Comparing the embolic potential of open and closed cell stents during carotid angioplasty and stenting

被引:20
|
作者
Tadros, Rami O. [1 ]
Spyris, Constantinos T. [1 ]
Vouyouka, Ageliki G. [1 ]
Chung, Christine [1 ]
Krishnan, Prakash [1 ]
Arnold, Margaret W. [1 ]
Marin, Michael L. [1 ]
Faries, Peter L. [1 ]
机构
[1] Mt Sinai Sch Med, Dept Vasc Surg, New York, NY 10029 USA
关键词
ENDARTERECTOMY; DESIGN; TRIAL; INTERVENTION; PREVENTION; IMPACT; RISK;
D O I
10.1016/j.jvs.2011.12.077
中图分类号
R61 [外科手术学];
学科分类号
摘要
Objective: We sought to determine the effects of open (O) and closed (C) cell stents on the size and number of embolic particles generated during carotid artery stenting (CAS) and assess the impact on outcome. Methods: Embolic debris from carotid filters after CAS was analyzed using photomicroscopy and imaging software. Patient comorbidities, preoperative cerebrovascular symptoms, stent type, and outcomes (perioperative major adverse events) were examined. Results: Carotid filters from 173 consecutive CAS procedures (O, 125 and C, 48) were reviewed. The mean age was 70.9 +/- 9.2 years; 58% were men. Mean stenosis was 88.2% +/- 8.1%; 36.6% had neurological symptoms preprocedurally. There was no difference in preoperative symptoms between the two groups (O, 38.7% vs C, 31.3%; P = not significant [NS]). However, closed cell stent use was associated with higher degree of stenosis (O, 87.2% +/- 8.0% vs C, 90.6% +/- 7.8%; P = .01), an older age (O, 70.0 +/- 8.6 years vs C, 73.4 +/- 10.2 years; P = .03), and peripheral arterial disease (21.1% vs 43.5%; P = .01). A larger mean particle size was observed in patients treated with open cell stents compared to closed cell stents (O, 416.5 +/- 335.7 mu m vs C, 301.1 +/- 251.3 mu m; P = .03). There was no significant difference in the total number of particles (O, 13.8 +/- 21.5 vs C, 17.6 +/- 19.9; P = NS), periprocedural stroke (P = NS), and major adverse events between the two groups (P = NS). Conclusions: Open cell stents are associated with a larger mean particle size compared to closed cell stents. No impact on procedural outcomes based on stent type was observed. (J Vasc Surg 2012;56:89-95.)
引用
收藏
页码:89 / 95
页数:7
相关论文
共 50 条
  • [21] A Microscopic and Biomarker Evaluation of Embolic Filter Debris Collected During Carotid Artery Stenting
    Yang, Mark
    Yu, Yan
    Walsh, William R.
    Yang, Jia-Lin
    Baker, Luke
    Lennox, Andrew F.
    Crowe, Philip J.
    Varcoe, Ramon L.
    JOURNAL OF ENDOVASCULAR THERAPY, 2016, 23 (02) : 275 - 284
  • [22] Outcome of Carotid Angioplasty With a Novel Open-Cell Carotid Stent System
    Langhoff, Ralf
    Mudra, Harald
    Waliszewski, Matthias
    Reimers, Bernhard
    Berge, Jerome
    Setacci, Carlo
    Beyssen, Bernard
    VASCULAR AND ENDOVASCULAR SURGERY, 2014, 48 (04) : 317 - 324
  • [23] Comparing open and closed cell stents in idiopathic intracranial hypertension: A comprehensive meta-analysis of clinical outcomes
    Batista, Savio
    Palavani, Lucca B.
    Verly, Gabriel
    Ferreira, Marcio Yuri
    Sanches, Joao Pedro Bittar
    Silva, Guilherme Melo
    Pinheiro, Agostinho C.
    Filho, Jose Alberto Almeida
    NEURORADIOLOGY JOURNAL, 2025, 38 (01) : 21 - 29
  • [24] Open-cell versus closed-cell stent design differences in blood flow velocities after carotid stenting
    Pierce, Damon S.
    Rosero, Eric B.
    Modrall, J. Gregory
    Adams-Huet, Beverley
    Valentine, R. James
    Clagett, G. Patrick
    Timaran, Carlos H.
    JOURNAL OF VASCULAR SURGERY, 2009, 49 (03) : 602 - 606
  • [25] Casper vs. Closed-Cell Stent Carotid Artery Stenting Randomized Trial
    Vanzin, Jose Ricardo
    de Castro-Afonso, Luis Henrique
    Santos, Moema Nene
    Manzato, Luciano Bambini
    Nakiri, Guilherme Seizem
    Monsignore, Lucas Moretti
    Trivelato, Felipe Padovani
    Rezende, Marco Tulio Salles
    Pontes-Neto, Octavio Marques
    Abud, Daniel Giansante
    CLINICAL NEURORADIOLOGY, 2021, 31 (03) : 843 - 852
  • [26] Optical coherence tomography evaluation of tissue prolapse after carotid artery stenting using closed cell design stents for unstable plaque
    Harada, Kei
    Oshikata, Shogo
    Kajihara, Masahito
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2018, 10 (03) : 229 - 234
  • [27] Double Filtration During Carotid Artery Stenting Using a Novel Post-Dilation Balloon With Integrated Embolic Protection
    Langhoff, Ralf
    Schofer, Joachim
    Scheinert, Dierk
    Schmidt, Andrej
    Sedgewick, Gerald
    Saylors, Elizabeth
    Sachar, Ravish
    Sievert, Horst
    Zeller, Thomas
    JACC-CARDIOVASCULAR INTERVENTIONS, 2019, 12 (04) : 395 - 403
  • [28] Comparison of Flow Impairment during Carotid Artery Stenting Using Two Types of Eccentric Filter Embolic Protection Devices
    Nii, Kouhei
    Tsutsumi, Masanori
    Maeda, Hitoshi
    Aikawa, Hiroshi
    Inoue, Ritsuro
    Eto, Ayumu
    Sakamoto, Kimiya
    Mitsutake, Takafumi
    Hanada, Hayatsura
    Kazekawa, Kiyoshi
    NEUROLOGIA MEDICO-CHIRURGICA, 2016, 56 (12) : 759 - 765
  • [29] Use of an embolic protection device during carotid artery stenting is associated with lower periprocedural risk
    Knappich, Christoph
    Bohmann, Bianca
    Kirchhoff, Felix
    Lohe, Vanessa
    Naher, Shamsun
    Kallmayer, Michael
    Eckstein, Hans-Henning
    Kuehnl, Andreas
    JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2024,
  • [30] Size and nature of emboli produced during carotid artery angioplasty and stenting: In vivo study
    Brightwell, R. E.
    Ryder, T. A.
    Hamady, M.
    Cheshire, N. J. W.
    INTERNATIONAL JOURNAL OF SURGERY, 2011, 9 (02) : 177 - 182