Single-shot spiral imaging enabled by an expanded encoding model: Demonstration in diffusion MRI

被引:50
|
作者
Wilm, Bertram J. [1 ,2 ]
Barmet, Christoph [1 ,2 ,3 ]
Gross, Simon [1 ,2 ]
Kasper, Lars [1 ,2 ,4 ]
Vannesjo, S. Johanna [1 ,2 ]
Haeberlin, Max [1 ,2 ]
Dietrich, Benjamin E. [1 ,2 ]
Brunner, David O. [1 ,2 ]
Schmid, Thomas [1 ,2 ]
Pruessmann, Klaas P. [1 ,2 ]
机构
[1] Univ Zurich, Inst Biomed Engn, Gloriastr 35, CH-8092 Zurich, Switzerland
[2] Swiss Fed Inst Technol, Gloriastr 35, CH-8092 Zurich, Switzerland
[3] Skope Magnetic Resonance Technol Inc, Zurich, Switzerland
[4] Univ Zurich, Inst Biomed Engn, Translat Neuromodeling Unit, Zurich, Switzerland
关键词
single-shot spiral; algebraic image reconstruction; magnetic field monitoring; magnetic field probes; DWI; DTI; OFF-RESONANCE CORRECTION; K-SPACE TRAJECTORIES; MAGNETIC-FIELD FLUCTUATIONS; QUANTITATIVE CHARACTERIZATION; RECONSTRUCTION; ECHO; SEPARATION; TENSOR; ACQUISITION; ARTIFACTS;
D O I
10.1002/mrm.26493
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeThe purpose of this work was to improve the quality of single-shot spiral MRI and demonstrate its application for diffusion-weighted imaging. MethodsImage formation is based on an expanded encoding model that accounts for dynamic magnetic fields up to third order in space, nonuniform static B-0, and coil sensitivity encoding. The encoding model is determined by B-0 mapping, sensitivity mapping, and concurrent field monitoring. Reconstruction is performed by iterative inversion of the expanded signal equations. Diffusion-tensor imaging with single-shot spiral readouts is performed in a phantom and in vivo, using a clinical 3T instrument. Image quality is assessed in terms of artefact levels, image congruence, and the influence of the different encoding factors. ResultsUsing the full encoding model, diffusion-weighted single-shot spiral imaging of high quality is accomplished both in vitro and in vivo. Accounting for actual field dynamics, including higher orders, is found to be critical to suppress blurring, aliasing, and distortion. Enhanced image congruence permitted data fusion and diffusion tensor analysis without coregistration. ConclusionUse of an expanded signal model largely overcomes the traditional vulnerability of spiral imaging with long readouts. It renders single-shot spirals competitive with echo-planar readouts and thus deploys shorter echo times and superior readout efficiency for diffusion imaging and further prospective applications. Magn Reson Med 77:83-91, 2017. (c) 2016 International Society for Magnetic Resonance in Medicine.
引用
收藏
页码:83 / 91
页数:9
相关论文
共 50 条
  • [21] Distortion correction of single-shot EPI enabled by deep-learning
    Hu, Zhangxuan
    Wang, Yishi
    Zhang, Zhe
    Zhang, Jieying
    Zhang, Huimao
    Guo, Chunjie
    Sun, Yuejiao
    Guo, Hua
    NEUROIMAGE, 2020, 221
  • [22] Clinical evaluation of single-shot and readout-segmented diffusion-weighted imaging in stroke patients at 3 T
    Morelli, John
    Porter, David
    Ai, Fei
    Gerdes, Clint
    Saettele, Megan
    Feiweier, Thorsten
    Padua, Abraham
    Dix, James
    Marra, Michael
    Rangaswamy, Rajesh
    Runge, Val
    ACTA RADIOLOGICA, 2013, 54 (03) : 299 - 306
  • [23] Considerations in high-resolution skeletal muscle diffusion tensor imaging using single-shot echo planar imaging with stimulated-echo preparation and sensitivity encoding
    Karampinos, Dimitrios C.
    Banerjee, Suchandrima
    King, Kevin F.
    Link, Thomas M.
    Majumdar, Sharmila
    NMR IN BIOMEDICINE, 2012, 25 (05) : 766 - 778
  • [24] Distortion Removal and Deblurring of Single-Shot DWI MRI Scans
    Choudhury, Ahana Roy
    Jambawalikar, Sachin R.
    Kumar, Piyush
    Bommireddy, Venkat Sumanth Reddy
    MACHINE LEARNING FOR MEDICAL IMAGE RECONSTRUCTION (MLMIR 2021), 2021, 12964 : 65 - 75
  • [25] Single-Shot Magnetic Resonance Spectroscopic Imaging with Partial Parallel Imaging
    Posse, Stefan
    Otazo, Ricardo
    Tsai, Shang-Yueh
    Yoshimoto, Akio Ernesto
    Lin, Fa-Hsuan
    MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (03) : 541 - 547
  • [26] Single-shot pseudo-centric EPI for magnetization-prepared imaging
    Lee, Hyun-Soo
    Hwang, Seon-Ha
    Park, Jaeseok
    Park, Sung-Hong
    MAGNETIC RESONANCE IN MEDICINE, 2021, 86 (05) : 2656 - 2665
  • [27] Trust Region Methods for the Estimation of a Complex Exponential Decay Model in MRI With a Single-Shot or Multi-Shot Trajectory
    Hu, Chenxi
    Reeves, Stanley J.
    IEEE TRANSACTIONS ON IMAGE PROCESSING, 2015, 24 (11) : 3694 - 3706
  • [28] Partial Fourier techniques in single-shot cross-term spatiotemporal encoded MRI
    Zhang, Zhiyong
    Frydman, Lucio
    MAGNETIC RESONANCE IN MEDICINE, 2018, 79 (03) : 1506 - 1514
  • [29] Diffusion-weighted imaging of the sellar region: A comparison study of BLADE and single-shot echo planar imaging sequences
    Lu Yiping
    Liu Hui
    Zhou Kun
    Geng Daoying
    Yin Bo
    EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (07) : 1239 - 1244
  • [30] Single-shot holography for depth resolved three dimensional imaging
    Koukourakis, Nektarios
    Kasseck, Christoph
    Rytz, Daniel
    Gerhardt, Nils C.
    Hofmann, Martin R.
    OPTICS EXPRESS, 2009, 17 (23): : 21015 - 21029