Single-shot spiral imaging enabled by an expanded encoding model: Demonstration in diffusion MRI

被引:50
|
作者
Wilm, Bertram J. [1 ,2 ]
Barmet, Christoph [1 ,2 ,3 ]
Gross, Simon [1 ,2 ]
Kasper, Lars [1 ,2 ,4 ]
Vannesjo, S. Johanna [1 ,2 ]
Haeberlin, Max [1 ,2 ]
Dietrich, Benjamin E. [1 ,2 ]
Brunner, David O. [1 ,2 ]
Schmid, Thomas [1 ,2 ]
Pruessmann, Klaas P. [1 ,2 ]
机构
[1] Univ Zurich, Inst Biomed Engn, Gloriastr 35, CH-8092 Zurich, Switzerland
[2] Swiss Fed Inst Technol, Gloriastr 35, CH-8092 Zurich, Switzerland
[3] Skope Magnetic Resonance Technol Inc, Zurich, Switzerland
[4] Univ Zurich, Inst Biomed Engn, Translat Neuromodeling Unit, Zurich, Switzerland
关键词
single-shot spiral; algebraic image reconstruction; magnetic field monitoring; magnetic field probes; DWI; DTI; OFF-RESONANCE CORRECTION; K-SPACE TRAJECTORIES; MAGNETIC-FIELD FLUCTUATIONS; QUANTITATIVE CHARACTERIZATION; RECONSTRUCTION; ECHO; SEPARATION; TENSOR; ACQUISITION; ARTIFACTS;
D O I
10.1002/mrm.26493
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe purpose of this work was to improve the quality of single-shot spiral MRI and demonstrate its application for diffusion-weighted imaging. MethodsImage formation is based on an expanded encoding model that accounts for dynamic magnetic fields up to third order in space, nonuniform static B-0, and coil sensitivity encoding. The encoding model is determined by B-0 mapping, sensitivity mapping, and concurrent field monitoring. Reconstruction is performed by iterative inversion of the expanded signal equations. Diffusion-tensor imaging with single-shot spiral readouts is performed in a phantom and in vivo, using a clinical 3T instrument. Image quality is assessed in terms of artefact levels, image congruence, and the influence of the different encoding factors. ResultsUsing the full encoding model, diffusion-weighted single-shot spiral imaging of high quality is accomplished both in vitro and in vivo. Accounting for actual field dynamics, including higher orders, is found to be critical to suppress blurring, aliasing, and distortion. Enhanced image congruence permitted data fusion and diffusion tensor analysis without coregistration. ConclusionUse of an expanded signal model largely overcomes the traditional vulnerability of spiral imaging with long readouts. It renders single-shot spirals competitive with echo-planar readouts and thus deploys shorter echo times and superior readout efficiency for diffusion imaging and further prospective applications. Magn Reson Med 77:83-91, 2017. (c) 2016 International Society for Magnetic Resonance in Medicine.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 50 条
  • [41] Referenceless one-dimensional Nyquist ghost correction in multicoil single-shot spatiotemporally encoded MRI
    Chen, Ying
    Liao, Yupeng
    Yuan, Lisha
    Liu, Hui
    Yun, Seong Dae
    Shah, Nadim Joni
    Chen, Zhong
    Zhong, Jianhui
    MAGNETIC RESONANCE IMAGING, 2017, 37 : 222 - 233
  • [42] Single-shot lensless imaging based on iterative denoising phase retrieval
    Wei, Jiahao
    Shen, Mengzhe
    Zuo, Chao
    HOLOGRAPHY, DIFFRACTIVE OPTICS, AND APPLICATIONS XII, 2022, 12318
  • [43] Center-out EPI (COEPI): A fast single-shot imaging technique with a short TE
    Chen, Xudong
    Zhu, Ante
    Du, Yiping P.
    MAGNETIC RESONANCE IN MEDICINE, 2020, 84 (02) : 787 - 799
  • [44] Rapid anatomical brain imaging using spiral acquisition and an expanded signal model
    Kasper, Lars
    Engel, Maria
    Barmet, Christoph
    Haeberlin, Maximilian
    Wilm, Bertram J.
    Dietrich, Benjamin E.
    Schmid, Thomas
    Gross, Simon
    Brunner, David O.
    Stephan, Klaas E.
    Pruessmann, Klaas P.
    NEUROIMAGE, 2018, 168 : 88 - 100
  • [45] Reconstruction as a source of artifact in nongated single-shot diffusion-weighted EPI
    Robson, MD
    Porter, DA
    MAGNETIC RESONANCE IMAGING, 2005, 23 (09) : 899 - 905
  • [46] Image quality assessment of single-shot turbo spin echo diffusion-weighted imaging with parallel imaging technique: a phantom study
    Yoshida, Tsukasa
    Urikura, Atsushi
    Shirata, Kensei
    Nakaya, Yoshihiro
    Terashima, Shingo
    Hosokawa, Yoichiro
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1065)
  • [47] On the application of a non-CPMG single-shot fast spin-echo sequence to diffusion tensor MRI of the human brain
    Bastin, ME
    Le Roux, P
    MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (01) : 6 - 14
  • [48] Motion-tolerant diffusion mapping based on single-shot overlapping-echo detachment (OLED) planar imaging
    Ma, Lingceng
    Cai, Congbo
    Yang, Hongyi
    Cai, Shuhui
    Qian, Junchao
    Xiao, Lizhi
    Zhong, Kai
    Chen, Zhong
    MAGNETIC RESONANCE IN MEDICINE, 2018, 80 (01) : 200 - 210
  • [49] Single-shot echo-planar imaging with Nyquist ghost compensation: Interleaved dual echo with acceleration (IDEA) echo-planar imaging (EPI)
    Poser, Benedikt A.
    Barth, Markus
    Goa, Pal-Erik
    Deng, Weiran
    Stenger, V. Andrew
    MAGNETIC RESONANCE IN MEDICINE, 2013, 69 (01) : 37 - 47
  • [50] Flexible reduced field of view magnetic resonance imaging based on single-shot spatiotemporally encoded technique
    Li Jing
    Cai Cong-Bo
    Chen Lin
    Chen Ying
    Qu Xiao-Bo
    Cai Shu-Hui
    CHINESE PHYSICS B, 2015, 24 (10)