Analytical performance of the plasma source RF quadrupole ion trap in elemental and isotopic MS

被引:30
作者
Eiden, GC [1 ]
Barinaga, CJ [1 ]
Koppenaal, DW [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
D O I
10.1039/a806433d
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Recent progress in the development of ion trap ICP-MS instrumentation is presented in terms of performance figures of merit for trace elemental and isotopic analysis. Conventional polyatomic ion interferences are avoided altogether using collisional dissociation techniques inherent to ion trap operation. The current detection limits are in the low pg mL(-1) range, as obtained over extended mass scans using multi-element solutions. The dynamic range linearity is nearly six orders of magnitude using variable injection times to control ion density in the trap. The abundance sensitivity is 10(5) to 10(6) using selective ion monitoring. Other performance figures approach or exceed those of conventional linear quadrupole based ICP-MS instrumentation.
引用
收藏
页码:1129 / 1132
页数:4
相关论文
共 23 条
[1]   REMOVAL OF ORGANIC-SOLVENTS BY CRYOGENIC DESOLVATION IN INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY [J].
ALVES, LC ;
MINNICH, MG ;
WIEDERIN, DR ;
HOUK, RS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1994, 9 (03) :399-403
[2]   EXTENDED THEORETICAL CONSIDERATIONS FOR MASS RESOLUTION IN THE RESONANCE EJECTION MODE OF QUADRUPOLE ION-TRAP MASS-SPECTROMETRY [J].
ARNOLD, NS ;
HARS, G ;
MEUZELAAR, HLC .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (07) :676-688
[3]   ION-TRAP MASS-SPECTROMETRY WITH AN INDUCTIVELY-COUPLED PLASMA SOURCE [J].
BARINAGA, CJ ;
KOPPENAAL, DW .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1994, 8 (01) :71-76
[4]  
DUCKWORTH DC, 1998, ICP INFORM NEWSL, V23, P298
[5]   Plasma source ion trap mass spectrometry: Enhanced abundance sensitivity by resonant ejection of atomic ions [J].
Eiden, GC ;
Barinaga, CJ ;
Koppenaal, DW .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1996, 7 (11) :1161-1171
[6]   Selective removal of plasma matrix ions in plasma source mass spectrometry [J].
Eiden, GC ;
Barinaga, CJ ;
Koppenaal, DW .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (04) :317-322
[7]  
Eiden GC, 1997, RAPID COMMUN MASS SP, V11, P37, DOI 10.1002/(SICI)1097-0231(19970115)11:1<37::AID-RCM815>3.0.CO
[8]  
2-C
[9]  
FRANKLIN TM, 1997, 45 ASMS C MASS SPECT, P1209
[10]  
FRUM CI, 1996, 44 ASMS C MASS SPECT, P831