Investigation of polycapillary half lenses for quantitative confocal micro-X-ray fluorescence analysis

被引:6
作者
Iro, Michael [1 ]
Ingerle, Dieter [1 ]
Radtke, Martin [2 ]
Buzanich, Ana Guilherme [2 ]
Kregsamer, Peter [1 ]
Streli, Christina [1 ]
机构
[1] TU Wien, Atominst, Stadionallee 2, A-1020 Vienna, Austria
[2] Bundesanstalt Mat Forsch & Prufung BAM, Richard Willstatter Str 11, D-12489 Berlin, Germany
关键词
polycapillary optics; confocal micro-XRF; X-ray fluorescence; 3D elemental analysis; X-ray optics;
D O I
10.1107/S1600577522009699
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The use of polycapillary optics in confocal micro-X-ray fluorescence analysis (CMXRF) enables the destruction-free 3D investigation of the elemental composition of samples. The energy-dependent transmission properties, concerning intensity and spatial beam propagation of three polycapillary half lenses, which are vital for the quantitative interpretation of such CMXRF measurements, are investigated in a monochromatic confocal laboratory setup at the Atominstitut of TU Wien, and a synchrotron setup on the BAMline beamline at the BESSY II Synchrotron, Helmholtz-Zentrum-Berlin. The empirically established results, concerning the intensity of the transmitted beam, are compared with theoretical values calculated with the polycap software package and a newly presented analytical model for the transmission function. The resulting form of the newly modelled energy-dependent transmission function is shown to be in good agreement with Monte Carlo simulated results for the complete energy regime, as well as the empirically established results for the energy regime between 6 keV and 20 keV. An analysis of possible fabrication errors was conducted via pinhole scans showing only minor fabrication errors in two of the investigated polycapillary optics. The energy-dependent focal spot size of the primary polycapillary was investigated in the laboratory via the channel-wise evaluation of knife-edge scans. Experimental results are compared with data given by the manufacturer as well as geometric estimations for the minimal focal spot size. Again, the resulting measurement points show a trend in agreement with geometrically estimated results and manufacturer data.
引用
收藏
页码:1376 / 1384
页数:9
相关论文
共 25 条
  • [1] GXPS: A Geant4-based Monte Carlo code for x-ray polycapillary simulation
    Chi, Zhijun
    [J]. JOURNAL OF APPLIED PHYSICS, 2020, 128 (12)
  • [2] Confocal MXRF in environmental applications
    Fittschen, Ursula Elisabeth Adriane
    Falkenberg, Gerald
    [J]. ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (06) : 1743 - 1750
  • [3] Examination of poly-capillary lenses for their use in micro-XRF spectrometers
    Haschke, M
    Haller, M
    [J]. X-RAY SPECTROMETRY, 2003, 32 (03) : 239 - 247
  • [4] Haschke M., 2014, Laboratory Micro-X-Ray Fluorescence Spectroscopy: Instrumentation and Applications, DOI [10.1007/978-3-319-04864-2, DOI 10.1007/978-3-319-04864-2]
  • [5] A monochromatic confocal micro-x-ray fluorescence (μXRF) spectrometer for the lab
    Ingerle, D.
    Swies, J.
    Iro, M.
    Wobrauschek, P.
    Streli, C.
    Hradil, K.
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 2020, 91 (12)
  • [6] Three-dimensional micro-XRF investigations of paint layers with a tabletop setup
    Kanngiesser, B
    Malzer, W
    Rodriguez, AF
    Reiche, I
    [J]. SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (01) : 41 - 47
  • [7] Klockenkämper R, 2015, CHEM ANAL SERIES MON, V181
  • [8] Kumakhov MA, 2000, X-RAY SPECTROM, V29, P343, DOI 10.1002/1097-4539(200009/10)29:5<343::AID-XRS414>3.0.CO
  • [9] 2-S
  • [10] CHANNELING OF PHOTONS AND NEW X-RAY OPTICS
    KUMAKHOV, MA
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1990, 48 (1-4) : 283 - 286