Direct sensitive quantitative LC/MS analysis of C-peptide from human urine by two dimensional reverse phase/reverse phase high-performance liquid chromatography

被引:17
作者
Rogatsky, E
Tomuta, V
Cruikshank, G
Vele, L
Jayatillake, H
Stein, D
机构
[1] Albert Einstein Coll Med, Gen Clin Res Ctr, Analyt Core Lab, Bronx, NY 10461 USA
[2] Albert Einstein Coll Med, Div Endocrine, Dept Med, Bronx, NY USA
关键词
2-D chromatography; column switching; C-peptide; isotope dilution assay; matrix effects;
D O I
10.1002/jssc.200500369
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We describe for the first time the direct quantitative analysis of human C-peptide from urine by isotope dilution assay. Implementation of 2-D reverse phase-reverse phase chromatography (2-D RP-RP) with SIM detection resulted in high assay sensitivity (LOQ= 5 pg on column), accuracy, and method ruggedness. Our experiments demonstrate the strong resolving capability of our proposed 2-D RP-RP platform which significantly reduces strong matrix effects and their resulting quantitation error over a wide range of urine concentrations. In contrast, single column methods (both SIM and multiple reaction monitoring) were found acceptable only for strongly diluted urine samples.
引用
收藏
页码:529 / 537
页数:9
相关论文
共 14 条
  • [1] Ion suppression in mass spectrometry
    Annesley, TM
    [J]. CLINICAL CHEMISTRY, 2003, 49 (07) : 1041 - 1044
  • [2] CHOI B, 2003, J CHROMATOGR A, V907, P337
  • [3] Assays for insulin, proinsulin(s) and C-peptide
    Clark, PM
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1999, 36 : 541 - 564
  • [4] A mass spectrometric method for quantitation of intact insulin in blood samples
    Darby, SM
    Miller, ML
    Allen, RO
    LeBeau, M
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (01) : 8 - 14
  • [5] Overcoming practical limitations for the application of ultrafiltration in sample preparation for liquid chromatography/mass spectrometry of small proteins
    Fierens, C
    Thienpont, LM
    Stöckl, D
    De Leenheer, AP
    [J]. ANALYTICAL BIOCHEMISTRY, 2000, 285 (01) : 168 - 169
  • [6] Fierens C, 2000, RAPID COMMUN MASS SP, V14, P936, DOI 10.1002/(SICI)1097-0231(20000530)14:10<936::AID-RCM960>3.0.CO
  • [7] 2-A
  • [8] Application of a C-peptide electrospray ionization-isotope dilution-liquid chromatography-tandem mass spectrometry measurement procedure for the evaluation of five C-peptide immunoassays for urine
    Fierens, C
    Stöckl, D
    Baetens, D
    De Leenheer, AP
    Thienpont, LM
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2003, 792 (02): : 249 - 259
  • [9] Quantitative analysis of urinary C-peptide by liquid chromatography-tandem mass spectrometry with a stable isotopically labelled internal standard
    Fierens, C
    Thienpont, LMR
    Stöckl, D
    Willekens, E
    De Leenheer, AP
    [J]. JOURNAL OF CHROMATOGRAPHY A, 2000, 896 (1-2) : 275 - 278
  • [10] Development of an isotope dilution assay for precise determination of insulin, C-peptide, and proinsulin levels in non-diabetic and type II diabetic individuals with comparison to immunoassay
    Kippen, AD
    Cerini, F
    Vadas, L
    Stocklin, R
    Vu, L
    Offord, RE
    Rose, K