Technical Note: Development of 3D-printed breast phantoms for end-to-end testing of whole breast volumetric arc radiotherapy

被引:4
作者
Delombaerde, Laurence [1 ]
Petillion, Saskia [2 ]
Weltens, Caroline [1 ,2 ]
De Roover, Robin [1 ]
Reynders, Truus [2 ]
Depuydt, Tom [1 ,2 ]
机构
[1] Katholieke Univ Leuven, Dept Oncol, Leuven, Belgium
[2] Univ Hosp Leuven, Dept Radiat Oncol, Leuven, Belgium
关键词
3D-printing; anthropomorphic phantom; breast VMAT-SIB; commissioning; dose verification; DOSIMETRY;
D O I
10.1002/acm2.12976
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
End-to-end testing of a new breast radiotherapy technique preferably requires realistic phantom geometries, which is challenging to achieve using currently commercially available solutions. We have developed a series of three-dimensional (3D)-printed breast phantoms, with ionization chamber and radiochromic film inserts, which can be attached to a commercial anthropomorphic thorax phantom. A contoured left breast from a patient's planning CT was mapped onto a CT of the CIRS E2E thorax phantom (CIRS Inc.) and cropped to fit the surface. Four versions of the breast were 3D printed, containing a cavity for an ionization chamber and slits for radiochromic film insertion in the three cardinal planes, respectively. The phantoms were fully compatible with surface scanning technology used for setup. The phantoms were validated using a whole-breast volumetric modulated arc therapy protocol with a simultaneous integrated boost to the tumor bed (VMAT-SIB). Six patient plans and one original plan on the breast phantom were verified with planar portal imaging, point dose, and film measurements in the MultiCube phantom and planar gamma-analysis using ArcCHECK diode array. Six patient plans were recalculated on the breast phantom (hybrid plans) and delivered with point dose and film measurements with 3% (local)/2 mm gamma-analysis. One complete end-to-end test on the breast phantom was performed. All plan quality verifications had point dose differences below 2.4% from the calculated dose and gamma-agreement scores (gamma AS) > 87.3% for film measurements in the MultiCube, portal dosimetry, and ArcCHECK. Point dose differences in the 3D-printed phantoms were below 2.6% (median -1.4%, range -2.6%; 0.3%). Median gamma AS was 96.4% (range 80.1%-99.7%) for all film inserts. The proposed 3D-printed attachable breast dosimetry phantoms have been shown to be a valuable tool for end-to-end testing of a new radiotherapy protocol. The workflow described in this report can aid users to create their own phantom-specific breast 3D-printed phantoms.
引用
收藏
页码:315 / 320
页数:6
相关论文
共 23 条
[1]   3D surface imaging for monitoring intrafraction motion in frameless stereotactic body radiotherapy of lung cancer [J].
Alderliesten, Tanja ;
Sonke, Jan-Jakob ;
Betgen, Anja ;
van Vliet-Vroegindeweij, Corine ;
Remeijer, Peter .
RADIOTHERAPY AND ONCOLOGY, 2012, 105 (02) :155-160
[2]   Preparation and fabrication of a full-scale, sagittal-sliced, 3D-printed, patient-specific radiotherapy phantom [J].
Craft, Daniel F. ;
Howell, Rebecca M. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05) :285-292
[3]   Calibrating page sized Gafchromic EBT3 films [J].
Crijns, W. ;
Maes, F. ;
van der Heide, U. A. ;
Van den Heuvel, F. .
MEDICAL PHYSICS, 2013, 40 (01)
[4]   A quantitative evaluation of IMRT dose distributions: refinement and clinical assessment of the gamma evaluation [J].
Depuydt, T ;
Van Esch, A ;
Huyskens, DP .
RADIOTHERAPY AND ONCOLOGY, 2002, 62 (03) :309-319
[5]   Patient specific 3D printed phantom for IMRT quality assurance [J].
Ehler, Eric D. ;
Barney, Brett M. ;
Higgins, Patrick D. ;
Dusenbery, Kathryn E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (19) :5763-5773
[6]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[7]   3D Slicer as an image computing platform for the Quantitative Imaging Network [J].
Fedorov, Andriy ;
Beichel, Reinhard ;
Kalpathy-Cramer, Jayashree ;
Finet, Julien ;
Fillion-Robin, Jean-Christophe ;
Pujol, Sonia ;
Bauer, Christian ;
Jennings, Dominique ;
Fennessy, Fiona ;
Sonka, Milan ;
Buatti, John ;
Aylward, Stephen ;
Miller, James V. ;
Pieper, Steve ;
Kikinis, Ron .
MAGNETIC RESONANCE IMAGING, 2012, 30 (09) :1323-1341
[8]   Design, development, and implementation of the Radiological Physics Center's pelvis and thorax anthropomorphic quality assurance phantoms [J].
Followill, David S. ;
Evans, DeeAnn Radford ;
Cherry, Christopher ;
Molineu, Andrea ;
Fisher, Gary ;
Hanson, William F. ;
Lbbott, Geoffrey S. .
MEDICAL PHYSICS, 2007, 34 (06) :2070-2076
[9]   Planning strategies in volumetric modulated arc therapy for breast [J].
Giorgia, Nicolini ;
Antonella, Fogliata ;
Alessandro, Clivio ;
Eugenio, Vanetti ;
Luca, Cozzi .
MEDICAL PHYSICS, 2011, 38 (07) :4025-4031
[10]   Evaluation of a 3D surface imaging system for deep inspiration breath-hold patient positioning and intra-fraction monitoring [J].
Hamming, Vincent C. ;
Visser, Christa ;
Batin, Estelle ;
McDermott, Leah N. ;
Busz, Dianne M. ;
Both, Stefan ;
Langendijk, Johannes A. ;
Sijtsema, Nanna M. .
RADIATION ONCOLOGY, 2019, 14 (1)