Use of polarity switching for the simultaneous bioanalysis of analytes with three orders of magnitude difference in concentration by LC-MS/MS

被引:10
作者
Bergeron, Melanie [1 ]
Bergeron, Annik [1 ]
van Amsterdam, Peter [2 ]
Furtado, Milton [1 ]
Garofolo, Fabio [1 ]
机构
[1] Algorithme Pharma Inc, Laval, PQ H7V 4B3, Canada
[2] Abbott Healthcare Prod, NL-1381 CP Weesp, Netherlands
关键词
CHROMATOGRAPHY-MASS-SPECTROMETRY; ELECTROSPRAY-IONIZATION; MEBEVERINE; SENSITIVITY; MATRIX;
D O I
10.4155/bio.13.142
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Background: The challenge of quantifying two compounds in a single assay with drastic dynamic ranges is to obtain linearity without source or detector saturation at the mass spectrometer. Results: In positive-ionization mode, the nonlinear relationships for Desmethyl Mebeverine Acid (DMAC) were demonstrated using three common strategies to overcome this issue: using offset voltage parameters, less-sensitive product ion or 13C mass SRM transitions. On the contrary, nonlinear relationships for DMAC were overcome if negative-ionization mode was used. Due to Mebeverine analytical LLOQ, dilution was not suitable for a single assay of Mebeverine and DMAC. However, polarity switching in negative mode for DMAC was successfully found to compensate for the nonlinearity at the mass spectrometer while preserving Mebeverine linear regression model in positive mode. Conclusion: The polarity switching strategy has demonstrated the advantage of improving linearity for analytes having different ionization polarities and three orders of magnitude difference in concentration.
引用
收藏
页码:1911 / 1918
页数:8
相关论文
共 21 条