Venous blood redistribution alters the accuracy of Omron HEM-705CP

被引:5
作者
Czarkowski, M
Mikulska, M
Zebrowski, M
Baran, A
Rózanowski, K
机构
[1] Med Univ Warsaw, Dept Internal Med & Endocrinol, PL-02097 Warsaw, Poland
[2] Mil Inst Aerosp Med, Dept Physiol, Warsaw, Poland
关键词
blood pressure measurement; oscillometric method; OMRON HEM-705CP; lower body negative pressure stimulus;
D O I
10.1097/00126097-200402000-00003
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Background Validation procedures of automated blood pressure measuring devices are carried out at rest only. Design We aimed to determine whether venous blood redistribution affects the accuracy of the OMRON HEM-705CP (Osc). Lower body negative pressure stimulus (-40 mmHg) (LBNP) was used to simulate changes of body position. Methods Fifty-four young healthy volunteers had their blood pressure (BP) measured simultaneously by Osc and mercury sphygmomanometer (HgS) at rest (at 3 and 5 min) and during LBNP (at 7 and 9 min). Results Differences (A) between HgS and Osc for systolic and diastolic BP (SBP and DBP) were calculated. The DeltaSBP obtained in minute 9 (DeltaSBP-9) was bigger than the one taken in minute 5 (DeltaSBP-5) (-2.89 +/- 4.3 versus -1.35 +/- 5.2 mmHg, P < 0.02). The rest of the DeltaSBP values and all the DeltaDBP values were constant at rest and during venous blood redistribution caused by LBNP. We calculated the fall in SBP separately for HgS and Osc by subtracting SBP-9 from SBP-5 for each patient. The fall measured by Osc was bigger than by HgS (-5.87 +/- 4.0 versus -4.3 +/- 4.5 mmHg; P < 0.02). When graded according to British Hypertension Society protocol, Osc remained in the class of recommendation (A or B) for every minute of the procedure. Conclusions Even if an oscillometric device passes various validation tests, they do not give. sufficient information about its accuracy during venous blood redistribution. This knowledge is of particular importance when BP regulation abnormalities occurring during changes in body position are diagnosed with the oscillometric method. (C) 2004 Lippincott Williams Wilkins.
引用
收藏
页码:9 / 12
页数:4
相关论文
共 18 条
[1]  
[Anonymous], 1987, AM NAT STAND EL AUT
[2]   CHARACTERIZATION OF THE OSCILLOMETRIC METHOD FOR MEASURING INDIRECT BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, M ;
COMBS, C ;
REINER, D ;
BABBS, CF .
ANNALS OF BIOMEDICAL ENGINEERING, 1982, 10 (06) :271-280
[3]   FOREARM VASCULAR RESPONSES TO LOWER BODY NEGATIVE PRESSURE AND ORTHOSTASIS [J].
GILBERT, CA ;
STEVENS, PM .
JOURNAL OF APPLIED PHYSIOLOGY, 1966, 21 (04) :1265-&
[4]  
Iqbal P., 1996, Blood Pressure, V5, P222, DOI 10.3109/08037059609079675
[5]   2003 European society of hypertension -: European Society of Cardiology guidelines for the management of arterial hypertension [J].
Mancia, G ;
Rosei, EA ;
Cifkova, R ;
DeBacker, G ;
Erdine, S ;
Fagard, R ;
Farsang, C ;
Heagerty, AM ;
Kawecka-Jaszcs, K ;
Kiowski, W ;
Kjeldsen, S ;
Lüscher, T ;
McInnes, G ;
Mallion, JM ;
Brien, EO ;
Poulter, NR ;
Priori, SG ;
Rahn, KH ;
Rodicio, JL ;
Ruilope, LM ;
Safar, M ;
Staessen, JA ;
van Zwieten, P ;
Waeber, B ;
Williams, B ;
Zanchetti, A ;
Zannad, F .
JOURNAL OF HYPERTENSION, 2003, 21 (06) :1011-1053
[6]   The mercury sphygmomanometer should be abandoned before it is proscribed [J].
Markandu, ND ;
Whitcher, F ;
Arnold, A ;
Carney, C .
JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (01) :31-36
[7]  
MUSGRAVE FS, 1971, AEROSPACE MED, V42, P1065
[8]  
MUSGRAVE FS, 1969, AEROSPACE MED, V40, P602
[9]  
O'Brien, 1996, Blood Press Monit, V1, P55
[10]   Replacing the mercury sphygmomanometer [J].
O'Brien, E .
BRITISH MEDICAL JOURNAL, 2000, 320 (7238) :815-816