摘要
A new cavity for pyrgeometer characterizations has been built at the Infrared Radiometry Section of the World Radiation Center (WRC-IRS) at PMOD/WRC. The calculated effective emissivity of 0.99993 +/- 0.00033 was obtained from Monte Carlo simulations taking into account the geometry and the measured temperature distributions of the cavity. The cavity is operated in a temperature range of -30 degrees C to +30 degrees C and is initially flushed with nitrogen to reduce the relative humidity in the cavity. The estimated uncertainties of retrieved pyrgeometer parameters k(1), k(2), and k(3) are +/- 0.024, +/- 0.0008, and +/- 0.03, respectively. The relative uncertainty of pyrgeometer sensitivity C is 0.8%. The comparison with the cavity used since 1995 at PMOD/WRC gave average differences of 0.005, 0.00026, and 0.08 for k(1), k(2), and k(3), respectively. The pyrgeometer sensitivity retrieved with the new cavity is on average 1.0% higher than with the original cavity. (C) 2008 Optical Society of America.
引用
收藏
页码:4441 / 4447
页数:7
相关论文
共 14 条
[1]  
ALBRECHT B, 1977, J APPL METEOROL, V16, P188, DOI 10.1175/1520-0450(1977)016<0190:PFIPP>2.0.CO
[2]  
2
[3]  
BETZ C, 1992, INTWICKLUNG EICHSTRA
[4]   CHARACTERIZATION OF AN ABSOLUTE CRYOGENIC RADIOMETER AS A STANDARD DETECTOR FOR RADIANT-POWER MEASUREMENTS [J].
DATLA, RU ;
STOCK, K ;
PARR, AC ;
HOYT, CC ;
MILLER, PJ ;
FOUKAL, PV .
APPLIED OPTICS, 1992, 31 (34) :7219-7225
[5]  
Fairall CW, 1998, J ATMOS OCEAN TECH, V15, P1229, DOI 10.1175/1520-0426(1998)015<1229:ANLACA>2.0.CO
[6]  
2
[7]  
Payne RE, 1999, J ATMOS OCEAN TECH, V16, P739, DOI 10.1175/1520-0426(1999)016<0739:ANLACA>2.0.CO
[8]  
2
[9]  
Philipona R, 1998, J ATMOS OCEAN TECH, V15, P687, DOI 10.1175/1520-0426(1998)015<0687:TBSRNP>2.0.CO
[10]  
2