Effects of acquisition time and reconstruction algorithm on image quality, quantitative parameters, and clinical interpretation of myocardial perfusion imaging

被引:15
作者
Enevoldsen, Lotte H. [1 ]
Menashi, Changez A. K. [1 ]
Andersen, Ulrik B. [1 ]
Jensen, Lars T. [1 ]
Henriksen, Otto M. [1 ]
机构
[1] Glostrup Cty Hosp, Sect Clin Physiol & Nucl Med, Dept Diagnost, DK-2600 Glostrup, Denmark
关键词
Myocardial perfusion imaging; single photon emission computed tomography; stress testing; coronary artery disease; Tc-99m MIBI; SUBSET EXPECTATION MAXIMIZATION; RESOLUTION RECOVERY; FILTERED BACKPROJECTION; SPECT; HALF; DISEASE;
D O I
10.1007/s12350-013-9775-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Recently introduced iterative reconstruction algorithms with resolution recovery (RR) and noise-reduction technology seem promising for reducing scan time or radiation dose without loss of image quality. However, the relative effects of reduced acquisition time and reconstruction software have not previously been reported. The aim of the present study was to investigate the influence of reduced acquisition time and reconstruction software on quantitative and qualitative myocardial perfusion single photon emission computed tomography (SPECT) parameters using full time (FT) and half time (HT) protocols and Evolution for Cardiac Software. We studied 45 consecutive, non-selected patients referred for a clinically indicated routine 2-day stress/rest Tc-99m-Sestamibi myocardial perfusion SPECT. All patients underwent an FT and an HT scan. Both FT and HT scans were processed according to our standard procedure with both ordered-subset expectation maximization (OSEM) + filtered back projection (FBP) reconstructions and a second reconstruction of HT scans was performed with the RR software producing three datasets for each patient for visual analysis (FT-OSEM, HT-OSEM, and HT-RR) and for quantitative analysis (FT-FBP, HT-FBP, and HT-RR). The datasets were analyzed using commercially available QGS/QPS software and read by two observers evaluating image quality and clinical interpretation. Image quality was assessed on a 10-cm visual analog scale score. HT imaging was associated with loss of image quality that was compensated for by RR reconstruction. HT imaging was also associated with increasing perfusion defect extents, an effect more pronounced using RR than FBP reconstruction. Compared to standard FT-FBP, HT-RR significantly reduced left ventricular volumes whereas HT-FBP increased end-systolic volume. HT imaging had no effect on measured left ventricular ejections fraction or measures of reversibility. Image interpretation found a higher level of concordance between FT-OSEM and HT-RR than between FT-OSEM and HT-OSEM without any observable systematic effects. Use of RR reconstruction algorithms compensates for loss of image quality associated with reduced scan time. Both HT acquisition and RR reconstruction algorithm had significant effects on motion and perfusion parameters obtained with standard software, but these effects were relatively small and probably of limited clinical importance. Although no systematic effects on image interpretation were observed, the influence on diagnostic accuracy remains to be determined.
引用
收藏
页码:1086 / 1092
页数:7
相关论文
共 21 条
  • [1] Half-Time SPECT Myocardial Perfusion Imaging with Attenuation Correction
    Ali, Iftikhar
    Ruddy, Terrence D.
    Almgrahi, Abdulaziz
    Anstett, Frank G.
    Wells, R. Glenn
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2009, 50 (04) : 554 - 562
  • [2] Reduced-count myocardial perfusion SPECT with resolution recovery
    Armstrong, Ian S.
    Arumugam, Parthiban
    James, Jackie M.
    Tonge, Christine M.
    Lawson, Richard S.
    [J]. NUCLEAR MEDICINE COMMUNICATIONS, 2012, 33 (02) : 121 - 129
  • [3] Belhocine T., 2008, EVOLUTION CARDIAC IN
  • [4] Clinical results of a novel wide beam reconstruction method for shortening scan time of Tc-99m cardiac SPECT perfusion studies
    Borges-Neto, Salvador
    Pagnanelli, Robert A.
    Shaw, Linda K.
    Honeycutt, Emily
    Shwartz, Shuli C.
    Adams, George L.
    Coleman, Ralph Edward
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2007, 14 (04) : 555 - 565
  • [5] Daou D, 2003, J NUCL MED, V44, P155
  • [6] De LA, 2010, NUCL MED COMMUN, V31, P552
  • [7] Ordered subset expectation maximization and wide beam reconstruction "half-time" gated myocardial perfusion SPECT functional imaging: A comparison to "full-time" filtered backprojection
    DePuey, E. Gordon
    Gadiraju, Ramesh
    Clark, John
    Thompson, Linda
    Anstett, Frank
    Shwartz, Shuli C.
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2008, 15 (04) : 547 - 563
  • [8] A comparison of the image quality of full-time myocardial perfusion SPECT vs wide beam reconstruction half-time and half-dose SPECT
    DePuey, E. Gordon
    Bommireddipalli, Srinivas
    Clark, John
    Leykekhman, Anna
    Thompson, Linda B.
    Friedman, Marvin
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2011, 18 (02) : 273 - 280
  • [9] Wide beam reconstruction "quarter-time" gated myocardial perfusion SPECT functional imaging: a comparison to "full-time" ordered subset expectation maximum
    DePuey, E. Gordon
    Bommireddipalli, Srinivas
    Clark, John
    Thompson, Linda
    Srour, Yossi
    [J]. JOURNAL OF NUCLEAR CARDIOLOGY, 2009, 16 (05) : 736 - 752
  • [10] DEPUEY EG, 1995, J NUCL MED, V36, P569