Quantification with a dedicated breast PET/CT scanner

被引:11
作者
Bowen, Spencer L. [1 ]
Ferrero, Andrea [2 ]
Badawi, Ramsey D. [3 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA 02129 USA
[2] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[3] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
基金
美国国家卫生研究院;
关键词
PET/CT; positron emission mammography; dedicated; breast cancer; POSITRON EMISSION MAMMOGRAPHY; COMPONENT-BASED NORMALIZATION; 3D PET; ATTENUATION CORRECTION; IMAGING-SYSTEM; COMPUTED-TOMOGRAPHY; DETECTOR EFFICIENCY; SCATTER CORRECTION; COMPTON SCATTER; CANCER;
D O I
10.1118/1.3703593
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dedicated breast PET/CT is expected to have utility in local staging, surgical planning, monitoring of therapy response, and detection of residual disease for breast cancer. Quantitative metrics will be integral to several such applications. The authors present a validation of fully 3D data correction schemes for a custom built dedicated breast PET/CT (DbPET/CT) scanner via F-18-FDG phantom scans. Methods: A component-based normalization was implemented, live-time was estimated with a multicomponent model, and a variance reduced randoms estimate was computed from delayed coincidences. Attenuation factors were calculated by using a CT based segmentation scheme while scatter was computed using a Monte Carlo (MC) simulation method. As no performance standard currently exists for breast PET systems, custom performance tests were created based on prior patient imaging results. Count-rate linearity for live-time and randoms corrections was measured with a decay experiment for a solid polyethylene cylinder phantom with an offset line source. A MC simulation was used to validate attenuation correction, a multicompartment phantom with asymmetric activity distribution provided an assessment of scatter correction, and image uniformity after geometric and detector normalization was measured from a high count scan of a uniform cylinder phantom. Raw data were reconstructed with filtered back projection (FBP) after Fourier rebinning. To quantify performance absolute activity concentrations, contrast recovery coefficients and image uniformity were calculated through region of interest analysis. Results: The most significant source of error was attributed to mispositioning of events due to pileup, presenting in count-related axial and transaxial nonuniformities that were not corrected for with the normalization method used here. Within the range of singles counts observed during clinical trials residual error after applying all corrections was comparable to that of a commercial whole body PET/CT system. Conclusions: The results suggest that DbPET/CT is capable of producing quantitative images under the operating conditions expected during patient imaging. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3703593]
引用
收藏
页码:2694 / 2707
页数:14
相关论文
共 50 条
[31]   Comparison of Nongated Chest CT and Dedicated Calcium Scoring CT for Coronary Calcium Quantification Using a 256-Dector Row CT Scanner [J].
Chen, Yuhuan ;
Hu, Zhijun ;
Li, Michelle ;
Jia, Yongjun ;
He, Taiping ;
Liu, Zhentang ;
Wei, Donghong ;
Yu, Yong .
ACADEMIC RADIOLOGY, 2019, 26 (10) :E267-E274
[32]   Dedicated Breast CT: Screening Technique of the Future [J].
Aminololama-Shakeri S. ;
Hargreaves J.B. ;
Boone J.M. ;
Lindfors K.K. .
Current Breast Cancer Reports, 2016, 8 (4) :242-247
[33]   Higher breast cancer conspicuity on dbPET compared to WB-PET/CT [J].
Nishimatsu, Kayo ;
Nakamoto, Yuji ;
Miyake, Kanae K. ;
Ishimori, Takayoshi ;
Kanao, Shotaro ;
Toi, Masakazu ;
Togashi, Kaori .
EUROPEAN JOURNAL OF RADIOLOGY, 2017, 90 :138-145
[34]   PET/CT and breast cancer [J].
Barbara Zangheri ;
Cristina Messa ;
Maria Picchio ;
Luigi Gianolli ;
Claudio Landoni ;
Ferruccio Fazio .
European Journal of Nuclear Medicine and Molecular Imaging, 2004, 31 :S135-S142
[35]   PET/CT and breast cancer [J].
Zangheri, B ;
Messa, C ;
Picchio, M ;
Gianolli, L ;
Landoni, C ;
Fazio, F .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (Suppl 1) :S135-S142
[36]   Distribution pattern of FDG uptake using ring-type dedicated breast PET in comparison to whole-body PET/CT scanning in invasive breast cancer [J].
Rena Sakaguchi ;
Masako Kataoka ;
Shotaro Kanao ;
Kanae K. Miyake ;
Yuji Nakamoto ;
Tomohatru Sugie ;
Masakazu Toi ;
Yoshiki Mikami ;
Kaori Togashi .
Annals of Nuclear Medicine, 2019, 33 :570-578
[37]   Performance Evaluation of a New Dedicated Breast PET Scanner Using NEMA NU4-2008 Standards [J].
Miyake, Kanae K. ;
Matsumoto, Keiichi ;
Inoue, Mika ;
Nakamoto, Yuji ;
Kanao, Shotaro ;
Oishi, Tae ;
Kawase, Shigeto ;
Kitamura, Keishi ;
Yamakawa, Yoshiyuki ;
Akazawa, Ayako ;
Kobayashi, Tetsuya ;
Ohi, Junichi ;
Togashi, Kaori .
JOURNAL OF NUCLEAR MEDICINE, 2014, 55 (07) :1198-1203
[38]   CT-Based Attenuation Correction on the FLEX Triumph Preclinical PET/CT Scanner [J].
Prasad, Rameshwar ;
Ay, Mohammad R. ;
Ratib, Osman ;
Zaidi, Habib .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (01) :66-75
[39]   Evaluation of a Hanging-Breast PET System for Primary Tumor Visualization in Patients With Stage I-III Breast Cancer: Comparison With Standard PET/CT [J].
Teixeira, Suzana C. ;
Rebolleda, Jose Ferrer ;
Koolen, Bas B. ;
Wesseling, Jelle ;
Jurado, Raul Sanchez ;
Stokkel, Marcel P. M. ;
Santiago, Maria del Puig Cozar ;
van der Noort, Vincent ;
Rutgers, Emiel J. Th. ;
Olmos, Renato A. Valdes .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2016, 206 (06) :1307-1314
[40]   Quantification of PET and CT Data Misalignment Errors in Cardiac PET/CT: Clinical and Phantom Studies [J].
Ghafarian, P. ;
Aghamiri, S. M. R. ;
Ay, M. R. ;
Rahmim, A. ;
Zaidi, H. .
2009 3RD INTERNATIONAL CONFERENCE ON BIOINFORMATICS AND BIOMEDICAL ENGINEERING, VOLS 1-11, 2009, :2263-+