Effect of pulse repetition frequency of high-intensity focused ultrasound on in vitro thrombolysis

被引:36
作者
Yang, Wenjing [1 ]
Zhou, Yufeng [1 ]
机构
[1] Nanyang Technol Univ, Sch Mech & Aerosp Engn, Singapore, Singapore
关键词
Thrombolysis; High-intensity focused ultrasound (HIFU); Pulse repetition frequency (PRF); Clot motion; Passive cavitation detection (PCD); ACUTE ISCHEMIC-STROKE; DEEP-VEIN THROMBOSIS; VENOUS THROMBOEMBOLISM; BUBBLE CAVITATION; HISTOTRIPSY; SONOTHROMBOLYSIS; THROMBECTOMY; MICROTRIPSY; MANAGEMENT; PARAMETER;
D O I
10.1016/j.ultsonch.2016.09.014
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Vascular occlusion by the thrombi is the main reason for ischemic stroke and deep vein thrombosis. High intensity focused ultrasound (HIFU) and histotripsy or microtripsy pulses can effectively dissolve the blood clot with no use of thrombolytic agent and ultrasound contrast agent (microbubbles). In this study, HIFU bursts at the same duty cycle (2%) but varied pulse repetition frequency (PRF) from 1 Hz to 1000 Hz were delivered to in vitro porcine blood clot for 30 s. Thrombolysis efficiency initially increases slightly with the PRF, 86.4 +/- 103%, 89.9 +/- 11.9, and 92.9 +/- 12.8% at the PRF of 1 Hz, 10 Hz, and 100 Hz, respectively, without significant difference (p > 0.05), but then drops dramatically to 37.9 +/- 6.9% at the PRF of 1000 Hz (p < 0.05). The particle size in the supernatant of dissolution is 547.1 +/- 129.5 nm, which suggests the disruption of thrombi into the subcellular level. Thrombi motion during HIFU exposure shows violent motion and significant curling at the low PRF, rotation about its axis with occasional curling at the moderate PRF, and localized vibration at the high PRF due to the generation of acoustic radiation force and streaming. Quantitative analysis of recorded motion shows the axial displacement decreases with the PRF of delivered HIFU bursts, from 3.9 +/- 1.5 mm at 1 Hz to 0.7 +/- 0.4 mm at 1000 Hz. Bubble cavitation during HIFU exposure to the blood clot was also monitored. The increase of PRF led to the increase of inertial cavitation but the decrease of stable cavitation. In summary, the PRF of delivered HIFU bursts at the same output energy has a significant effect on the thrombi motion, bubble cavitation activities, and subsequently thrombolysis efficiencies. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:152 / 160
页数:9
相关论文
共 44 条
[1]   Guidelines for thrombolytic therapy for acute stroke: A supplement to the guidelines for the management of patients with acute ischemic stroke - A statement for healthcare professionals from a special writing group of the stroke council, American Heart Association [J].
Adams, HP ;
Brott, TG ;
Furlan, AJ ;
Gomez, CR ;
Grotta, J ;
Helgason, CM ;
Kwiatkowski, T ;
Lyden, PD ;
Marler, JR ;
Torner, J ;
Feinberg, W ;
Mayberg, M ;
Thies, W .
CIRCULATION, 1996, 94 (05) :1167-1174
[2]   Numerical simulation of acoustic streaming generated by finite-amplitude resonant oscillations in an enclosure [J].
Aktas, MK ;
Farouk, B .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2004, 116 (05) :2822-2831
[3]   SHAKEN AND STIRRED: MECHANISMS OF ULTRASOUND-ENHANCED THROMBOLYSIS [J].
Bader, Kenneth B. ;
Gruber, Matthew J. ;
Holland, Christy K. .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2015, 41 (01) :187-196
[4]   Treatment of deep-vein thrombosis [J].
Bates, SM ;
Ginsberg, JS .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 351 (03) :268-277
[5]   Venous Thromboembolism A Public Health Concern [J].
Beckman, Michele G. ;
Hooper, W. Craig ;
Critchley, Sara E. ;
Ortel, Thomas L. .
AMERICAN JOURNAL OF PREVENTIVE MEDICINE, 2010, 38 (04) :S495-S501
[6]   Sonothrombolysis for acute ischemic stroke: a systematic review of randomized controlled trials [J].
Bor-Seng-Shu, Edson ;
Nogueira, Ricardo De Carvalho ;
Figueiredo, Eberval G. ;
Evaristo, Eli Faria ;
Conforto, Adriana Bastos ;
Teixeira, Manoel Jacobsen .
NEUROSURGICAL FOCUS, 2012, 32 (01) :E5
[7]   High-Intensity Focused Ultrasound (HIFU) for Dissolution of Clots in a Rabbit Model of Embolic Stroke [J].
Burgess, Alison ;
Huang, Yuexi ;
Waspe, Adam C. ;
Ganguly, Milan ;
Goertz, David E. ;
Hynynen, Kullervo .
PLOS ONE, 2012, 7 (08)
[8]   Measurement of the ultrasonic attenuation coefficient of human blood plasma during clotting in the frequency range of 8 to 22 MHZ [J].
Calor-Filho, Marcos Muniz ;
Machado, Joao Carlos .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2006, 32 (07) :1055-1064
[9]   Review of LDA and PIV applied to the measurement of sound and acoustic streaming [J].
Campbell, M ;
Cosgrove, JA ;
Greated, CA ;
Jack, S ;
Rockliff, D .
OPTICS AND LASER TECHNOLOGY, 2000, 32 (7-8) :629-639
[10]   OBSERVATIONS OF THE DYNAMICS AND ACOUSTICS OF TRAVELING BUBBLE CAVITATION [J].
CECCIO, SL ;
BRENNEN, CE .
JOURNAL OF FLUID MECHANICS, 1991, 233 :633-660