Effect of specimen fixation method on pullout tests of pedicle screws

被引:80
作者
Pfeiffer, M
Gilbertson, LG
Goel, VK
Griss, P
Keller, JC
Ryken, TC
Hoffman, HE
机构
[1] UNIV IOWA,IOWA SPINE RES CTR,IOWA CITY,IA
[2] UNIV MARBURG,ORTHOPAD KLIN,MARBURG,GERMANY
[3] UNIV PITTSBURGH,DEPT ORTHOPAED SURG,MUSCULOSKELETAL RES CTR,PITTSBURGH,PA
[4] UNIV IOWA,DEPT DENT,IOWA CITY,IA 52242
[5] UNIV IOWA,DEPT NEUROSURG,IOWA CITY,IA 52242
关键词
biomechanics; spine; lumbar vertebrae; pedicle screw; PMMA; scanning electron microscopy; energy dispersive x-ray spectroscopy; dual energy x-ray absorptiometry;
D O I
10.1097/00007632-199605010-00009
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. Experimental axial pullout tests of a new type of pedicle screw were done on cadaveric lumbar vertebrae. The manner in which specimens were secured in the testing apparatus was varied to determine, influence of Specimen fixation method on the maximum pedicle screw pullout force. Objectives. To determine the appropriateness of embedding (i.e., potting) spinal specimens in polymer resin (e.g., bone cement or Plastic Padding. [Plastic Padding Ltd., High Wycombe, Buckinghamshire, Engrand]) for axial pullout tests of pedicle screws. Several different specimen fixation methods were examined to make recommendations for the standardization of future experimental testing protocols. Summary of Background Data. Axial pullout of transpedicular screws, although not a likely clinical mode of failure, is a popular experimental testing mode for evaluating screw-bone biomechanics. A wide variety of techniques for securing a vertebral specimen to counter the axial pullout force has been reported (including the use of polymer resin) with a correspondingly wide range in the resulting axial pullout strengths. The possible influence of the specimen fixation method on pedicle screw axial pullout strength has not been addressed previously. Methods. Axial pullout tests of pedicle screws (DDS, Plus Endoprothetik, Rotkreuz, Switzerland) from the pedicles of 27 isolated lumbar vertebral bodies were done using a Model 810 MTS Universal Testing Machine (MTS Systems, Inc., Minneapolis; Minnesota). The specimens were secured in a custom-made vise fixture either as is or after the vertebral bodies were potted in Plastic Padding up to the pedicle origin. Some of the potted specimens were wrapped first in latex to prevent polymer resin intrusion, and the others were unprotected. Pullout tests were attempted on both the left and right pedicles of each specimen, and the maximum pedicle screw pullout force was recorded. Measurement of bone mineral density by means of dual energy x-ray absorptiometry, in addition to macroscopic and scanning electron microscopy histologic analyses, microradiography, and energy dispersive X-ray spectroscopy, was done post-test to assist in the interpretation of the data. Results. The maximum pedicle screw pullout force was found to be dependent on both the bone mineral density and the mode of fixation of the vertebrae. Embedding in polymer resin without protection of the specimen (i.e., latex wrapping) led to several instances of well-documented polymer resin intrusion; in these specimens, mean maximum pedicle screw pullout force was significantly greater than that of specimens secured without polymer resin and that of embedded specimens for which intrusion did not occur. Conclusions. Polymer resin intrusion can have a significant effect on the biomechanical characteristics of the bone-pedicle screw interface. When polymer resins are used to secure vertebral specimens for in vitro biomechanical tests of the bone-pedicle screw interface, it is important to either prevent intrusion (e.g., with a latex wrapping) or document post-test (e.g., through the methods described in this article) that intrusion did not occur for the specimens included in the analysis.
引用
收藏
页码:1037 / 1044
页数:8
相关论文
共 19 条
  • [1] INTERSITE COMPARISON OF THE HOLOGIC QDR-1000 DUAL ENERGY X-RAY BONE DENSITOMETER
    BLAKE, GM
    TONG, CM
    FOGELMAN, I
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (761) : 440 - 446
  • [2] INFLUENCE OF BONE-MINERAL DENSITY ON THE FIXATION OF THORACOLUMBAR IMPLANTS - A COMPARATIVE-STUDY OF TRANSPEDICULAR SCREWS, LAMINAR HOOKS, AND SPINOUS PROCESS WIRES
    COE, JD
    WARDEN, KE
    HERZIG, MA
    MCAFEE, PC
    [J]. SPINE, 1990, 15 (09) : 902 - 907
  • [3] GEORGE DC, 1991, SPINE, V16, P181, DOI 10.1097/00007632-199116020-00015
  • [4] RELATION OF VERTEBRAL BONE SCREW AXIAL PULLOUT STRENGTH TO QUANTITATIVE COMPUTED TOMOGRAPHIC TRABECULAR BONE-MINERAL CONTENT
    GILBERT, SG
    JOHNS, PC
    CHOW, DC
    BLACK, RC
    [J]. JOURNAL OF SPINAL DISORDERS, 1993, 6 (06): : 513 - 521
  • [5] Krag M H, 1988, J Spinal Disord, V1, P287
  • [6] KRAG MH, 1986, CLIN ORTHOP RELAT R, V203, P75
  • [7] KRIZAN K, 1994, J HISTOTECHNOL, V75, P333
  • [8] DUAL-ENERGY X-RAY ABSORPTIOMETRY OF LUMBAR VERTEBRAE - RELATIVE CONTRIBUTION OF BODY AND POSTERIOR ELEMENTS AND ACCURACY IN RELATION WITH NEUTRON-ACTIVATION ANALYSIS
    LOUIS, O
    VANDENWINKEL, P
    COVENS, P
    SCHOUTENS, A
    OSTEAUX, M
    [J]. BONE, 1992, 13 (04) : 317 - 320
  • [9] Magerl F[., 1982, Current Concepts of External Fixation of Fractures, P353
  • [10] BONE DENSITOMETRY OF EXCISED VERTEBRAE - ANATOMICAL RELATIONSHIPS
    MAZESS, RB
    PEDERSEN, P
    VETTER, J
    BARDEN, HS
    [J]. CALCIFIED TISSUE INTERNATIONAL, 1991, 48 (06) : 380 - 386