Blood Pressure Estimation From Time-Domain Features of Oscillometric Waveforms Using Long Short-Term Memory Recurrent Neural Networks

被引:33
作者
Argha, Ahmadreza [1 ]
Celler, Branko G. [1 ]
机构
[1] Univ New South Wales, Sch Elect Engn & Telecommun, Sydney, NSW 2052, Australia
基金
澳大利亚研究理事会;
关键词
Long short-term memory recurrent neural network (LSTM-RNN); systolic blood pressure (SBP) and diastolic blood pressure (DBP) estimation; time-domain features of oscillometric waveform (OW); MAXIMUM AMPLITUDE ALGORITHM; ACCURACY;
D O I
10.1109/TIM.2019.2941037
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This article presents a novel method for estimating systolic blood pressure (SBP) and diastolic blood pressure (DBP) from time-domain features extracted from oscillometric waveforms (OWs) using a long short-term memory (LSTM) recurrent neural network (RNN) method. First, we extract seven time-domain features from each cycle of OW, including the cuff pressure, the cardiac period, the trough-to-peak amplitude of OW, the time between the trough and the peak of the OW, the slopes of the oscillometric waveform envelope (OWE), and the maximum upslope of individual OWs. Second, we locate each feature vector in an noninvasive blood pressure ( NIBP) record in one of three different phases (classes), namely, presystolic (PS), between systolic and diastolic ( BSD), and after diastolic (AD), and form a target sequence. Then, we propose an LSTM-RNN approach to effectively learn the complex nonlinear relationship between the feature vector sequences and the target sequence. The SBP and DBP points are then obtained by mapping the beats at which the network output sequence switches from PS phase to BSD phase and from BSD phase to AD phase, respectively, to the deflation curve. Adopting a tenfold cross-validation scheme and using a database of 350 NIBP recordings gave an average mean error of -1.2 +/- 5.9 mmHg for SBP and 1.8 +/- 8.8 mmHg for DBP relative to reference values derived from a visual method of determining SBP and DBP. The proposed RNN-based approach uses all time-domain features available from each NIBP recording and can outperform traditional methods in blood pressure (BP) estimation.
引用
收藏
页码:3614 / 3622
页数:9
相关论文
共 44 条
[21]   CHARACTERIZATION OF THE OSCILLOMETRIC METHOD FOR MEASURING INDIRECT BLOOD-PRESSURE [J].
GEDDES, LA ;
VOELZ, M ;
COMBS, C ;
REINER, D ;
BABBS, CF .
ANNALS OF BIOMEDICAL ENGINEERING, 1982, 10 (06) :271-280
[22]   LSTM: A Search Space Odyssey [J].
Greff, Klaus ;
Srivastava, Rupesh K. ;
Koutnik, Jan ;
Steunebrink, Bas R. ;
Schmidhuber, Juergen .
IEEE TRANSACTIONS ON NEURAL NETWORKS AND LEARNING SYSTEMS, 2017, 28 (10) :2222-2232
[23]  
Koohi R, 2016, IEEE INSTRU MEAS MAG, V19, P26, DOI 10.1109/MIM.2016.7579067
[24]  
Lee S.-K., 2011, Public Diplomacy and Soft Power in East Asia, P1, DOI [DOI 10.1007/S11042-011-0850-X, DOI 10.1057/9780230118447-1, 10.1109/ICISA.2011.5772401, DOI 10.1109/IMTC.2011.5944338]
[25]   Deep learning ensemble with asymptotic techniques for oscillometric blood pressure estimation [J].
Lee, Soojeong ;
Chang, Joon-Hyuk .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2017, 151 :1-13
[26]   Deep Belief Networks Ensemble for Blood Pressure Estimation [J].
Lee, Soojeong ;
Chang, Joon-Hyuk .
IEEE ACCESS, 2017, 5 :9962-9972
[27]   Oscillometric Blood Pressure Estimation Based on Deep Learning [J].
Lee, Soojeong ;
Chang, Joon-Hyuk .
IEEE TRANSACTIONS ON INDUSTRIAL INFORMATICS, 2017, 13 (02) :461-472
[28]   Improved Gaussian Mixture Regression Based on Pseudo Feature Generation Using Bootstrap in Blood Pressure Estimation [J].
Lee, Soojeong ;
Park, Chee-Hyun ;
Chang, Joon-Hyuk .
IEEE TRANSACTIONS ON INDUSTRIAL INFORMATICS, 2016, 12 (06) :2269-2280
[29]   Oscillometric Blood Pressure Estimation Based on Maximum Amplitude Algorithm Employing Gaussian Mixture Regression [J].
Lee, Soojeong ;
Chang, Joon-Hyuk ;
Nam, Sang Won ;
Lim, Chungsoo ;
Rajan, Sreeraman ;
Dajani, Hilmi R. ;
Groza, Voicu Z. .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2013, 62 (12) :3387-3389
[30]   Confidence Interval Estimation for Oscillometric Blood Pressure Measurements Using Bootstrap Approaches [J].
Lee, Soojeong ;
Bolic, Miodrag ;
Groza, Voicu Z. ;
Dajani, Hilmi R. ;
Rajan, Sreeraman .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (10) :3405-3415