A solid-state, open-system, differential calorimeter

被引:3
|
作者
Lacouture, Shelby [1 ]
Dardik, Trevor [1 ]
Van der Vliet, Dennis [2 ]
Akene, Jephtah [3 ]
Adeosun, Samuel [1 ]
Duncan, Robert V. [1 ]
机构
[1] Texas Tech Univ, Dept Phys & Astron, Lubbock, TX 79410 USA
[2] Honeywell UOP, 50E Algonquin Rd, Des Plaines, IL 60017 USA
[3] Southwest Res Inst, 6220 Culebra Rd, San Antonio, TX 78238 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2020年 / 91卷 / 09期
关键词
This article details the design; modeling; construction; and evaluation of an open system calorimeter that operates in a normal room environment to measure endothermic or exothermic events in a system subjected to a steady heat flux. The calorimeter is unique because it allows the measurement of energy and power from an open system; where a heat flux enters and leaves the calorimetric boundary in a well-controlled manner. It is also novel because it utilizes a solid state heating and cooling assembly that acts as an electronic heat reservoir. The system is capable of measuring power levels from a few milliwatts to several watts; and it has been designed and optimized to be nearly immune to variations at ambient temperature and room airflow. The calorimeter was modeled using lumped parameter electrical-thermal equivalent circuits in SPICE software. This modeling in the electrical domain led to the use of a mathematical correction factor that mitigates mismatches in thermal conduction paths between an active and a passive cell as well as correcting differences in the sensitivities of the flux sensors employed for heat flow measurement. After obtaining a viable design; a prototype was constructed and validated with precise input power delivered via electric joule heating of a resistive element. © 2020 American Institute of Physics Inc.. All rights reserved;
D O I
10.1063/5.0013591
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
This article details the design, modeling, construction, and evaluation of an open system calorimeter that operates in a normal room environment to measure endothermic or exothermic events in a system subjected to a steady heat flux. The calorimeter is unique because it allows the measurement of energy and power from an "open" system, where a heat flux enters and leaves the calorimetric boundary in a well-controlled manner. It is also novel because it utilizes a solid state heating and cooling assembly that acts as an electronic heat reservoir. The system is capable of measuring power levels from a few milliwatts to several watts, and it has been designed and optimized to be nearly immune to variations at ambient temperature and room airflow. The calorimeter was modeled using lumped parameter electrical-thermal equivalent circuits in SPICE software. This modeling in the electrical domain led to the use of a mathematical correction factor that mitigates mismatches in thermal conduction paths between an active and a passive cell as well as correcting differences in the sensitivities of the flux sensors employed for heat flow measurement. After obtaining a viable design, a prototype was constructed and validated with precise input power delivered via electric joule heating of a resistive element.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] INFLUENCE OF OPEN-POROUS SYSTEM ON THE SOLID-STATE PHASE TRANSITION IN 1-OCTADECENE
    Alekseev, O. M.
    Alekseev, S. O.
    Zabashta, Yu F.
    Lazarenko, M. M.
    Hnatiuk, K., I
    Lazarenko, M., V
    Dinzhos, R., V
    Simeonov, M. S.
    UKRAINIAN JOURNAL OF PHYSICS, 2019, 64 (04): : 340 - 347
  • [42] A COMPARISON OF CLOSED-SYSTEM AND OPEN-SYSTEM MODELS FOR POREWATER PH AND CALCITE-SATURATION STATE
    BOUDREAU, BP
    CANFIELD, DE
    GEOCHIMICA ET COSMOCHIMICA ACTA, 1993, 57 (02) : 317 - 334
  • [43] Open-system dynamics and mixing in magma mushes
    G. W. Bergantz
    J. M. Schleicher
    A. Burgisser
    Nature Geoscience, 2015, 8 : 793 - 796
  • [44] Open-System Dynamics of the Geometrical Reperesentation for a Qutrit
    Cheng, Shuming
    Liu, Lijun
    Li, Li
    Xi, Zairong
    2015 34TH CHINESE CONTROL CONFERENCE (CCC), 2015, : 8319 - 8324
  • [45] TRANSMITTER SYSTEM COMMANDED BY SOLID-STATE AMPS
    BROWNE, J
    MICROWAVES & RF, 1989, 28 (04) : 125 - &
  • [46] SOLID-STATE KEY LOCK SECURITY SYSTEM
    FRAGANO, M
    CROSETTO, P
    IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 1984, 30 (04) : 604 - 607
  • [47] Chaos in a multimode solid-state laser system
    Bracikowski, C.
    Roy, Rajarshi
    CHAOS, 1991, 1 (01) : 49 - 64
  • [48] CELLULASE PRODUCTION BY A SOLID-STATE CULTURE SYSTEM
    RYU, DDY
    TAGUCHI, H
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1983, 186 (AUG): : 84 - MBTD
  • [49] MODULAR SOLID-STATE BEHAVIORAL TRAINING SYSTEM
    SMITH, GR
    WASSON, SL
    BEHAVIOR RESEARCH METHODS & INSTRUMENTATION, 1974, 6 (05): : 481 - 487
  • [50] DEVELOPMENT OF A SOLID-STATE SOUND RECORDER SYSTEM
    OKUZAWA, K
    ONISHI, N
    NISHIDA, H
    HIRAI, H
    SHINOHARA, M
    INOUE, T
    TANAKA, H
    IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 1992, 38 (03) : 364 - 370