Digital mammography in direct magnification technique using a large-area amorphous silicon X-ray detector.

被引:15
作者
Hermann, KP
Hundertmark, C
Funke, M
von Brenndorff, A
Grabbe, E
机构
[1] Univ Gottingen, Inst Med Phys & Biophys, D-37073 Gottingen, Germany
[2] Univ Gottingen, Zentrum Radiol, Abt Rontgendiagnost 1, D-3400 Gottingen, Germany
[3] Feinfocus Med Technik GmbH, Garbsen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1999年 / 170卷 / 05期
关键词
digital mammography; flat panel X-ray detector; direct magnification technique; contrast-detail resolution;
D O I
10.1055/s-2007-1011080
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the image quality of a new direct digital mammography system using a large-area amorphous silicon Xray detector in a phantom study. Materials and Methods: The contrast-detail resolution as a function of the tube voltages, the magnification factors and the mean glandular doses were investigated using dedicated test objects. Results: The contrast-detail resolution was significantly improved in comparison with conventional screen-film mammography. Usually, the doses necessary to obtain these high-quality survey mammograms were smaller. Conclusions: By combining the direct magnification technique and a digital flat panel detector the limited spatial resolution of such image receptors can be overcome. With this direct digital mammography technique, a digital image was directly captured without an intermediate step of optical or mechanical scan.
引用
收藏
页码:503 / 506
页数:4
相关论文
共 8 条
[1]  
[Anonymous], 1996, EUROPEAN PROTOCOL DO
[2]   A magnification mammography technique using storage phosphor plates: Experimental investigations concerning spatial resolution and detection of microcalcifications. [J].
Funke, M ;
Hermann, KP ;
Breiter, N ;
HundertMark, C ;
Sachs, J ;
Gruhl, T ;
Sperner, W ;
Grabbe, E .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (02) :174-179
[3]   DIMA-magnification mammography: use of phantoms for the investigation of image quality and radiation exposure. [J].
Hermann, KP ;
Funke, M ;
Bruggemeyer, H ;
Sperner, W ;
Gruhl, T ;
Koch, R .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1997, 167 (03) :304-310
[4]   Determination of average glandular dose with modern mammography units for two large groups of patients [J].
Klein, R ;
Aichinger, H ;
Dierker, J ;
Jansen, JTM ;
JoiteBarfuss, S ;
Sabel, M ;
SchulzWendtland, R ;
Zoetelief, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (04) :651-671
[5]  
NEITZEL U, 1997, KONTRASTE, V11, P16
[6]  
Spahn M., 1997, Electromedica, V65, P37
[7]   New amorphous-silicon image sensor for x-ray diagnostic medical imaging applications [J].
Weisfield, RL ;
Hartney, MA ;
Street, RA ;
Apte, RB .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :444-452
[8]  
WEISFIELD RL, 1997, MED IMAGING SPIE, V3032, P9