Quantification with a dedicated breast PET/CT scanner

被引:11
作者
Bowen, Spencer L. [1 ]
Ferrero, Andrea [2 ]
Badawi, Ramsey D. [3 ]
机构
[1] Massachusetts Gen Hosp, Dept Radiol, Athinoula A Martinos Ctr Biomed Imaging, Charlestown, MA 02129 USA
[2] Univ Calif Davis, Dept Biomed Engn, Davis, CA 95616 USA
[3] Univ Calif Davis, Med Ctr, Dept Radiol, Sacramento, CA 95817 USA
基金
美国国家卫生研究院;
关键词
PET/CT; positron emission mammography; dedicated; breast cancer; POSITRON EMISSION MAMMOGRAPHY; COMPONENT-BASED NORMALIZATION; 3D PET; ATTENUATION CORRECTION; IMAGING-SYSTEM; COMPUTED-TOMOGRAPHY; DETECTOR EFFICIENCY; SCATTER CORRECTION; COMPTON SCATTER; CANCER;
D O I
10.1118/1.3703593
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Dedicated breast PET/CT is expected to have utility in local staging, surgical planning, monitoring of therapy response, and detection of residual disease for breast cancer. Quantitative metrics will be integral to several such applications. The authors present a validation of fully 3D data correction schemes for a custom built dedicated breast PET/CT (DbPET/CT) scanner via F-18-FDG phantom scans. Methods: A component-based normalization was implemented, live-time was estimated with a multicomponent model, and a variance reduced randoms estimate was computed from delayed coincidences. Attenuation factors were calculated by using a CT based segmentation scheme while scatter was computed using a Monte Carlo (MC) simulation method. As no performance standard currently exists for breast PET systems, custom performance tests were created based on prior patient imaging results. Count-rate linearity for live-time and randoms corrections was measured with a decay experiment for a solid polyethylene cylinder phantom with an offset line source. A MC simulation was used to validate attenuation correction, a multicompartment phantom with asymmetric activity distribution provided an assessment of scatter correction, and image uniformity after geometric and detector normalization was measured from a high count scan of a uniform cylinder phantom. Raw data were reconstructed with filtered back projection (FBP) after Fourier rebinning. To quantify performance absolute activity concentrations, contrast recovery coefficients and image uniformity were calculated through region of interest analysis. Results: The most significant source of error was attributed to mispositioning of events due to pileup, presenting in count-related axial and transaxial nonuniformities that were not corrected for with the normalization method used here. Within the range of singles counts observed during clinical trials residual error after applying all corrections was comparable to that of a commercial whole body PET/CT system. Conclusions: The results suggest that DbPET/CT is capable of producing quantitative images under the operating conditions expected during patient imaging. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.3703593]
引用
收藏
页码:2694 / 2707
页数:14
相关论文
共 50 条
  • [21] The Role of PET/CT in Breast Cancer
    Hadebe, Bawinile
    Harry, Lerwine
    Ebrahim, Tasmeera
    Pillay, Venesen
    Vorster, Mariza
    DIAGNOSTICS, 2023, 13 (04)
  • [22] In-line Phase Contrast Tomography of the Breast with a Dedicated Micro-CT Scanner
    Mettivier, Giovanni
    Bliznakova, Kristina
    Sarno, Antonio
    Di Lillo, Francesca
    Castriconi, Roberta
    Russo, Paolo
    2016 IEEE NUCLEAR SCIENCE SYMPOSIUM, MEDICAL IMAGING CONFERENCE AND ROOM-TEMPERATURE SEMICONDUCTOR DETECTOR WORKSHOP (NSS/MIC/RTSD), 2016,
  • [23] Dedicated scanner for laboratory investigations on cone-beam CT/SPECT imaging of the breast
    Mettivier, Giovanni
    Russo, Paolo
    Cesarelli, Mario
    Ospizio, Roberto
    Passeggio, Giuseppe
    Roscilli, Lorenzo
    Pontoriere, Giuseppe
    Rocco, Raffaele
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 629 (01) : 350 - 356
  • [24] Spectral optimization for dedicated breast CT
    Weigel, Michaela
    Vollmar, Sabrina V.
    Kalender, Willi A.
    MEDICAL PHYSICS, 2011, 38 (01) : 114 - 124
  • [25] A Proposed Dedicated Breast PET Lexicon: Standardization of Description and Reporting of Radiotracer Uptake in the Breast
    Miyake, Kanae K.
    Kataoka, Masako
    Ishimori, Takayoshi
    Matsumoto, Yoshiaki
    Torii, Masae
    Takada, Masahiro
    Satoh, Yoko
    Kubota, Kazunori
    Satake, Hiroko
    Yakami, Masahiro
    Isoda, Hiroyoshi
    Ikeda, Debra M.
    Toi, Masakazu
    Nakamoto, Yuji
    DIAGNOSTICS, 2021, 11 (07)
  • [26] Dedicated Breast Gamma Camera Imaging and Breast PET: Current Status and Future Directions
    Narayanan, Deepa
    Berg, Wendie A.
    PET CLINICS, 2018, 13 (03) : 363 - +
  • [27] Preliminary performance characterization of DbPET2.1, a PET scanner dedicated to the imaging of the breast and extremities
    Ferrero, Andrea
    Peng, Qiyu
    Burkett, George W., Jr.
    Sumanasena, Buddika
    Moses, William W.
    Badawi, Ramsey D.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2015, 1 (01):
  • [28] First clinical experience with a dedicated PET for hanging breast molecular imaging
    Koolen, B. B.
    Aukema, T. S.
    Gonzalez Martinez, A. J.
    Vogel, W. V.
    Caballero Ontanaya, L.
    Peeters, M. -J. T. F. D. Vrancken
    Vroonland, C. J. J.
    Rutgers, E. J. Th
    Benlloch Baviera, J. M.
    Olmos, R. A. Valdes
    QUARTERLY JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 57 (01) : 92 - 100
  • [29] Investigation of a Dedicated, High Resolution PET/CT Scanner for Staging and Treatment Planning of Head and Neck Cancer
    Raylman, Raymond R.
    Stolin, Alexander V.
    Sompalli, Prashanth
    Randall, Nicole Bunda
    Martone, Peter F.
    Clinthorne, Neal H.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2015, 62 (05) : 2057 - 2066
  • [30] Initial clinical test of a breast-PET scanner
    Raylman, Raymond R.
    Abraham, Jame
    Hazard, Hannah
    Koren, Courtney
    Filburn, Shannon
    Schreiman, Judith S.
    Kurian, Sobha
    Majewski, Stan
    Marano, Gary D.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2011, 55 (01) : 58 - 64