Improvement of axial resolution of spectral domain optical coherence tomography with wide band PLC splitter

被引:0
作者
Eom, Joo Beom [1 ]
Min, Eun Jung [2 ]
Lee, Byeong Ha [2 ]
机构
[1] Korea Photon Technol Inst, 5 Cheomdan 4 Gil, Kwangju 500770, South Korea
[2] GIST, Sch Informat & Commun, Gwanju 500712, South Korea
来源
BIOPHOTONICS: PHOTONIC SOLUTIONS FOR BETTER HEALTH CARE III | 2012年 / 8427卷
关键词
Wide band optical coupler; Planar lightwave circuit (PLC); OCT;
D O I
10.1117/12.922157
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We report the development and performance of high resolution spectral domain optical coherence tomography (SD-OCT) system based on 2x2 planar lightwave circuit (PLC) splitter that was designed as a single mode splitter at near infrared and used as the beam splitter for a SD-OCT system. The splitter has been made by coupling SMFs to a planar lightwave circuit (PLC) splitter chip. The PLC splitter chip was fabricated to have a single mode property with 740 nm cutoff wavelength and the SMFs, which have 730 nm cutoff wavelength, were securely connected to the PLC chip through fiber block arrays having lithographically fabricated V grooves. With the implemented PLC splitter, we have obtained a low excess loss of 0.4 dB at 840 nm with wide band coupling property. With the proposed 2x2 PLC splitter and fabricated WDM coupler, SD-OCT images of samples successfully obtained by using combined source with 840 nm and 880 nm SLDs
引用
收藏
页数:6
相关论文
共 9 条
[1]   2x2 photonic crystal fiber splitter based on silica-based planar lightwave circuits [J].
Eom, Joo Beom ;
Park, Jong-Hyuk ;
Lee, Byeong Ha .
OPTICS LETTERS, 2009, 34 (23) :3737-3739
[2]   MEASUREMENT OF INTRAOCULAR DISTANCES BY BACKSCATTERING SPECTRAL INTERFEROMETRY [J].
FERCHER, AF ;
HITZENBERGER, CK ;
KAMP, G ;
ELZAIAT, SY .
OPTICS COMMUNICATIONS, 1995, 117 (1-2) :43-48
[3]   Optical coherence tomography for ultrahigh resolution in vivo imaging [J].
Fujimoto, JG .
NATURE BIOTECHNOLOGY, 2003, 21 (11) :1361-1367
[4]   OPTICAL COHERENCE TOMOGRAPHY [J].
HUANG, D ;
SWANSON, EA ;
LIN, CP ;
SCHUMAN, JS ;
STINSON, WG ;
CHANG, W ;
HEE, MR ;
FLOTTE, T ;
GREGORY, K ;
PULIAFITO, CA ;
FUJIMOTO, JG .
SCIENCE, 1991, 254 (5035) :1178-1181
[5]   Photonic crystal fiber coupler [J].
Lee, BH ;
Eom, LB ;
Kim, J ;
Moon, DS ;
Paek, UC ;
Yang, GH .
OPTICS LETTERS, 2002, 27 (10) :812-814
[6]   Full-range, high-speed, high-resolution 1-μm spectral-domain optical coherence tomography using BM-scan for volumetric imaging of the human posterior eye [J].
Makita, Shuichi ;
Fabritius, Tapio ;
Yasuno, Yoshiaki .
OPTICS EXPRESS, 2008, 16 (12) :8406-8420
[7]   Ultra-high resolution optical coherence tomography (OCT) using a halogen lamp as the light source [J].
Ohmi, M ;
Haruna, M .
OPTICAL REVIEW, 2003, 10 (05) :478-481
[8]   Advances in broad bandwidth light sources for ultrahigh resolution optical coherence tomography [J].
Unterhuber, A ;
Povazay, B ;
Bizheva, K ;
Hermann, B ;
Sattmann, H ;
Stingl, A ;
Le, T ;
Seefeld, M ;
Menzel, R ;
Preusser, M ;
Budka, H ;
Schubert, C ;
Reitsamer, H ;
Ahnelt, PK ;
Morgan, JE ;
Cowey, A ;
Drexler, W .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (07) :1235-1246
[9]   Ultrahigh-resolution optical coherence tomography by broadband continuum generation from a photonic crystal fiber [J].
Wang, YM ;
Zhao, YH ;
Nelson, JS ;
Chen, ZP ;
Windeler, RS .
OPTICS LETTERS, 2003, 28 (03) :182-184