Placement of thoracolumbar pedicle screws using O-arm-based navigation: technical note on controlling the operational accuracy of the navigation system

被引:21
作者
Ammirati, Mario [1 ]
Salma, Asem [2 ]
机构
[1] Ohio State Univ, Med Ctr, Dept Neurol Surg, Columbus, OH 43210 USA
[2] Univ Illinois, Coll Med, Dept Neurol Surg, Illinois Neurol Inst, Peoria, IL 61656 USA
关键词
Image guidance; Spine surgery; Pedicle screws; Operational accuracy; O-arm; COMPUTED-TOMOGRAPHY; CLINICAL ARTICLE; IMAGE GUIDANCE; FLUOROSCOPY; INSERTION; FUSION; INSTRUMENTATION; SPINE;
D O I
10.1007/s10143-012-0421-2
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Suboptimal placements of pedicle screws may lead to neurological and vascular complications. Computer-assisted image guidance has been shown to improve accuracy in spinal instrumentation. Checking the accuracy of the navigation system during pedicle screw placement is fundamental. We describe a novel technique of using continuous accuracy check of the navigation system during O-arm-based neuronavigation to instrument the thoracolumbar region. Forty thoracic and 42 lumbar screws were inserted in 12 patients. The Mirza evaluation system was used to evaluate the accuracy of the inserted screws. There was no neurological injury and no need to reposition any screw. The accuracy of the screws placement was excellent. Our technique of continuous at will operational accuracy check of the neuronavigation system is associated with extreme accuracy of screw placement, no need to bring a patient back to the operating room to reposition a pedicle screw, and with excellent outcome.
引用
收藏
页码:157 / 162
页数:6
相关论文
共 19 条
  • [1] Accuracy of pedicle screw placement in the lumbosacral spine using conventional technique: computed tomography postoperative assessment in 102 consecutive patients Clinical article
    Amato, Vincenzo
    Giannachi, Luigi
    Irace, Claudio
    Corona, Claudio
    [J]. JOURNAL OF NEUROSURGERY-SPINE, 2010, 12 (03) : 306 - 313
  • [2] Benefit and accuracy of intraoperative 3D-imaging after pedicle screw placement: a prospective study in stabilizing thoracolumbar fractures
    Beck, Markus
    Mittlmeier, Thomas
    Gierer, Philip
    Harms, Christoph
    Gradl, Georg
    [J]. EUROPEAN SPINE JOURNAL, 2009, 18 (10) : 1469 - 1477
  • [3] A COMPARISON OF FLUOROSCOPY AND COMPUTED TOMOGRAPHY-DERIVED VOLUMETRIC MULTIPLE EXPOSURE TRANSMISSION HOLOGRAPHY FOR THE GUIDANCE OF LUMBAR PEDICLE SCREW INSERTION
    BENZEL, EC
    RUPP, FW
    MCCORMACK, BM
    BALDWIN, NG
    ANSON, JA
    ADAMS, MS
    [J]. NEUROSURGERY, 1995, 37 (04) : 711 - 716
  • [4] Personalised image-based templates for intra-operative guidance
    Berry, E
    Cuppone, M
    Porada, S
    Millner, PA
    Rao, A
    Chiverton, N
    Seedhom, BB
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2005, 219 (H2) : 111 - 118
  • [5] Accuracy of upper thoracic pedicle screw placement using three-dimensional image guidance
    Bledsoe, Jonathan M.
    Fenton, Doug
    Fogelson, Jeremy L.
    Nottmeier, Eric W.
    [J]. SPINE JOURNAL, 2009, 9 (10) : 817 - 821
  • [6] Long-term functional outcome of pedicle screw instrumentation as a support for posterolateral spinal fusion -: Randomized clinical study with a 5-year follow-up
    Christensen, FB
    Hansen, ES
    Laursen, M
    Thomsen, K
    Bünger, CE
    [J]. SPINE, 2002, 27 (12) : 1269 - 1277
  • [7] Computer-aided pedicle screw placement using frameless stereotaxis
    Glossop, ND
    Hu, RW
    Randle, JA
    [J]. SPINE, 1996, 21 (17) : 2026 - 2034
  • [8] Pedicle screw placement in the thoracic spine - A comparison of image-guided and manual techniques in cadavers
    Hart, RA
    Hansen, BL
    Shea, M
    Hsu, F
    Anderson, GJ
    [J]. SPINE, 2005, 30 (12) : E326 - E331
  • [9] Pedicle screw placement accuracy - A meta-analysis
    Kosmopoulos, Victor
    Schizas, Constantin
    [J]. SPINE, 2007, 32 (03) : E111 - E120
  • [10] Accuracy of thoracic vertebral body screw placement using standard fluoroscopy, fluoroscopic image guidance, and computed tomographic image guidance - A cadaver study
    Mirza, SK
    Wiggins, GC
    Kuntz, C
    York, JE
    Bellabarba, C
    Knonodi, MA
    Chapman, JR
    Shaffrey, CI
    [J]. SPINE, 2003, 28 (04) : 402 - 413