Hydrophilic interaction chromatography in nonaqueous elution mode for separation of hydrophilic analytes on silica-based packings with noncharged polar bondings

被引:78
作者
Bicker, Wolfgang [1 ]
Wu, JunYan [1 ]
Laemmerhofer, Michael [1 ]
Lindner, Wolfgang [1 ]
机构
[1] Univ Vienna, Christian Doppler Lab Mol Recognit Mat, Dept Analyt Chem & Food Chem, A-1090 Vienna, Austria
关键词
diol stationary phases; high-performance liquid chromatography; hydrophilic interaction chromatography; nonaqueous chromatography; polar-organic elution mode;
D O I
10.1002/jssc.200800246
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Chromatographic effects of dedicated stationary and mobile phase variations in hydrophilic interaction chromatography (HILIC) were investigated using a set of nucleobases, nucleosides and deoxynucleosides as polar test solutes. Retention and selectivity profiles were comparatively mapped on four in-house developed silica materials modified with short alkyl chains (C-4, C-5) which carry hydroxyl functionalities (including diol motifs) as well as embedded sulphide or sulphoxide groups. These data were complemented by results obtained with two commercially available diol-type phases and a bare silica column. Besides elucidation of packing-related aspects this work concentrated specifically on extending aqueous HILIC (AQ-HILIC) to nonaqueous polar-organic elution conditions herein termed NA-HILIC. The exchange of the polar modifier water by various alcohols in ACN-rich mobile phases containing 5 mM ammonium acetate decreased the eluotropic strength of the resulting eluents. The gain in retention largely followed the order ethanol (EtOH) > methanol (MeOH) > 1,2-ethanediol (Et(OH)(2) and was accompanied by distinct effects on chromatographic selectivity. For example, on the most polar home-made packin'a the purine nucleoside selectivity guanosine/adenosine increased from 2.25 in the AQ-HILIC (k(guanosine) = 8.3) to 7.33 (k(guanosine) = 59) in the NA-HILIC mode when EtOH was employed as NA modifier while this value was 5.84 and 2.93 with MeOH and Et(OH)2, respectively (eluent: 5 mM ammonium acetate in ACN/modifier 90:10 v/v). Besides the type of protic modifier its percentage as well the retention and selectivity effects upon varying the ammonium acetate concentration and column temperature, respectively, were also investigated. Notable inter-column differences were found for all of these elution parameters. A mixed-mode retention model composed of partitioning and adsorption is proposed for both AQ- and NA-HILIC retention processes. The potential of (i) the implementation of novel polar bondings (such as ones containing sulphoxide functionalities) and (ii) the comprehensive exploitation of elution variables (type of protic modifiers, salt, ctc.) for providing new selectivity increments to the separation of polar analytes in HILIC is emphasised.
引用
收藏
页码:2971 / 2987
页数:17
相关论文
共 38 条
  • [1] HYDROPHILIC-INTERACTION CHROMATOGRAPHY FOR THE SEPARATION OF PEPTIDES, NUCLEIC-ACIDS AND OTHER POLAR COMPOUNDS
    ALPERT, AJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1990, 499 : 177 - 196
  • [2] Retention pattern profiling of fungal metabolites on mixed-mode reversed-phase/weak anion exchange stationary phases in comparison to reversed-phase and weak anion exchange separation materials by liquid chromatography-electrospray ionisation-tandem mass spectrometry
    Apfelthaler, Elisabeth
    Bicker, Wolfgang
    Laemmerhofer, Michael
    Sulyok, Michael
    Krska, Rudolf
    Lindner, Wolfgang
    Schuhmacher, Rainer
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2008, 1191 (1-2) : 171 - 181
  • [3] Practice and mechanism of HPLC oligosaccharide separation with a cyclodextrin bonded phase
    Berthod, A
    Chang, SSC
    Kullman, JPS
    Armstrong, DW
    [J]. TALANTA, 1998, 47 (04) : 1001 - 1012
  • [4] Direct high-performance liquid chromatographic method for enantioselective and diastereoselective determination of selected pyrethroic acids
    Bicker, W
    Lämmerhofer, M
    Lindner, W
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2004, 1035 (01) : 37 - 46
  • [5] validated method for the determination of the ethanol consumption markers ethyl glucuronide, ethyl phosphate, and ethyl sulfate in human urine by reversed-phase/weak anion exchange liquid chromatography-tandem mass spectrometry
    Bicker, Wolfgang
    Laemmerhofer, Michael
    Keller, Thomas
    Schuhmacher, Rainer
    Krska, Rudolf
    Lindner, Wolfgang
    [J]. ANALYTICAL CHEMISTRY, 2006, 78 (16) : 5884 - 5892
  • [6] EVALUATION OF A NEW POLAR ORGANIC HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC MOBILE PHASE FOR CYCLODEXTRIN-BONDED CHIRAL STATIONARY PHASES
    CHANG, SC
    REID, GL
    CHEN, S
    CHANG, CD
    ARMSTRONG, DW
    [J]. TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1993, 12 (04) : 144 - 153
  • [7] Effect of phase ratio on van't Hoff analysis in reversed-phase liquid chromatography, and phase-ratio-independent estimation of transfer enthalpy
    Chester, TL
    Coym, JW
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2003, 1003 (1-2) : 101 - 111
  • [8] Hydrophilic interaction liquid chromatography-mass spectrometry for the analysis of paralytic shellfish poisoning (PSP) toxins
    Dell'Aversano, C
    Hess, P
    Quilliam, MA
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2005, 1081 (02) : 190 - 201
  • [9] Grumbach ES, 2004, LC GC N AM, V22, P1010
  • [10] Investigating the effect of chromatographic conditions on retention of organic acids in hydrophilic interaction chromatography using a design of experiment
    Guo, Y.
    Srinivasan, S.
    Gaiki, S.
    [J]. CHROMATOGRAPHIA, 2007, 66 (3-4) : 223 - 229