共 50 条
Catheter-delivered ultrasound potentiates in vitro thrombolysis
被引:30
|作者:
ShlanskyGoldberg, RD
[1
]
Cines, DB
[1
]
Sehgal, CM
[1
]
机构:
[1] HOSP UNIV PENN,DEPT PATHOL & LAB MED,PHILADELPHIA,PA 19104
关键词:
D O I:
10.1016/S1051-0443(96)72861-0
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
PURPOSE: To develop a catheter-directed method to enhance urokinase-mediated thrombolysis with use of ultrasound. MATERIALS AND METHODS: A prototype catheter was constructed by using a 9-F piezoelectric crystal capable of producing 640-kHz pulsed ultrasound energy. Clots formed in vitro from whole blood were trace-labeled with iodine-125 fibrinogen, and the release of radiolabeled fibrin degradation products was measured in the presence of urokinase, ultrasound, or a combination of urokinase and ultrasound. RESULTS: By 30 minutes, clot lysis was more complete with urokinase plus ultrasound (78.7% +/- 5.3 [mean +/- SD]) than with ultrasound alone (19.3% +/- 10.0) or urokinase alone (47.9% +/- 10.0) (P <.001 for ultrasound and urokinase vs either alone), The time to 50% clot lysis was shortened by 46% on average with the application of urokinase and ultrasound compared with urokinase alone (P <.03). CONCLUSIONS: Catheter-based ultrasound enhances enzymatic thrombolysis in vitro and may be a practical means to reduce the dose of enzyme and the time needed to achieve clot lysis in vivo.
引用
收藏
页码:313 / 320
页数:8
相关论文