Measurement techniques and results of an intercomparison for RF power in a 3.5 mm coaxial line up to 26 GHz

被引:3
作者
Janik, D [1 ]
Ascroft, JT
Clark, RF
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
[2] Natl Phys Lab, Teddington TW11 0LW, Middx, England
[3] Natl Res Council Canada, INMS, Ottawa, ON K1A 0R6, Canada
关键词
bolometer mounts; calibration; calorimetry; electromagnetic measurements; measurement standards; microwave power;
D O I
10.1109/19.769554
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The methods and results of an international comparison are presented for measurements of RF power in a 3.5 mm coaxial line at seven frequencies up to 26 GHz. Three national laboratories of metrology having different measurement techniques participated in the comparison. The agreement between the mean values of the calibration factor of the traveling standard reported by the participants is well within their reported uncertainties (k = 2), which are between 0.5% and 3.1%.