Bench Testing of a Knee Joint Arthrometer

被引:2
作者
Barcellona, Massimo G. [1 ]
Christopher, Tony [2 ]
Morrissey, Matthew C. [3 ]
机构
[1] Kings Coll London, Sch Med, Acad Dept Physiotherapy, London, England
[2] Kings Coll London, Ctr Human & Aerosp Physiol Sci, London, England
[3] Univ Ljubljana, Fac Hlth Sci, Ljubljana 1000, Slovenia
关键词
RADIOSTEREOMETRIC ANALYSIS; DEFICIENT KNEE; LAXITY; INSTABILITY; RECONSTRUCTION; AGREEMENT; DRAWER;
D O I
10.3928/01477447-20130724-14
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The KT1000 and KT2000 knee joint arthrometers (MEDmetric Corp, San Diego, California) have been shown to overestimate the measurement of knee joint sagittal laxity. The purpose of this study was to investigate the accuracy of the KT arthrometers as measures of anterior and posterior linear displacement. The anterior and posterior linear displacements of 3 KT arthrometers (2 KT1000 arthrometers and 1 KT2000 arthrometer) were compared with the simultaneous displacement measured by a precision linear Vernier Dial Test Indicator (Davenport Ltd, London, United Kingdom). The displacement calculated using the analog output of the KT2000 was also compared with the values on the KT2000 displacement dial. Compared with the Vernier Dial Test Indicator, the KT arthrometers overestimated anterior linear displacement by between 22% and 24%. True anterior displacement for all 3 arthrometers, as recorded by the Vernier Dial Test Indicator, was found by multiplying the KT value by 0.79. When compared with the Vernier Dial Test Indicator, the KT arthrometers underestimated posterior linear displacement by between 18% and 19%. True posterior displacement, as recorded by the Vernier Dial Test Indicator, was found by multiplying the KT1000 value by 1.17 and the KT2000 value by 1.16. The internal apparatus of the KT2000 and KT1000 knee joint arthrometers overestimates anterior displacement and underestimates posterior displacement with a predictable relative systematic error. Future validation studies should use these correction equations to assess the accuracy of the KT arthrometers. Sagittal plane knee laxity measured with the KT devices requires systematic correction for optimal accuracy.
引用
收藏
页码:E1000 / E1006
页数:7
相关论文
共 17 条
[1]   The effect of coordinate system choice and segment reference on RSA-based knee translation measures [J].
Beardsley, Christina L. ;
Paller, David J. ;
Peura, Glenn D. ;
Brattbakk, Bjarne ;
Beynnon, Bruce D. .
JOURNAL OF BIOMECHANICS, 2007, 40 (06) :1417-1422
[2]  
Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   Kinematics and laxity of the knee joint after anterior cruciate ligament reconstruction -: Pre- and postoperative radiostereometric studies [J].
Brandsson, S ;
Karlsson, J ;
Swärd, L ;
Kartus, J ;
Eriksson, BI ;
Kärrholm, J .
AMERICAN JOURNAL OF SPORTS MEDICINE, 2002, 30 (03) :361-367
[5]   LIGAMENTOUS RESTRAINTS TO ANTERIOR-POSTERIOR DRAWER IN THE HUMAN KNEE - BIOMECHANICAL STUDY [J].
BUTLER, DL ;
NOYES, FR ;
GROOD, ES .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1980, 62 (02) :259-270
[6]   INSTRUMENTED MEASUREMENT OF ANTERIOR KNEE LAXITY IN PATIENTS WITH ACUTE ANTERIOR CRUCIATE LIGAMENT DISRUPTION [J].
DANIEL, DM ;
STONE, ML ;
SACHS, R ;
MALCOM, L .
AMERICAN JOURNAL OF SPORTS MEDICINE, 1985, 13 (06) :401-407
[7]   ACCURACY AND REPRODUCIBILITY OF INSTRUMENTED KNEE-DRAWER TESTS [J].
EDIXHOVEN, P ;
HUISKES, R ;
DEGRAAF, R ;
VANRENS, TJG ;
SLOOFF, TJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1987, 5 (03) :378-387
[8]   Measurement of anterior-posterior knee laxity: a comparison of three techniques [J].
Fleming, BC ;
Brattbakk, B ;
Peura, GD ;
Badger, GJ ;
Beynnon, BD .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (03) :421-426
[9]  
FRIDEN T, 1992, ACTA ORTHOP SCAND, V63, P593
[10]   EVALUATION OF AN ELECTROGONIOMETRIC INSTRUMENT FOR MEASUREMENT OF LAXITY OF THE KNEE [J].
GRANBERRY, WM ;
NOBLE, PC ;
WOODS, GW .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (09) :1316-1322