Analysis of Stereotactic Accuracy in Patients Undergoing Deep Brain Stimulation Using Nexframe and the Leksell Frame

被引:53
作者
Bot, Maarten [1 ,4 ]
van den Munckhof, Pepijn [4 ]
Bakay, Roy [2 ]
Sierens, Diane [3 ]
Stebbins, Glenn [1 ]
Metman, Leo Verhagen [1 ]
机构
[1] Rush Univ, Med Ctr, Dept Neurol Sci, Chicago, IL 60612 USA
[2] Rush Univ, Med Ctr, Dept Neurosurg, Chicago, IL 60612 USA
[3] Rockford Mem Hosp, Dept Neurosurg, Rockford, IL USA
[4] Univ Amsterdam, Acad Med Ctr, Dept Neurosurg, NL-1105 AZ Amsterdam, Netherlands
关键词
Frameless stereotaxy; Frame-based stereotaxy; Deep brain stimulation; Stereotactic neurosurgery; Movement disorders; SUBTHALAMIC NUCLEUS; PARKINSONS-DISEASE; COMPUTED-TOMOGRAPHY; X-RAY; ELECTRODES; SURGERY; SHIFT; IMPLANTATION; PLACEMENT; DISTORTION;
D O I
10.1159/000375178
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Objective: To determine and compare the accuracy of Nexframe and the Leksell stereotactic frame in deep brain stimulation (DBS) procedures. Background: The 'frameless' Nexframe uses bone fiducials rather than a head-mounted frame, which offers potential benefits for both the patient and the surgical team. Accuracy of lead implantation and factors affecting this accuracy are of crucial importance but have not been extensively studied for the frameless system. Design/Methods: The location of 194 (Leksell frame, n = 116; Nexframe, n = 78) DBS leads was determined on postoperative MRI. Obtained stereotactic coordinates were compared with expected intraoperative target coordinates. Resulting absolute errors in the X (medial-lateral), Y (anteri-or-posterior), and Z (dorsal-ventral) coordinates (Delta X, Delta Y, and Delta Z) were then used to calculate the vector error (VE). The vector error describes the total error in 3-D space and represents our main outcome measure. Results: The vector error mean +/- SD) was 2.8 +/- 1.3 for Nexframe and 2.5 +/- 1.2 for the Leksell frame (p = 0.43). For Nexframe, absolute X, Y, and Z errors were 1.4 +/- 1.3, 1.7 +/- 1.2, and 1.0 +/- 0.9 mm. For the Leksell frame, the absolute X, Y, and Z errors were 1.4 +/- 1.0, 1.2 +/- 1.0, and 1.3 +/- 0.9 mm. On the anterior-posterior plane (Y coordinate), the Leksell frame was more accurate than Nexframe (p = 0.04). In contrast, Nexframe was more accurate on the dorsal-ventral plane (Z coordinate) (p = 0.04). There was no difference in accuracy between the two methods on the medial-lateral plane (X coordinate). Conclusion: This comparison of Nexframe and the Leksell frame shows that both techniques have equivalent overall 3-D accuracy. (C) 2015 S. Karger AG, Basel
引用
收藏
页码:316 / 325
页数:10
相关论文
共 30 条
[1]   Comparison of accuracy and precision between frame-based and frameless stereotactic navigation for deep brain stimulation electrode implantation [J].
Bjartmarz, Hjalmar ;
Rehncrona, Stig .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2007, 85 (05) :235-242
[2]   Clinical Motor Outcome of Bilateral Subthalamic Nucleus Deep-Brain Stimulation for Parkinson's Disease Using Image-Guided Frameless Stereotaxy [J].
Bronte-Stewart, Helen ;
Louie, Stephanie ;
Batya, Sara ;
Henderson, Jaimie M. .
NEUROSURGERY, 2010, 67 (04) :1088-1093
[3]  
CLARKE RH, 1906, BRIT MED J, V2, P1799
[4]   Cortical and subcortical brain shift during stereotactic procedures [J].
Elias, W. Jeffrey ;
Fu, Kai-Ming ;
Frysinger, Robert C. .
JOURNAL OF NEUROSURGERY, 2007, 107 (05) :983-988
[5]   Accuracy of stereotactic electrode placement in deep brain stimulation by intraoperative computed tomography [J].
Fiegele, Thomas ;
Feuchtner, Gudrun ;
Sohm, Florian ;
Bauer, Richard ;
Anton, Juergen Volker ;
Gotwald, Thaddaeus ;
Twerdy, Klaus ;
Eisner, Wilhelm .
PARKINSONISM & RELATED DISORDERS, 2008, 14 (08) :595-599
[6]   Nexframe Frameless Stereotaxy with Multitract Microrecording: Accuracy Evaluated by Frame-Based Stereotactic X-Ray [J].
Fukaya, Chikashi ;
Sumi, Koichiro ;
Otaka, Toshiharu ;
Obuchi, Toshiki ;
Kano, Toshikazu ;
Kobayashi, Kazutaka ;
Oshima, Hideki ;
Yamamoto, Takamitsu ;
Katayama, Yoichi .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2010, 88 (03) :163-168
[7]   Brain shift during deep brain stimulation surgery for Parkinson's disease [J].
Halpern, Casey H. ;
Danish, Shabbar F. ;
Baltuch, Gordon H. ;
Jaggi, Jurg L. .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2008, 86 (01) :37-43
[8]   Frameless stereotaxy using bone fiducial markers for deep brain stimulation [J].
Holloway, KL ;
Gaede, SE ;
Starr, PA ;
Rosenow, JM ;
Ramakrishnan, V ;
Henderson, JM .
JOURNAL OF NEUROSURGERY, 2005, 103 (03) :404-413
[9]   Analysis of Stereotactic Accuracy of the Cosman-Robert-Wells Frame and Nexframe Frameless Systems in Deep Brain Stimulation Surgery [J].
Kelman, Craig ;
Ramakrishnan, V. ;
Davies, Alex ;
Holloway, Kathryn .
STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2010, 88 (05) :288-295
[10]   Is MRI a reliable tool to locate the electrode after deep brain stimulation surgery? Comparison study of CT and MRI for the localization of electrodes after DBS [J].
Lee, Ji Yeoun ;
Kim, Jin Wook ;
Lee, Jee-Young ;
Lim, Yong Hoon ;
Kim, Cheolyoung ;
Kim, Dong Gyu ;
Jeon, Beom Seok ;
Paek, Sun Ha .