Non-contact video-based vital sign monitoring using ambient light and auto-regressive models

被引:343
作者
Tarassenko, L. [1 ]
Villarroel, M. [1 ]
Guazzi, A. [1 ]
Jorge, J. [1 ]
Clifton, D. A. [1 ]
Pugh, C. [2 ]
机构
[1] Univ Oxford, Inst Biomed Engn, Dept Engn Sci, Oxford OX1 2JD, England
[2] Univ Oxford, Nuffield Dept Med, Oxford OX1 2JD, England
基金
英国惠康基金; 英国工程与自然科学研究理事会;
关键词
non-contact monitoring; PPG; video; auto-regressive models; vital signs; dialysis; PULSE OXIMETRY; REFLECTANCE;
D O I
10.1088/0967-3334/35/5/807
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Remote sensing of the reflectance photoplethysmogram using a video camera typically positioned 1 m away from the patient's face is a promising method for monitoring the vital signs of patients without attaching any electrodes or sensors to them. Most of the papers in the literature on non-contact vital sign monitoring report results on human volunteers in controlled environments. We have been able to obtain estimates of heart rate and respiratory rate and preliminary results on changes in oxygen saturation from double-monitored patients undergoing haemodialysis in the Oxford Kidney Unit. To achieve this, we have devised a novelmethod of cancelling out aliased frequency components caused by artificial light flicker, using auto-regressive (AR) modelling and pole cancellation. Secondly, we have been able to construct accurate maps of the spatial distribution of heart rate and respiratory rate information from the coefficients of the AR model. In stable sections with minimal patient motion, the mean absolute error between the camera-derived estimate of heart rate and the reference value from a pulse oximeter is similar to the mean absolute error between two pulse oximeter measurements at different sites (finger and earlobe). The activities of daily living affect the respiratory rate, but the cameraderived estimates of this parameter are at least as accurate as those derived from a thoracic expansion sensor (chest belt). During a period of obstructive sleep apnoea, we tracked changes in oxygen saturation using the ratio of normalized reflectance changes in two colour channels (red and blue), but this required calibration against the reference data from a pulse oximeter.
引用
收藏
页码:807 / 831
页数:25
相关论文
共 42 条
[1]   Forehead pulse oximetry - Headband use helps alleviate false low readings likely related to venous pulsation artifact [J].
Agashe, Geeta S. ;
Coakley, Joseph ;
Mannheimer, Paul D. .
ANESTHESIOLOGY, 2006, 105 (06) :1111-1116
[2]  
[Anonymous], 2012, ACM Transactions on Graphics (TOG), DOI DOI 10.1145/2185520.2185561
[3]   Detecting Pulse from Head Motions in Video [J].
Balakrishnan, Guha ;
Durand, Fredo ;
Guttag, John .
2013 IEEE CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR), 2013, :3430-3437
[4]  
Borhani Y, 2010, COMPUT CARDIOL CONF, V37, P967
[5]   Continuous wavelet filtering on webcam photoplethysmographic signals to remotely assess the instantaneous heart rate [J].
Bousefsaf, Frederic ;
Maaoui, Choubeila ;
Pruski, Alain .
BIOMEDICAL SIGNAL PROCESSING AND CONTROL, 2013, 8 (06) :568-574
[6]   Tracking poles with an autoregressive model: a confidence index for the analysis of the intrapartum cardiotocogram [J].
Cazares, S ;
Moulden, M ;
Redman, CWG ;
Tarassenko, L .
MEDICAL ENGINEERING & PHYSICS, 2001, 23 (09) :603-614
[7]  
Cennini G, 2010, WO Patent Application, Patent No. 2010100593
[8]   Heart rate monitoring via remote photoplethysmography with motion artifacts reduction [J].
Cennini, Giovanni ;
Arguel, Jeremie ;
Aksit, Kaan ;
van Leest, Arno .
OPTICS EXPRESS, 2010, 18 (05) :4867-4875
[9]   Pulse oximetry: Understanding its basic principles facilitates appreciation of its limitations [J].
Chan, Edward D. ;
Chan, Michael M. ;
Chan, Mallory M. .
RESPIRATORY MEDICINE, 2013, 107 (06) :789-799
[10]  
Cretikos MA, 2008, MED J AUSTRALIA, V188, P657