Standoff Raman spectrometry for the non-invasive detection of explosives precursors in highly fluorescing packaging

被引:15
作者
Izake, Emad L. [1 ]
Sundarajoo, Shankaran [1 ]
Olds, William [1 ]
Cletus, Biju [1 ]
Jaatinen, Esa [2 ]
Fredericks, Peter M. [1 ]
机构
[1] Queensland Univ Technol, Chem Discipline, Fac Sci & Technol, Brisbane, Qld 4001, Australia
[2] Queensland Univ Technol, Phys Discipline, Fac Sci & Technol, Brisbane, Qld 4001, Australia
关键词
Deep Raman spectroscopy; Highly fluorescing packaging; Standoff detection; Explosives precursors; Homeland security; CONCEALED CONTENT; SPECTROSCOPY; DEPTH; SYSTEM; IDENTIFICATION; REJECTION;
D O I
10.1016/j.talanta.2012.09.055
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Noninvasive standoff deep Raman spectroscopy has been utilised for the detection of explosives precursors in highly fluorescing packaging from 15 m. To our knowledge this is the first time standoff deep Raman spectroscopy of concealed substances in highly fluorescing coloured packaging is demonstrated. Time-resolved Raman spectroscopy, spatially offset Raman spectroscopy and time-resolved spatially offset Raman spectroscopy have been compared to identify their selectivity towards the deep layers of a sample. The selectivity of time-resolved Raman spectroscopy towards the concealed chemical substances was found to be comparable to that of spatially offset Raman spectroscopy. However, time-resolved Raman spectroscopy did not require precise translation of the laser excitation beam onto the surface of the interrogated packaging as in the case of spatially offset Raman spectroscopy. Our results confirm that standoff time-resolved spatially offset Raman spectroscopy has significantly higher selectivity towards the deep layers of a sample when compared to the other deep Raman spectroscopy modes. The developed spectrometer was capable of detecting the concealed substances within 5 s of data acquisition. By using time-resolved spatially Raman spectroscopy, a Raman spectrum that is representative of the content alone was acquired without the use of sophisticated algorithms to eliminate the spectral contributions of the packaging material within the acquired spectrum as in the case of time-resolved Raman spectroscopy and spatially offset Raman spectroscopy. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:20 / 27
页数:8
相关论文
共 47 条
[1]  
Akeson M., 2011, P SOC PHOTO-OPT INS, V8017, P1
[2]   Picosecond Raman spectroscopy with a fast intensified CCD camera for depth analysis of diffusely scattering media [J].
Ariese, Freek ;
Meuzelaar, Heleen ;
Kerssens, Marleen M. ;
Buijs, Joost B. ;
Gooijer, Cees .
ANALYST, 2009, 134 (06) :1192-1197
[3]   Time-resolved Raman spectroscopy for in situ planetary mineralogy [J].
Blacksberg, Jordana ;
Rossman, George R. ;
Gleckler, Anthony .
APPLIED OPTICS, 2010, 49 (26) :4951-4962
[4]   MODEL FOR PHOTON MIGRATION IN TURBID BIOLOGICAL MEDIA [J].
BONNER, RF ;
NOSSAL, R ;
HAVLIN, S ;
WEISS, GH .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1987, 4 (03) :423-432
[5]   Non-invasive analysis of turbid samples using deep Raman spectroscopy [J].
Buckley, Kevin ;
Matousek, Pavel .
ANALYST, 2011, 136 (15) :3039-3050
[6]   Standoff detection of high explosive materials at 50 meters in ambient light conditions using a small Raman instrument [J].
Carter, JC ;
Angel, SM ;
Lawrence-Snyder, M ;
Scaffidi, J ;
Whipple, RE ;
Reynolds, JG .
APPLIED SPECTROSCOPY, 2005, 59 (06) :769-775
[7]  
Cletus B, 2011, P SOC PHOTO-OPT INS, V8032, P1
[8]  
Cletus B., 2012, P SOC PHOTO-OPT INS, V8374
[9]   Combined time- and space-resolved Raman spectrometer for the non-invasive depth profiling of chemical hazards [J].
Cletus, Biju ;
Olds, William ;
Izake, Emad L. ;
Sundarajoo, Shankaran ;
Fredericks, Peter M. ;
Jaatinen, Esa .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2012, 403 (01) :255-263
[10]   Fluorescence rejection in resonance Raman spectroscopy using a picosecond-gated intensified charge-coupled device camera [J].
Efremov, Evtim V. ;
Buijs, Joost B. ;
Gooijer, Cees ;
Ariese, Freek .
APPLIED SPECTROSCOPY, 2007, 61 (06) :571-578