Acute performance evaluation of a new ventricular automatic capture algorithm

被引:10
作者
Sperzel, J
Nowak, B
Himmrich, E
Zhang, G
König, A
Willems, R
Reister, C
Sathaye, A
Fröhlig, G
机构
[1] Kerckhoff Klin, Dept Cardiol & Electrophysiol, D-61231 Bad Nauheim, Germany
[2] Cardioangiol Ctr Bethanien, D-60389 Frankfurt, Germany
[3] Univ Mainz Klinikum, D-55101 Mainz, Germany
[4] Guidant Corp, St Paul, MN 55112 USA
[5] Guidant Europe, Dept Clin Res, B-1831 Diegem, Belgium
[6] Univ Klinikum Saarlandes, D-66421 Homburg Saar, Germany
来源
EUROPACE | 2006年 / 8卷 / 01期
关键词
pacemaker; automatic capture verification; pacing artifact; polarization; pacing threshold;
D O I
10.1093/europace/euj008
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims This study evaluated the acute clinical performance of a new ventricular automatic capture algorithm developed to work with all lead types and pacing vectors. Methods and results During regular pacemaker implant or replacement, AutoThreshold and manual threshold tests were performed in ventricular unipolar (UP) and bipolar (BP, if applicable) pacing using a customized external prototype INSIGNIA (TM) pacemaker. The success rate and accuracy of two different modes (commanded and ambulatory) of the automatic capture algorithm were used to evaluate the performance. Loss-of-capture events (two consecutive non-captured beats without backup pacing) were used to assess safety. Data of 53 patients (33 DDD/20 VVI) from four medical centres were analysed. Tested leads included 43 BP and 10 UP from nine manufacturers, and seven had etectrodes with low polarization. The rate of successful commanded and ambulatory AutoThreshold tests was 96 and 94%, respectively, with an average absolute threshold difference compared with manual threshold of < 0.1 V at 0.4 ms (commanded 0.07 +/- 0.07 V and ambulatory 0.08 +/- 0.07 V). There was no significant difference in performance between UP/BP pacing, polarization, and lead type. No loss-of-capture event was observed. Conclusion When successful, the ventricular automatic capture algorithm accurately determined pacing thresholds in either a UP or BP pacing configuration among all leads tested.
引用
收藏
页码:65 / 69
页数:5
相关论文
共 21 条
  • [1] FEASIBILITY OF USING INTRACARDIAC IMPEDANCE MEASUREMENTS FOR CAPTURE DETECTION
    ALT, E
    KRIEGLER, C
    FOTUHI, P
    WILLHAUS, R
    COMBS, W
    HEINZ, M
    HAYES, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1992, 15 (11): : 1873 - 1879
  • [2] BOLZ A, 1993, MED PROG TECHNOL, V19, P129
  • [3] Impact of automatic adjustment of stimulation outputs on pacemaker longevity in a new dual-chamber pacing system
    Brockes, C
    Rahn-Schönbeck, M
    Duru, F
    Candinas, R
    Turina, M
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2003, 8 (01) : 45 - 48
  • [4] Automatic adjustment of pacemaker stimulation output correlated with continuously monitored capture thresholds:: A multicenter study
    Clarke, M
    Liu, B
    Schüller, H
    Binner, L
    Kennergren, C
    Guerola, M
    Weinmann, P
    Ohm, OJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (08): : 1567 - 1575
  • [5] An algorithm for automatic measurement of stimulation thresholds: Clinical performance and preliminary results
    Danilovic, D
    Ohm, OJ
    Stroebel, J
    Breivik, K
    Hoff, PI
    Markowitz, T
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (05): : 1058 - 1068
  • [6] Pacing threshold trends and variability in modern tined leads assessed using high resolution automatic measurements: Conversion of pulse width into voltage thresholds
    Danilovic, D
    Ohm, OJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (04): : 567 - 587
  • [7] Dohrmann M L, 1985, Cardiol Clin, V3, P527
  • [8] Automatic capture verification by charge-neutral sensing
    Kadhiresan, VA
    Olive, A
    Gornick, C
    Spinelli, J
    Villalta, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1999, 22 (01): : 73 - 78
  • [9] Clinical experience with an automatic threshold tracking algorithm study
    Kennergren, C
    Larsson, B
    Uhrenius, Å
    Gadler, F
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2003, 26 (12): : 2219 - 2224
  • [10] A cardiac evoked response algorithm providing threshold tracking: A North American multicenter study
    Lau, C
    Cameron, DA
    Nishimura, SC
    Ahern, T
    Freedman, RA
    Ellenbogen, K
    Greenberg, S
    Baker, J
    Meacham, D
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2000, 23 (06): : 953 - 959