EXPERIMENTAL VALIDATION OF ALGORITHMS USED TO ESTIMATE EFFECTIVE DOSE DURING INTERVENTIONAL RADIOLOGY PROCEDURES

被引:1
作者
Negri, P. [1 ]
Campi, F. [2 ]
De Crescenzo, S. [3 ]
Garlati, L. [2 ]
Tambussi, O. [2 ]
机构
[1] CERAP SA, Octeville, France
[2] Politecn Milan, CeSNEF Nucl Engn, Milan, Italy
[3] ASST Niguarda, Fis Sanit, Milan, Italy
关键词
OCCUPATIONALLY EXPOSED PERSONS; DOUBLE DOSIMETRY ALGORITHMS; DIAGNOSTIC-RADIOLOGY; PROTECTIVE APRONS; ARTERIOGRAPHY;
D O I
10.1093/rpd/ncz136
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
In IR procedures, in order to evaluate the effective dose, the importance of the double dosemeter has been recognised, one worn above and one under the apron. Over the last few decades several algorithms have been developed to combine the readings of the dosemeters, however currently there is no international consensus on which is the best one. In this work, eight irradiations, corresponding to typical interventional radiology procedures, were carried out in order to experimentally verify the accuracy of the algorithms. The patient was substituted by solid water-equivalent (RW3) layers, while effective dose to personnel was calculated by TLDs inside the Alderson Rando phantom. The results show that most of the algorithms, with a few exceptions, are too conservative, however there are many factors which can affect their accuracy, so it is impossible to achieve a high level of precision in the evaluation of the effective dose.
引用
收藏
页码:42 / 49
页数:8
相关论文
共 23 条
[1]  
[Anonymous], ANN ICRP
[2]  
[Anonymous], 2002, 6 EUR ALARA NETW WOR
[3]  
Borasi G., 2001, FIS MED, V3, P221
[4]   Double-dosimetry algorithm for workers in interventional radiology [J].
Clerinx, Peter ;
Buls, N. ;
Bosmans, H. ;
de Mey, J. .
RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) :321-327
[5]   THE RELATIONSHIP OF EFFECTIVE DOSE TO PERSONNEL AND MONITOR READING FOR SIMULATED FLUOROSCOPIC IRRADIATION CONDITIONS [J].
FAULKNER, K ;
MARSHALL, NW .
HEALTH PHYSICS, 1993, 64 (05) :502-508
[6]  
Hubbel J. H., 2004, TABLE XRAY MASS ATTE
[7]  
International Commission on Radiobiological Protection, 2007, ICRP PUBL, V37
[8]  
International Commission on Radiological Protection, 2018, ANN ICRP, V47
[9]  
International Commission on Radiological Protection, 2002, ANN ICRP, V32
[10]   COMPARISON OF DOUBLE DOSIMETRY ALGORITHMS FOR ESTIMATING THE EFFECTIVE DOSE IN OCCUPATIONAL DOSIMETRY OF INTERVENTIONAL RADIOLOGY STAFF [J].
Jarvinen, H. ;
Buls, N. ;
Clerinx, P. ;
Miljanic, S. ;
Nikodemova, D. ;
Ranogajec-Komor, M. ;
Struelens, L. ;
d'Errico, F. .
RADIATION PROTECTION DOSIMETRY, 2008, 131 (01) :80-86