Biophysical Modeling to Determine the Optimization of Left Ventricular Pacing Site and AV/VV Delays in the Acute and Chronic Phase of Cardiac Resynchronization Therapy

被引:21
作者
Lee, Angela W. C. [1 ]
Crozier, Andrew [2 ]
Hyde, Eoin R. [1 ]
Lamata, Pablo [1 ]
Truong, Michael [1 ]
Sohal, Manav [1 ]
Jackson, Thomas [1 ]
Behar, Jonathan M. [1 ]
Claridge, Simon [1 ]
Shetty, Anoop [1 ]
Sammut, Eva [1 ]
Plank, Gernot [2 ]
Rinaldi, Christopher Aldo [1 ,3 ]
Niederer, Steven [1 ]
机构
[1] Kings Coll London, Div Imaging Sci & Biomed Engn, London, England
[2] Med Univ Graz, Inst Biophys, Graz, Austria
[3] Guys & St Thomas NHS Fdn Trust, Cardiovasc Dept, London, England
基金
英国工程与自然科学研究理事会;
关键词
atrioventricular delay; cardiac resynchronization therapy; computer modeling/simulations; interventricular delay; left bundle branch block; left ventricular lead placement; ACUTE HEMODYNAMIC-RESPONSE; HEART-FAILURE; LEAD POSITION; IMPROVEMENT; STIMULATION; MULTISITE; SURVIVAL; IMPACT; TRIAL;
D O I
10.1111/jce.13134
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Device Optimization for Acute and Chronic CRT. Background: Cardiac anatomy and function adapt in response to chronic cardiac resynchronization therapy (CRT). The effects of these changes on the optimal left ventricle (LV) lead location and timing delay settings have yet to be fully explored. Objective: To predict the effects of chronic CRT on the optimal LV lead location and device timing settings over time. Methods: Biophysical computational cardiac models were generated for 3 patients, immediately post-implant (ACUTE) and after at least 6 months of CRT (CHRONIC). Optimal LV pacing area and device settings were predicted by pacing the ACUTE and CHRONIC models across the LV epicardium (49 sites each) with a range of 9 pacing settings and simulating the acute hemodynamic response (AHR) of the heart. Results: There were statistically significant differences between the distribution of the AHR in the ACUTE and CHRONIC models (P < 0.0005 in all cases). The site delivering the maximal AHR shifted location between the ACUTE and CHRONIC models but provided a negligible improvement (<2%). The majority of the acute optimal LV pacing regions (76-100%) and device settings (76-91%) remained optimal chronically. Conclusion: Optimization of the LV pacing location and device settings were important at the time of implant, with a reduced benefit over time, where the majority of the acute optimal LV pacing region and device settings remained optimal with chronic CRT.
引用
收藏
页码:208 / 215
页数:8
相关论文
共 35 条
  • [31] Guidelines for cardiac pacing and cardiac resynchronization therapy
    Vardas, Panos
    [J]. EUROPACE, 2007, 9 (10): : 959 - 998
  • [32] Optimal Left Ventricular Lead Position Predicts Reverse Remodeling and Survival After Cardiac Resynchronization Therapy
    Ypenburg, Claudia
    van Bommel, Rutger J.
    Delgado, Victoria
    Mollema, Sjoerd A.
    Bleeker, Gabe B.
    Boersma, Eric
    Schalij, Martin J.
    Bax, Jeroen J.
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2008, 52 (17) : 1402 - 1409
  • [33] Left ventricular reverse remodeling but not clinical improvement predicts long-term survival after cardiac resynchronization therapy
    Yu, CM
    Bleeker, GB
    Fung, JWH
    Schalij, MJ
    Zhang, Q
    van der Wall, EE
    Chan, YS
    Kong, SL
    Bax, JJ
    [J]. CIRCULATION, 2005, 112 (11) : 1580 - 1586
  • [34] Multipoint pacing by a left ventricular quadripolar lead improves the acute hemodynamic response to CRT compared with conventional biventricular pacing at any site
    Zanon, Francesco
    Baracca, Enrico
    Pastore, Gianni
    Marcantoni, Lina
    Fraccaro, Chiara
    Lanza, Daniela
    Picariello, Claudio
    Aggio, Silvio
    Roncon, Loris
    Dell'Avvocata, Fabio
    Rigatelli, GianLuca
    Pacetta, Domenico
    Noventa, Franco
    Prinzen, Frits W.
    [J]. HEART RHYTHM, 2015, 12 (05) : 975 - 981
  • [35] Determination of the Longest Intrapatient Left Ventricular Electrical Delay May Predict Acute Hemodynamic Improvement in Patients After Cardiac Resynchronization Therapy
    Zanon, Francesco
    Baracca, Enrico
    Pastore, Gianni
    Fraccaro, Chiara
    Roncon, Loris
    Aggio, Silvio
    Noventa, Franco
    Mazza, Alberto
    Prinzen, Frits
    [J]. CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2014, 7 (03) : 377 - 383