Biomechanical, morphological, and histological analysis of early failures in hip resurfacing arthroplasty

被引:125
作者
Morlock, MM
Bishop, N
Ruether, W
Delling, G
Hanh, M
机构
[1] Hamburg Univ Technol, Biomech Sect, Hamburg, Germany
[2] Univ Hamburg, Hosp Eppendorf, Inst Orthopaed Surg, D-20246 Hamburg, Germany
[3] Univ Hamburg, Hosp Eppendorf, Inst Osteopathol, D-20246 Hamburg, Germany
[4] Univ Hamburg, Hosp Eppendorf, Biomech Sect, D-20246 Hamburg, Germany
关键词
wear; deformation; necrosis; failure mechanism; metal-on-metal; surface replacement; learning curve; cementing technique;
D O I
10.1243/095441105X69015
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present revival of hip resurfacing arthroplasty may be related to an increase in early failures owing to the challenging technique of the procedure. Fifty-five retrieved implants were analysed with respect to wear, cement mantle and cement penetration, fracture and head morphology, as well as standard histology. Femoral neck fractures occurred in median after 102 days. The time to failure was shorter for older women. Major deviations from the suggested cement mantle thickness and cement penetration were found. Indications for high trauma during implantation leading to early failure due to weakening of the femoral neck were also observed. Some failures had signs of pseudarthrosis beneath the implant. Four different fracture patterns with different mean survival times were identified. Observed wear was minor with the exception of that due to alignment mistakes (rim loading). The cups were not damaged by the failures. Histological results indicate that avascular necrosis is not necessarily connected with this kind of endoprosthetic surgery Most of the failures analysed can probably be attributed to the 'learning curve' effect, which is an unsatisfactory situation.
引用
收藏
页码:333 / 344
页数:12
相关论文
共 20 条
[1]   Metal-on-metal hybrid surface arthroplasty:: Two to six-year follow-up study [J].
Amstutz, HC ;
Beaulé, PE ;
Dorey, FJ ;
Le Duff, MJ ;
Campbell, PA ;
Gruen, TA .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (01) :28-39
[2]   Fracture of the neck of the femur after surface arthroplasty of the hip [J].
Amstutz, HC ;
Campbell, PA ;
Le Duff, MJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :1874-1877
[3]   Metal-on-metal surface arthroplasty with a cemented femoral component -: A 7-10 year follow-up study [J].
Beaulé, PE ;
Le Duff, M ;
Campbell, P ;
Dorey, FJ ;
Park, SH ;
Amstutz, HC .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :17-22
[4]   Orientation of the femoral component in surface arthroplasty of the hip -: A biomechanical and clinical analysis [J].
Beaulé, PE ;
Lee, JL ;
Le Duff, MJ ;
Amstutz, HC ;
Ebramzadeh, E .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2004, 86A (09) :2015-2021
[5]   A STUDY OF IMPLANT FAILURE IN THE WAGNER RESURFACING ARTHROPLASTY [J].
BELL, RS ;
SCHATZKER, J ;
FORNASIER, VL ;
GOODMAN, SB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1985, 67A (08) :1165-1175
[6]   Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty [J].
Brodner, W ;
Grübl, A ;
Jankovsky, R ;
Meisinger, V ;
Lehr, S ;
Gottsauner-Wolf, F .
JOURNAL OF ARTHROPLASTY, 2004, 19 (08) :66-70
[7]   Metal-on-metal resurfacing of the hip in patients under the age of 55 years with osteoarthritis [J].
Daniel, J ;
Pynsent, PB ;
McMinn, DJW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (02) :177-184
[8]  
GOLDBERG V, 2005, INSTRUCTIONAL COURSE
[9]   UNDECALCIFIED PREPARATION OF BONE TISSUE - REPORT OF TECHNICAL EXPERIENCE AND DEVELOPMENT OF NEW METHODS [J].
HAHN, M ;
VOGEL, M ;
DELLING, G .
VIRCHOWS ARCHIV A-PATHOLOGICAL ANATOMY AND HISTOPATHOLOGY, 1991, 418 (01) :1-7
[10]  
HERBERTS P, 2004, ANN REP 2003