Factors Affecting Perceptual Threshold in Argus II Retinal Prosthesis Subjects

被引:101
作者
Ahuja, A. K. [1 ,3 ]
Yeoh, J. [2 ]
Dorn, J. D. [1 ]
Caspi, A. [1 ]
Wuyyuru, V. [1 ]
McMahon, M. J. [1 ]
Humayun, M. S. [3 ]
Greenberg, R. J. [1 ]
daCruz, L. [2 ]
机构
[1] Second Sight Med Prod Inc, Sylmar, CA USA
[2] Moorfields Eye Hosp, London, England
[3] Keck Sch Med, Doheny Eye Inst, Los Angeles, CA USA
关键词
retinal prosthesis; retinal degeneration; retinitis pigmentosa;
D O I
10.1167/tvst.2.4.1
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: The Argus II epiretinal prosthesis has been developed to provide partial restoration of vision to subjects blinded from outer retinal degenerative disease. Participants were surgically implanted with the system in the United States and Europe in a single arm, prospective, multicenter clinical trial. The purpose of this investigation was to determine which factors affect electrical thresholds in order to inform surgical placement of the device. Methods: Electrode-retina and electrode-fovea distances were determined using SD-OCT and fundus photography, respectively. Perceptual threshold to electrical stimulation of electrodes was measured using custom developed software, in which current amplitude was varied until the threshold was found. Full field stimulus light threshold was measured using the Espion D-FST test. Relationships between electrical threshold and these three explanatory variables (electrode-retina distance, electrode-fovea distance, and monocular light threshold) were quantified using regression. Results: Regression analysis showed a significant correlation between electrical threshold and electrode-retina distance (R-2 = 0.50, P = 0.0002; n = 703 electrodes). 90.3% of electrodes in contact with the macula (n = 207) elicited percepts at charge densities less than 1 mC/cm(2)/phase. These threshold data also correlated well with ganglion cell density profile (P = 0.03). A weaker, but still significant, inverse correlation was found between light threshold and electrical threshold (R-2 < 0.52, P = 0.01). Multivariate modeling indicated that electrode-retina distance and light threshold are highly predictive of electrode threshold (R-2 = 0.87; P < 0.0005). Conclusions: Taken together, these results suggest that while light threshold should be used to inform patient selection, macular contact of the array is paramount. Translational Relevance: Reported Argus II clinical study results are in good agreement with prior in vitro and in vivo studies, and support the development of higher-density systems that employ smaller diameter electrodes. (clinicaltrials. gov identifier: NCT00407602)
引用
收藏
页数:15
相关论文
共 52 条
[1]   An in vitro model of a retinal prosthesis [J].
Ahuja, Ashish K. ;
Behrend, Matthew R. ;
Kuroda, Masako ;
Humayun, Mark S. ;
Weiland, James D. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2008, 55 (06) :1744-1753
[2]   Retinal Laminar architecture in human retinitis pigmentosa caused by Rhodopsin gene mutations [J].
Aleman, Tomas S. ;
Cideciyan, Artur V. ;
Sumaroka, Alexander ;
Windsor, Elizabeth A. M. ;
Herrera, Waldo ;
White, D. Alan ;
Kaushal, Shalesh ;
Naidu, Anjani ;
Roman, Alejandro J. ;
Schwartz, Sharon B. ;
Stone, Edwin M. ;
Jacobson, Samuel G. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2008, 49 (04) :1580-1590
[3]   Inner retinal abnormalities in X-linked retinitis pigmentosa with RPGR mutations [J].
Aleman, Tomas S. ;
Cideciyan, Artur V. ;
Sumaroka, Alexander ;
Schwartz, Sharon B. ;
Roman, Alejandro J. ;
Windsor, Elizabeth A. M. ;
Steinberg, Janet D. ;
Branham, Kari ;
Othman, Mohammad ;
Swaroop, Anand ;
Jacobson, Samuel G. .
INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2007, 48 (10) :4759-4765
[4]   Resolution of the Epiretinal Prosthesis is not Limited by Electrode Size [J].
Behrend, Matthew R. ;
Ahuja, Ashish K. ;
Humayun, Mark S. ;
Chow, Robert H. ;
Weiland, James D. .
IEEE TRANSACTIONS ON NEURAL SYSTEMS AND REHABILITATION ENGINEERING, 2011, 19 (04) :436-442
[5]   Selective labeling of retinal ganglion cells with calcium indicators by retrograde loading in vitro [J].
Behrend, Matthew R. ;
Ahuja, Ashish K. ;
Humayun, Mark S. ;
Weiland, James D. ;
Chowe, Robert H. .
JOURNAL OF NEUROSCIENCE METHODS, 2009, 179 (02) :166-172
[6]  
Boiko T, 2003, J NEUROSCI, V23, P2306
[7]   Feasibility Study of a Retinal Prosthesis [J].
Caspi, Avi ;
Dorn, Jessy D. ;
McClure, Kelly H. ;
Humayun, Mark S. ;
Greenberg, Robert J. ;
McMahon, Matthew J. .
ARCHIVES OF OPHTHALMOLOGY, 2009, 127 (04) :398-401
[8]  
Chen Shih-Jen, 2006, Trans Am Ophthalmol Soc, V104, P252
[9]   RETINAL EXCITATION AND INHIBITION FROM DIRECT ELECTRICAL STIMULATION [J].
CRAPPER, DR ;
NOELL, WK .
JOURNAL OF NEUROPHYSIOLOGY, 1963, 26 (06) :924-&
[10]   TOPOGRAPHY OF GANGLION-CELLS IN HUMAN RETINA [J].
CURCIO, CA ;
ALLEN, KA .
JOURNAL OF COMPARATIVE NEUROLOGY, 1990, 300 (01) :5-25