distributions with sub-millimeter accuracy and precision to intracranial and extracranial lesions. Moreover, total clinical accuracy of the investigated G4 system was found to be improved for the skull and fiducial tracking methods
引用
收藏
页码:4697 / 4718
页数:22
相关论文
共 45 条
  • [11] Duggan D M, 1998, Med Dosim, V23, P153, DOI 10.1016/S0958-3947(98)00013-2
  • [12] Use of a new type of radiochromic film, a new parallel-plate micro-chamber, MOSFETs, and TLD 800 microcubes in the dosimetry of small beams
    Francescon, P
    Cora, S
    Cavedon, C
    Scalchi, P
    Reccanello, S
    Colombo, F
    [J]. MEDICAL PHYSICS, 1998, 25 (04) : 503 - 511
  • [13] FRANCESCON P, 2005, CYBERKNIFE DOSIMETRI, V1, P271
  • [14] FU D, 2005, SPIE MED IMAGING IMA, P366
  • [15] A fast, accurate, and automatic 2D-3D image registration for image-guided cranial radiosurgery
    Fu, Dongshan
    Kuduvalli, Gopinath
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 2180 - 2194
  • [16] Image-guided radiotherapy using megavoltage cone-beam computed tomography for treatment of paraspinous tumors in the presence of orthopedic hardware
    Hansen, Eric K.
    Larson, David A.
    Aubin, Michele
    Chen, Josephine
    Descovich, Martina
    Gillis, Amy M.
    Morin, Olivier
    Xia, Ping
    Pouliot, Jean
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (02): : 323 - 326
  • [17] A comparison of dosimetry techniques in stereotactic radiosurgery
    Heydarian, M
    Hoban, PW
    Beddoe, AH
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) : 93 - 110
  • [18] A study of the accuracy of CyberKnife spinal radiosurgery using skeletal structure tracking
    Ho, Anthony K.
    Fu, Dongshan
    Cotrutz, Cristian
    Hancock, Steven L.
    Chang, Steven D.
    Gibbs, Iris C.
    Maurer, Calvin R., Jr.
    Adler, John R., Jr.
    [J]. NEUROSURGERY, 2007, 60 (02) : 147 - 156
  • [19] *IPSM, 1988, 53 IPSM
  • [20] Dose verification of single shot gamma knife applications using VIPAR polymer gel and MRI
    Karaiskos, P
    Petrokokkinos, L
    Tatsis, E
    Angelopoulos, A
    Baras, P
    Kozicki, M
    Papagiannis, P
    Rosiak, JM
    Sakelliou, L
    Sandilos, P
    Vlachos, L
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (06) : 1235 - 1250