Visual Strehl Performance of IOL Designs with Extended Depth of Focus

被引:27
作者
Arias Gallego, Augusto [1 ,2 ]
Bara, Salvador [3 ]
Jaroszewicz, Zbigniew [4 ]
Kolodziejczyk, Andrzej [5 ]
机构
[1] Inst Tecnol Metropolitano, Res Ctr, Medellin 54954, Colombia
[2] Univ Nacl Colombia, Sch Phys, Medellin, Colombia
[3] Univ Santiago Compostela, EU Opt & Optometria, Area Opt, Galicia, Spain
[4] Inst Natl Telecommun, Inst Appl Opt, Warsaw, Poland
[5] Warsaw Univ Technol, Fac Phys, Warsaw, Poland
关键词
vision; presbyopia; intraocular lens; extended depth of focus; INTRAOCULAR-LENS IMPLANTATION; DIFFRACTIVE ELEMENTS; CATARACT-SURGERY; FOCAL DEPTH; QUALITY; ABERRATION; COMPENSATION; PRESBYOPIA; CONTRAST; METRICS;
D O I
10.1097/OPX.0b013e3182775e1a
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. The use of monofocal, bifocal, or multifocal intraocular lenses (IOLs) is a common option to restore the refractive power of the eye in aphakic patients after cataract surgery. In this article, we study the predicted performance of two new designs of IOL, both with extended depth of focus: the quartic axicon and the light sword optical element (LSOE). These elements provide continuous focal segments spanning the range of dioptric power required for presbyopia compensation. Methods. The performance analysis is based on the visual Strehl ratio (SR) computed in the spatial frequency domain, a neuro-optical quality metric that takes into account the effects of both the optical transfer function and the neural contrast sensitivity function. Furthermore, the classical SR and compound modulation transfer function were calculated. Some conventional transmittances of commercially available IOLs are also analyzed. Results. The LSOE design has a more homogeneous behavior than other available solutions, providing a more uniform image quality over a significant fraction of the required addition range. Conclusions. The angular average of the visual SR computed in the spatial frequency domain and compound modulation transfer function indicate that both designs of extended depth of focus elements provide better optical quality on the whole addition range, except at the end points, than the discrete focus designs. The LSOE performed better than the quartic axicon in terms of uniformity of image quality within the same range. (Optom Vis Sci 2012; 89: 1702-1707)
引用
收藏
页码:1702 / 1707
页数:6
相关论文
共 25 条
  • [1] Abbott Medial Optics Inc. (AMO), TECN ASPH IOL
  • [2] Presbyopia compensation with a quartic axicon
    Ares, J
    Flores, R
    Bara, S
    Jaroszewicz, Z
    [J]. OPTOMETRY AND VISION SCIENCE, 2005, 82 (12) : 1071 - 1078
  • [3] AMO array multifocal lens versus monofocal correction in cataract surgery
    Brydon, KW
    Tokarewicz, AC
    Nichols, BD
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2000, 26 (01) : 96 - 100
  • [4] OPTICAL AND RETINAL FACTORS AFFECTING VISUAL RESOLUTION
    CAMPBELL, FW
    GREEN, DG
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1965, 181 (03): : 576 - +
  • [5] Predicting subjective judgment of best focus with objective image quality metrics
    Cheng, X
    Bradley, A
    Thibos, LN
    [J]. JOURNAL OF VISION, 2004, 4 (04): : 310 - 321
  • [6] Optical surface optimization for the correction of presbyopia
    Dai, Guang-Ming
    [J]. APPLIED OPTICS, 2006, 45 (17) : 4184 - 4195
  • [7] Quality of vision after cataract surgery after Tecnis Z9000 intraocular lens implantation - Effect of contrast sensitivity and wavefront aberration improvements on the quality of daily vision
    Denoyer, Alexandre
    Le Lez, Marie-Laure
    Majzoub, Samuel
    Pisella, Pierre-Jean
    [J]. JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2007, 33 (02) : 210 - 216
  • [8] Imaging with extended focal depth by means of the refractive light sword optical element
    Garcia, J. Ares
    Bara, S.
    Gomez Garcia, M.
    Jaroszewicz, Z.
    Kolodziejczyk, A.
    Petelczyc, K.
    [J]. OPTICS EXPRESS, 2008, 16 (22): : 18371 - 18378
  • [9] Goodman JW, 1968, INTRODUCTION TO FOUR
  • [10] Henry JM, 2001, B SOC BELG OPHTHALM, V282, P33