Determination of trace bisphenol A in environmental water by high-performance liquid chromatography using magnetic reduced graphene oxide based solid-phase extraction coupled with dispersive liquid-liquid microextraction

被引:71
作者
Li, Danping [1 ]
Ma, Xiaoguo [1 ]
Wang, Rui [1 ]
Yu, Yumian [1 ]
机构
[1] Guangdong Univ Technol, Sch Environm Sci & Engn, Guangzhou 510006, Guangdong, Peoples R China
基金
中国国家自然科学基金;
关键词
Bisphenol A; Magnetic reduced graphene oxide; Solid-phase extraction; Dispersive liquid-liquid microextraction; Gas chromatography-mass spectrometry; Environmental water; CARBON NANOTUBES; AQUEOUS-SOLUTION; IONIC LIQUID; SAMPLES; NANOPARTICLES; ADSORPTION; SPECIATION; SORBENT; MERCURY;
D O I
10.1007/s00216-016-0087-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Bisphenol A (BPA), an endocrine-disrupting chemical, has received much attention from researchers and the general public. In this paper, a novel method of determining BPA at trace levels was developed, using magnetic reduced graphene oxide (rGO-Fe3O4)-based solid-phase extraction coupled with dispersive liquid-liquid microextraction (DLLME), followed by high-performance liquid chromatographic determination. The rGO-Fe3O4 was prepared and then characterized by Fourier transform infrared spectroscopy, transmission electron microscopy, and vibrating sample magnetometry. The greatest saturation magnetization of rGO-Fe3O4 was up to 43.8 emu g(-1), which allowed rapid isolation of the rGO-Fe3O4 from solutions upon applying an appropriate magnetic field. The effects of solution pH, adsorbent amount, type and volume of eluent and extraction solvent, extraction time, and salt concentration on the extraction efficiency of BPA were examined and optimized. Under the optimum conditions, an enrichment factor of 5217 and an LOD of 0.01 mu g L-1 for BPA were obtained. The reusability of rGO-Fe3O4 for at least 12 repeated cycles without any significant decrease in the extraction recovery of BPA was demonstrated. The proposed method was applied to the determination of BPA in different real water samples, with relative recoveries of 84.8-104.9 % and RSDs of 0.8-8.3 % in the spiked concentration range 1-10 mu g L-1.
引用
收藏
页码:1165 / 1172
页数:8
相关论文
共 35 条
  • [1] Combined coarse-grained and atomistic simulation of liquid bisphenol A-polycarbonate: Liquid packing and intramolecular structure
    Abrams, CF
    Kremer, K
    [J]. MACROMOLECULES, 2003, 36 (01) : 260 - 267
  • [2] Micro versus macro solid phase extraction for monitoring water contaminants: A preliminary study using trihalomethanes
    Alexandrou, Lydon D.
    Spencer, Michelle J. S.
    Morrison, Paul D.
    Meehan, Barry J.
    Jones, Oliver A. H.
    [J]. SCIENCE OF THE TOTAL ENVIRONMENT, 2015, 512 : 210 - 214
  • [3] Determination of Trace Amounts of Cd(II), Cu(II), and Ni(II) in Food Samples Using a Novel Functionalized Magnetic Nanosorbent
    Bagheri, Hasan
    Asgharinezhad, Ali Akbar
    Ebrahimzadeh, Homeira
    [J]. FOOD ANALYTICAL METHODS, 2016, 9 (04) : 876 - 888
  • [4] Occurrence of bisphenol A in surface water and uptake in fish: evaluation of field measurements
    Belfroid, A
    van Velzen, M
    van der Horst, B
    Vethaak, D
    [J]. CHEMOSPHERE, 2002, 49 (01) : 97 - 103
  • [5] Multiwalled carbon nanotubes as a solid-phase extraction adsorbent for the determination of bisphenol a, 4-n-nonylphenol, and 4-tert-octylphenol
    Cai, YQ
    Jiang, GB
    Liu, JF
    Zhou, QX
    [J]. ANALYTICAL CHEMISTRY, 2003, 75 (10) : 2517 - 2521
  • [6] Water-Dispersible Magnetite-Reduced Graphene Oxide Composites for Arsenic Removal
    Chandra, Vimlesh
    Park, Jaesung
    Chun, Young
    Lee, Jung Woo
    Hwang, In-Chul
    Kim, Kwang S.
    [J]. ACS NANO, 2010, 4 (07) : 3979 - 3986
  • [7] Use of a new magnetic ion-imprinted nanocomposite adsorbent for selective and rapid preconcentration and determination of trace nickel by flame atomic absorption spectrometry
    Chen, Yuan
    Ma, Xiaoguo
    Huang, Minghua
    Peng, Junbiao
    Li, Chunying
    [J]. ANALYTICAL METHODS, 2016, 8 (04) : 824 - 829
  • [8] Silica Gel Chemically Modified with Ionic Liquid as Novel Sorbent for Solid-Phase Extraction and Preconcentration of Lead from Beer and Tea Drink Samples Followed by Flame Atomic Absorption Spectrometric Determination
    Cheng, Jufang
    Ma, Xiaoguo
    Wu, Yuguang
    [J]. FOOD ANALYTICAL METHODS, 2014, 7 (05) : 1083 - 1089
  • [9] Nano polypyrrole-coated magnetic solid phase extraction followed by dispersive liquid phase microextraction for trace determination of megestrol acetate and levonorgestrel
    Ebrahimpour, Behnam
    Yamini, Yadollah
    Seidi, Shahram
    Tajik, Mohammad
    [J]. ANALYTICA CHIMICA ACTA, 2015, 885 : 98 - 105
  • [10] Fan YC, 2008, CHINESE J ANAL CHEM, V36, P1157