Optimal design of label-free silicon "lab on a chip" biosensors

被引:10
作者
Zhang, Yaping [1 ]
机构
[1] Univ Nottingham Ningbo, Fac Engn, Ningbo 315100, Zhejiang, Peoples R China
关键词
Mach-Zehnder interferometer; Young interferometer; Lab on a chip; Arrayed biosensors; Evanescent field detection; INTERFEROMETER IMMUNOSENSOR; CHALCOGENIDE GLASSES; WAVE-GUIDES; FABRICATION; TECHNOLOGY; ARRAY;
D O I
10.1016/j.pnsc.2013.09.002
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This paper reported the optimal design of label-free silicon on insulator (SOI "lab on a chip" biosensors. These devices are designed on the basis of the evanescent field detection principles and interferometer technologies. The well-established silicon device process technology can be applied to fabricate and test these biosensor devices. In addition, these devices can be monolithically integrated with CMOS electronics and microfluidics. For these biosensor devices, multi-mode interferometer (MMI) was employed to combine many stand-alone biosensors to form chip-level biosensor arrays, is enable real-time and label free monitoring and parallel detection of various analytes in multiple test samples. This sensing and detection technology features the highest detection sensitivity, which can detect analytes at extremely low concentrations instantaneously. This research can lead to innovative commercial development of the new generation of high sensitivity bioseasors for a wide range of applications in many fields, such as environmental monitoring, food security control, medical and biological applications. (C) 2013 Chinese Materials Research Society. Production and hosting by Elsevier B.V. All rights reserved.
引用
收藏
页码:481 / 486
页数:6
相关论文
共 17 条
  • [1] New trends in instrumental design for surface plasmon resonance-based biosensors
    Abbas, Abdennour
    Linman, Matthew J.
    Cheng, Quan
    [J]. BIOSENSORS & BIOELECTRONICS, 2011, 26 (05) : 1815 - 1824
  • [2] Generalized Mach-Zehnder interferometers for sensing applications
    Bernini, R
    Cusano, A
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2004, 100 (1-2) : 72 - 74
  • [3] Brandenburg A., 1997, SENSOR ACTUAT B-CHEM, V38-39, P266
  • [4] An integrated fluorescence array as a platform for lab-on-a-chip technology using multimode interference splitters
    Cleary, A
    García-Blanco, S
    Glidle, A
    Aitchison, JS
    Laybourn, P
    Cooper, JM
    [J]. IEEE SENSORS JOURNAL, 2005, 5 (06) : 1315 - 1320
  • [5] A silicon-on-insulator photonic wire based evanescent field sensor
    Densmore, A.
    Xu, D. X.
    Waldron, P.
    Janz, S.
    Cheben, P.
    Lapointe, J.
    Delage, A.
    Lamontagne, B.
    Schmid, J. H.
    Post, E.
    [J]. IEEE PHOTONICS TECHNOLOGY LETTERS, 2006, 18 (21-24) : 2520 - 2522
  • [6] Silicon photonic wire biosensor array for multiplexed real-time and label-free molecular detection
    Densmore, A.
    Vachon, M.
    Xu, D. -X.
    Janz, S.
    Ma, R.
    Li, Y. -H.
    Lopinski, G.
    Delage, A.
    Lapointe, J.
    Luebbert, C. C.
    Liu, Q. Y.
    Cheben, P.
    Schmid, J. H.
    [J]. OPTICS LETTERS, 2009, 34 (23) : 3598 - 3600
  • [7] Fan XD, 2011, NAT PHOTONICS, V5, P591, DOI [10.1038/NPHOTON.2011.206, 10.1038/nphoton.2011.206]
  • [8] Remote opto-chemical sensing with extreme sensitivity: design, fabrication and performance of a pigtailed integrated optical phase-modulated Mach-Zehnder interferometer system
    Heideman, RG
    Lambeck, PV
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 1999, 61 (1-3) : 100 - 127
  • [9] Planar waveguide-coupled, high-index-contrast, high-Q resonators in chalcogenide glass for sensing
    Hu, Juejun
    Carlie, Nathan
    Feng, Ning-Ning
    Petit, Laeticia
    Agarwal, Ann
    Richardson, Kathleen
    Kimerling, Lionel
    [J]. OPTICS LETTERS, 2008, 33 (21) : 2500 - 2502
  • [10] Low loss silicon waveguides for the mid-infrared
    Mashanovich, Goran Z.
    Milosevic, Milan M.
    Nedeljkovic, Milos
    Owens, Nathan
    Xiong, Boqian
    Teo, Ee Jin
    Hu, Youfang
    [J]. OPTICS EXPRESS, 2011, 19 (08): : 7112 - 7119