Numerical study of the influence of water evaporation on radiofrequency ablation

被引:27
|
作者
Zhu, Qing [1 ]
Shen, Yuanyuan [2 ]
Zhang, Aili [1 ]
Xu, Lisa X. [1 ]
机构
[1] Shanghai Jiao Tong Univ, MedX Res Inst, Sch Biomed Engn, State Key Lab Oncogenes & Related Genes, Shanghai 200030, Peoples R China
[2] Shenzhen Univ, Sch Med, Dept Biomed Engn, Natl Reg Key Technol Engn Lab Med Ultrasound, Shenzhen 518060, Peoples R China
基金
中国国家自然科学基金;
关键词
Radiofrequency ablation; Water evaporation; Water diffusion; Mathematical model; TEMPERATURE DISTRIBUTION; ELECTRICAL-CONDUCTIVITY; DIELECTRIC-PROPERTIES; THERMAL-CONDUCTIVITY; TISSUE TEMPERATURE; BIOLOGICAL TISSUES; HEATING PATTERNS; SELF-DIFFUSION; RF ABLATION; TUMOR;
D O I
10.1186/1475-925X-12-127
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Radiofrequency ablation is a promising minimal invasive treatment for tumor. However, water loss due to evaporation has been a major issue blocking further RF energy transmission and correspondently eliminating the therapeutic outcome of the treatment. Method: A 2D symmetric cylindrical mathematical model coupling the transport of the electrical current, heat, and the evaporation process in the tissue, has been developed to simulate the treatment process and investigate the influence of the excessive evaporation of the water on the treatment. Results: Our results show that the largest specific absorption rate (QSAR) occurs at the edge of the circular surface of the electrode. When excessive evaporation takes place, the water dehydration rate in this region is the highest, and after a certain time, the dehydrated tissue blocks the electrical energy transmission in the radial direction. It is found that there is an interval as long as 65 s between the beginning of the evaporation and the increase of the tissue impedance. The model is further used to investigate whether purposely terminating the treatment for a while allowing diffusion of the liquid water into the evaporated region would help. Results show it has no obvious improvement enlarging the treatment volume. Treatment with the cooled-tip electrode is also studied. It is found that the cooling conditions of the inside agent greatly affect the water loss pattern. When the convection coefficient of the cooling agent increases, excessive evaporation will start from near the central axis of the tissue cylinder instead of the edge of the electrode, and the coagulation volume obviously enlarges before a sudden increase of the impedance. It is also found that a higher convection coefficient will extend the treatment time. Though the sudden increase of the tissue impedance could be delayed by a larger convection coefficient; the rate of the impedance increase is also more dramatic compared to the case with smaller convection coefficient. Conclusion: The mathematical model simulates the water evaporation and diffusion during radiofrequency ablation and may be used for better clinical design of radiofrequency equipment and treatment protocol planning.
引用
收藏
页数:16
相关论文
共 50 条
  • [41] Cardiac Radiofrequency Ablation
    Shoulders, Bridget
    Mauriello, Jillian
    Shellman, Tamika
    Follett, Corrinne
    DIMENSIONS OF CRITICAL CARE NURSING, 2016, 35 (05) : 255 - 267
  • [42] Influence of Surgical Staples on Radiofrequency Ablation Using Multitined Expandable Electrodes
    Yusuke Sakuhara
    Tadashi Shimizu
    Daisuke Abo
    Yu Hasegawa
    Fumi Kato
    Yoshihisa Kodama
    Hiroki Shirato
    CardioVascular and Interventional Radiology, 2007, 30 : 1201 - 1205
  • [43] Biophysics of radiofrequency ablation
    Haemmerich D.
    Critical Reviews in Biomedical Engineering, 2010, 38 (01) : 53 - 63
  • [44] Prostate Radiofrequency Focal Ablation (ProRAFT) Trial: A Prospective Development Study Evaluating a Bipolar Radiofrequency Device to Treat Prostate Cancer
    Orczyk, Clement
    Barratt, Dean
    Brew-Graves, Chris
    Hu, Yi Peng
    Freeman, Alex
    McCartan, Neil
    Potyka, Ingrid
    Ramachandran, Navin
    Rodell, Rachael
    Williams, Norman R.
    Emberton, Mark
    Ahmed, Hashim U.
    JOURNAL OF UROLOGY, 2021, 205 (04): : 1090 - 1098
  • [45] Stereotactic Radiofrequency Ablation
    Reto Bale
    Gerlig Widmann
    Marion Haidu
    CardioVascular and Interventional Radiology, 2011, 34 : 852 - 856
  • [46] Radiofrequency ablation of cancer
    Friedman, M
    Mikityansky, I
    Kam, A
    Libutti, SK
    Walther, MM
    Neeman, Z
    Locklin, JK
    Wood, BJ
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2004, 27 (05) : 427 - 434
  • [47] Pathomorphological changes after radiofrequency ablation in the liver
    Kasper, Hans-Udo
    Bangard, Christopher
    Gossmann, Axel
    Dienes, Hans Peter
    Stippel, Dirk Ludger
    PATHOLOGY INTERNATIONAL, 2010, 60 (03) : 149 - 155
  • [48] Radiofrequency Ablation: Temperature Distribution in Adjacent Tissues
    Bornemann, R.
    Groetz, S. F.
    Pennekamp, P. H.
    Wilhelm, K. E.
    Sander, K.
    Wirtz, D. C.
    Pflugmacher, R.
    ZEITSCHRIFT FUR ORTHOPADIE UND UNFALLCHIRURGIE, 2016, 154 (03): : 294 - 298
  • [49] Stereotactic Radiofrequency Ablation of Liver Tumors in Octogenarians
    Schullian, Peter
    Putzer, Daniel
    Silva, Michael A.
    Laimer, Gregor
    Kolbitsch, Christian
    Bale, Reto
    FRONTIERS IN ONCOLOGY, 2019, 9
  • [50] Percutaneous Radiofrequency Lung Ablation Combined with Transbronchial Saline Injection: An Experimental Study in Swine
    Kawai, T.
    Kaminou, T.
    Sugiura, K.
    Hashimoto, M.
    Ohuchi, Y.
    Adachi, A.
    Fujioka, S.
    Ito, H.
    Nakamura, K.
    Ihaya, T.
    Ogawa, T.
    CARDIOVASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 33 (01) : 143 - 149