Peripheral tissue oximetry: comparing three commercial near-infrared spectroscopy oximeters on the forearm

被引:45
作者
Hyttel-Sorensen, Simon [1 ]
Hessel, Trine Witzner [1 ]
Greisen, Gorm [1 ]
机构
[1] Univ Copenhagen Hosp, Rigshosp, Dept Neonatol, DK-2100 Copenhagen, Denmark
关键词
Near infrared spectroscopy (NIRS); Oxygenation; Muscle; Medical device; OXYGEN-SATURATION; CEREBRAL OXYGENATION;
D O I
10.1007/s10877-013-9507-9
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Estimation of regional tissue oxygenation (rStO(2)) by near infrared spectroscopy enables non-invasive end-organ oxygen balance monitoring and could be a valuable tool in intensive care. However, the diverse absolute values and dynamics of different devices, and overall poor repeatability of measurements are a problem. The aim of the present study is to test the hypothesis that INVOS 5100C, FORE-SIGHT and NONIN EQUANOX 7600 have similar properties concerning absolute values, repeatability, and sensitivity to changes in rStO(2). To test repeatability the sensors were repositioned 20 times during hemodynamic steady state on the adult forearm. Afterwards six vascular occlusions by inflation of an upper arm cuff were done to achieve low oxygenation in the forearm. Absolute values were compared by repeated-measures ANOVA, repeatability was estimated by the within-subject standard deviation, S-w, and response to changing oxygenation by the down slope of rStO(2) during vascular occlusion in the respective arm. 10 healthy adults, 21-29 years old, with double skinfolds on the forearm less than 10 mm participated. The median rStO(2) was 70.7 % (interquartile range (IQR) 7.7 %), 68.4 % (IQR 8.4 %), and 64.6 % (IQR 4.8) with INVOS, NONIN, and FORE-SIGHT, respectively, the median rate of decline was 13.2 %/min (IQR 9.6), 22.8 %/min (IQR 18.0), and 10.8 %/min (IQR 6.0), and the same-site repeatability was 2.9 % (95 % CI 2.4-3.3), 4.6 % (CI 3.9-5.3), and 2.0 % (CI 1.7-2.3). INVOS gave significantly higher steady state values than FORE-SIGHT, and NONIN had the steepest decline in rStO(2), but the poorest repeatability. Two measures of signal-to-noise were similar among devices. This suggests that good repeatability comes at the expense of low sensitivity to changes in oxygenation. Values of rStO(2) on the forearm from INVOS, NONIN and FORE-SIGTH cannot be used interchangeably.
引用
收藏
页码:149 / 155
页数:7
相关论文
共 20 条
  • [1] Assessment of tissue oxygen saturation during a vascular occlusion test using near-infrared spectroscopy: the role of probe spacing and measurement site studied in healthy volunteers
    Bezemer, Rick
    Lima, Alexandre
    Myers, Dean
    Klijn, Eva
    Heger, Michal
    Goedhart, Peter T.
    Bakker, Jan
    Ince, Can
    [J]. CRITICAL CARE, 2009, 13
  • [2] Bickler PE, 2011, AM SOC AN ANN M, P1
  • [3] Impact of Extracranial Contamination on Regional Cerebral Oxygen Saturation A Comparison of Three Cerebral Oximetty Technologies
    Davie, Sophie N.
    Grocott, Hilary P.
    [J]. ANESTHESIOLOGY, 2012, 116 (04) : 834 - 840
  • [4] Dullenkopf Alexander, 2005, Pediatr Crit Care Med, V6, P344, DOI 10.1097/01.PCC.0000161282.69283.75
  • [5] Biceps brachii myoelectric and oxygenation changes during static and sinusoidal isometric exercises
    Felici, Francesco
    Quaresima, Valentina
    Fattorini, Luigi
    Sbriccoli, Paola
    Filligoi, Gian Carlo
    Ferrari, Marco
    [J]. JOURNAL OF ELECTROMYOGRAPHY AND KINESIOLOGY, 2009, 19 (02) : E1 - E11
  • [6] Neonatal cerebral oximetry monitoring during ECMO cannulation
    Fenik, J. C.
    Rais-Bahrami, K.
    [J]. JOURNAL OF PERINATOLOGY, 2009, 29 (05) : 376 - 381
  • [7] The use of near-infrared spectroscopy in understanding skeletal muscle physiology: recent developments
    Ferrari, Marco
    Muthalib, Makii
    Quaresima, Valentina
    [J]. PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2011, 369 (1955): : 4577 - 4590
  • [8] Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm
    Hyttel-Sorensen, Simon
    Sorensen, Line C.
    Riera, Joan
    Greisen, Gorm
    [J]. BIOMEDICAL OPTICS EXPRESS, 2011, 2 (11): : 3047 - 3057
  • [9] Low tissue oxygen saturation at the end of early goal-directed therapy is associated with worse outcome in critically ill patients
    Lima, Alexandre
    van Bommel, Jasper
    Jansen, Tim C.
    Ince, Can
    Bakker, Jan
    [J]. CRITICAL CARE, 2009, 13
  • [10] MacLeod D., 2009, ANESTH ANALG S, V108, P1