Influence and Compensation of Truncation Artifacts in MR-Based Attenuation Correction in PET/MR

被引:29
作者
Schramm, G. [1 ,2 ]
Langner, J. [1 ]
Hofheinz, F. [1 ]
Petr, J. [1 ]
Lougovski, A. [1 ]
Beuthien-Baumann, B. [3 ]
Platzek, I. [4 ]
van den Hoff, J. [1 ,3 ]
机构
[1] Helmholtz Zentrum Dresden Rossendorf, Inst Radiopharmaceut Canc Res, Dept Positron Emiss Tomog, D-01328 Dresden, Germany
[2] Tech Univ Dresden, Fac Med Carl Gustav Carus, D-01187 Dresden, Germany
[3] Univ Hosp Carl Gustav Carus, Dept Nucl Med, D-01307 Dresden, Germany
[4] Univ Hosp Carl Gustav Carus, Dept Radiol, D-01307 Dresden, Germany
关键词
Attenuation correction; positron emission tomography/magnetic resonance imaging (PET/MRI); reduced field-of-view (FOV); truncation compensation; IMAGE ARTIFACTS; SURFACE COILS; CT;
D O I
10.1109/TMI.2013.2272660
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The goal of this article is to quantify the influence of truncation artifacts in the magnetic resonance (MR)-based attenuation map (MRMap) on reconstructed positron emission tomography (PET) image volumes and to propose a new method for minimizing this influence. Methods: PET data sets of 20 patients investigated in a Philips Ingenuity PET/MR were reconstructed with and without applying two different methods for truncation compensation (TC1 vendor-provided, TC2 newly developed). In this patient group, the extent of truncation artifacts and quality of the truncation compensation (TC) was assessed visually in the MRMaps. In three additional patients MRMaps generated by algorithm TC2 could be compared to the ground truth of transmission-based attenuation maps obtained with a Siemens ECAT HR+ scanner. The influence of truncation on regional SUVs in lesions, other hot structures (bladder, kidney, myocardium) and the arms was assessed in suitable volume of interests (VOI). Results: Truncation compensated MRMaps exhibited residual artifacts in the arms in 16 patients for algorithm TC1 and to a lesser extent in eight patients for algorithm TC2. Compared to the transmission-based attenuation maps algorithm TC2 slightly overestimated the size of the truncated arms by 0.3 cm in the radial direction. Without truncation compensation, VOIs located in the trunk showed an average SUVmax underestimation of less than 5.4% relative to the results obtained with TC2. Inside the patients' arms underestimations up to 46.5% were found. Conclusion: In the trunk, standardized uptake values (SUV) underestimations due to truncation artifacts in the MRMap are rather small. Inside the arms, severe SUV under-estimations can occur. Therefore, reliable TC is mandatory and can be achieved by applying the newly developed algorithm TC2 which has yielded promising results so far. Implementation of the proposed method is straightforward and should be easily adaptable to other PET/MR systems.
引用
收藏
页码:2056 / 2063
页数:8
相关论文
共 31 条
[1]  
Beyer T, 2006, J NUCL MED, V47, P91
[2]  
Beyer T, 2004, J NUCL MED, V45, p25S
[3]   Comparison of threshold-based and watershed-based segmentation for the truncation compensation of PET/MR images [J].
Blaffert, Thomas ;
Renisch, Steffen ;
Tang, Jing ;
Narayanan, Manoj ;
Hu, Zhiqiang .
MEDICAL IMAGING 2012: IMAGE PROCESSING, 2012, 8314
[4]  
Blumhagen J. O., 2012, MAGN RESON MED, P1
[5]   PET attenuation coefficients from CT images: experimental evaluation of the transformation of CT into PET 511-keV attenuation coefficients [J].
Burger, C ;
Goerres, G ;
Schoenes, S ;
Buck, A ;
Lonn, AHR ;
von Schulthess, GK .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2002, 29 (07) :922-927
[6]  
Delso G, 2011, J NUCL MED, V52, P1914, DOI 10.2967/jnumed.111.092726
[7]   The effect of limited MR field of view in MR/PET attenuation correction [J].
Delso, Gaspar ;
Martinez-Moeller, Axel ;
Bundschuh, Ralph A. ;
Nekolla, Stephan G. ;
Ziegler, Sibylle I. .
MEDICAL PHYSICS, 2010, 37 (06) :2804-2812
[8]   Automatic volume delineation in oncological PET Evaluation of a dedicated software tool and comparison with manual delineation in clinical data sets [J].
Hofheinz, F. ;
Poetzsch, C. ;
Oehme, L. ;
Beuthien-Baumann, B. ;
Steinbach, J. ;
Kotzerke, J. ;
van den Hoff, J. .
NUKLEARMEDIZIN-NUCLEAR MEDICINE, 2012, 51 (01) :9-16
[9]   Suitability of bilateral filtering for edge-preserving noise reduction in PET [J].
Hofheinz, Frank ;
Langner, Jens ;
Beuthien-Baumann, Bettina ;
Oehme, Liane ;
Steinbach, Joerg ;
Kotzerke, Joerg ;
van den Hoff, Joerg .
EJNMMI RESEARCH, 2011, 1 :1-9
[10]   Towards quantitative PET/MRI: a review of MR-based attenuation correction techniques [J].
Hofmann, Matthias ;
Pichler, Bernd ;
Schoelkopf, Bernhard ;
Beyer, Thomas .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2009, 36 :93-104