Quantification with a dedicated breast PET/CT scanner

被引:11
作者
Bowen, Spencer L. [1 ]
Ferrero, Andrea [2 ]
Badawi, Ramsey D. [3 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA 02129 USA
[2] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[3] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
基金
美国国家卫生研究院;
关键词
PET/CT; positron emission mammography; dedicated; breast cancer; POSITRON EMISSION MAMMOGRAPHY; COMPONENT-BASED NORMALIZATION; 3D PET; ATTENUATION CORRECTION; IMAGING-SYSTEM; COMPUTED-TOMOGRAPHY; DETECTOR EFFICIENCY; SCATTER CORRECTION; COMPTON SCATTER; CANCER;
D O I
10.1118/1.3703593
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dedicated breast PET/CT is expected to have utility in local staging, surgical planning, monitoring of therapy response, and detection of residual disease for breast cancer. Quantitative metrics will be integral to several such applications. The authors present a validation of fully 3D data correction schemes for a custom built dedicated breast PET/CT (DbPET/CT) scanner via F-18-FDG phantom scans. Methods: A component-based normalization was implemented, live-time was estimated with a multicomponent model, and a variance reduced randoms estimate was computed from delayed coincidences. Attenuation factors were calculated by using a CT based segmentation scheme while scatter was computed using a Monte Carlo (MC) simulation method. As no performance standard currently exists for breast PET systems, custom performance tests were created based on prior patient imaging results. Count-rate linearity for live-time and randoms corrections was measured with a decay experiment for a solid polyethylene cylinder phantom with an offset line source. A MC simulation was used to validate attenuation correction, a multicompartment phantom with asymmetric activity distribution provided an assessment of scatter correction, and image uniformity after geometric and detector normalization was measured from a high count scan of a uniform cylinder phantom. Raw data were reconstructed with filtered back projection (FBP) after Fourier rebinning. To quantify performance absolute activity concentrations, contrast recovery coefficients and image uniformity were calculated through region of interest analysis. Results: The most significant source of error was attributed to mispositioning of events due to pileup, presenting in count-related axial and transaxial nonuniformities that were not corrected for with the normalization method used here. Within the range of singles counts observed during clinical trials residual error after applying all corrections was comparable to that of a commercial whole body PET/CT system. Conclusions: The results suggest that DbPET/CT is capable of producing quantitative images under the operating conditions expected during patient imaging. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3703593]
引用
收藏
页码:2694 / 2707
页数:14
相关论文
共 50 条
[41]   A Combined PET/CT Scanner for Clinical Oncology [J].
Beyer, Thomas ;
Townsend, David W. ;
Brun, Tony ;
Kinahan, Paul E. ;
Charron, Martin ;
Roddy, Raymond ;
Jerin, Jeff ;
Young, John ;
Byars, Larry ;
Nutt, Ronald .
JOURNAL OF NUCLEAR MEDICINE, 2020, 61 :168S-177S
[42]   Dedicated Breast CT: Getting Ready for Prime Time [J].
Aminololama-Shakeri, Shadi ;
Boone, John M. .
JOURNAL OF BREAST IMAGING, 2024, 6 (05) :465-475
[43]   Is dedicated chest CT needed in addition to PET/CT for evaluation of pediatric oncology patients? [J].
Goodman, Sarah Z. ;
Rosenblum, Jeremy ;
Tuna, Ibrahim ;
Levsky, Jeffrey M. ;
Ricafort, Rosanna ;
Taragin, Benjamin .
CLINICAL IMAGING, 2015, 39 (05) :794-798
[44]   PET/CT in breast cancer: an update [J].
Groheux, D. ;
Moretti, J. -L. ;
Giacchetti, S. ;
Hindie, E. ;
Teyton, P. ;
Cuvier, C. ;
Bousquet, G. ;
Misset, J. -L. ;
Boin, C. ;
Espie, M. .
BULLETIN DU CANCER, 2009, 96 (11) :1053-1070
[45]   PET, PET/ CT, and PET/ MR Imaging Assessment of Breast Cancer [J].
Shamim, Shamim Ahmed ;
Torigian, Drew A. ;
Kumar, Rakesh .
PET CLINICS, 2008, 3 (03) :381-+
[46]   In-line spectroscopy for iodine quantification in dynamic contrast-enhanced dedicated breast CT [J].
Goris, Liselot C. ;
Mikerov, Mikhail ;
Pautasso, Juan J. ;
Sechopoulos, Ioannis .
MEDICAL PHYSICS, 2025, 52 (02) :1037-1044
[47]   Assessment of x-ray Scatter for the Micro-CT Subsystem of the FLEX Triumph™ Preclinical PET-CT Scanner [J].
Gutierrez, Daniel ;
Zaidi, Habib .
2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, :2216-2223
[48]   FDG-PET and PET/CT in breast cancer staging [J].
Avril, Norbert ;
Mather, Stephen J. ;
Roylance, Rebecca .
BREAST CARE, 2007, 2 (06) :372-377
[49]   Comparative image quality and dosimetric performance of two generations of dedicated breast CT systems [J].
Pautasso, Juan J. ;
Van Speybroeck, Camille D. E. ;
Michielsen, Koen ;
Sechopoulos, Ioannis .
MEDICAL PHYSICS, 2025, 52 (04) :2191-2200
[50]   Clinical practice guideline for dedicated breast PET [J].
Makoto Hosono ;
Tsuneo Saga ;
Kengo Ito ;
Shinichiro Kumita ;
Masayuki Sasaki ;
Michio Senda ;
Jun Hatazawa ;
Hiroshi Watanabe ;
Hiroshi Ito ;
Shinichi Kanaya ;
Yuichi Kimura ;
Hideo Saji ;
Seishi Jinnouchi ;
Hiroyoshi Fukukita ;
Koji Murakami ;
Seigo Kinuya ;
Junichi Yamazaki ;
Mayuki Uchiyama ;
Koichi Uno ;
Katsuhiko Kato ;
Tsuyoshi Kawano ;
Kazuo Kubota ;
Takashi Togawa ;
Norinari Honda ;
Hirotaka Maruno ;
Mana Yoshimura ;
Masami Kawamoto ;
Yukihiko Ozawa .
Annals of Nuclear Medicine, 2014, 28 :597-602