Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy

被引:79
作者
Fogliata, A. [1 ]
Garcia, R. [2 ]
Knoos, T. [3 ,4 ]
Nicolini, G. [1 ]
Clivio, A. [1 ]
Vanetti, E. [1 ]
Khamphan, C. [2 ]
Cozzi, L. [1 ]
机构
[1] Oncol Inst So Switzerland, Med Phys Unit, CH-6500 Bellinzona, Switzerland
[2] Inst St Catherine, Med Phys Unit, F-84000 Avignon, France
[3] Skane Univ Hosp, S-22185 Lund, Sweden
[4] Lund Univ, Dept Med Radiat Phys, S-22185 Lund, Sweden
关键词
quality assurance; flattening filter free beams; dosimetric parameter definition; FREE CLINICAL ACCELERATOR; MEDICAL LINEAR-ACCELERATOR; STOPPING-POWER RATIOS; MONTE-CARLO; FREE CLINAC; RADIOTHERAPY; NEUTRON;
D O I
10.1118/1.4754799
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Flattening filter free (FFF) beams generated by medical linear accelerators have recently started to be used in radiotherapy clinical practice. Such beams present fundamental differences with respect to the standard filter flattened (FF) beams, making the generally used dosimetric parameters and definitions not always viable. The present study will propose possible definitions and suggestions for some dosimetric parameters for use in quality assurance of FFF beams generated by medical linacs in radiotherapy. Methods: The main characteristics of the photon beams have been analyzed using specific data generated by a Varian TrueBeam linac having both FFF and FF beams of 6 and 10 MV energy, respectively. Results: Definitions for dose profile parameters are suggested starting from the renormalization of the with respect to the corresponding FF beam. From this point the flatness concept has been translated into one of "unflatness" and other definitions have been proposed, maintaining a strict parallelism between FFF and FF parameter concepts. Conclusions: Ideas for quality controls used in establishing a quality assurance program when introducing FFF, beams into the clinical environment are given here, keeping them similar to those used for standard FF beams. By following the suggestions in this report, the authors foresee that the introduction of FFF beams into a clinical radiotherapy environment will be as safe and well controlled as standard beam modalities using the existing guidelines. (C) 2012 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4754799]
引用
收藏
页码:6455 / 6464
页数:10
相关论文
共 31 条
[1]   AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams [J].
Almond, PR ;
Biggs, PJ ;
Coursey, BM ;
Hanson, WF ;
Huq, MS ;
Nath, R ;
Rogers, DWO .
MEDICAL PHYSICS, 1999, 26 (09) :1847-1870
[2]   The characterization of unflattened photon beams from a 6 MV linear accelerator [J].
Cashmore, Jason .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (07) :1933-1946
[3]   Prediction of stopping-power ratios in flattening-filter free beams [J].
Ceberg, Crister ;
Johnsson, Stefan ;
Lind, Marten ;
Knoos, Tommy .
MEDICAL PHYSICS, 2010, 37 (03) :1164-1168
[4]   A Monte Carlo study of a flattening filter-free linear accelerator verified with measurements [J].
Dalaryd, Marten ;
Kragl, Gabriele ;
Ceberg, Crister ;
Georg, Dietmar ;
McClean, Brendan ;
af Wetterstedt, Sacha ;
Wieslander, Elinore ;
Knoos, Tommy .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) :7333-7344
[5]   Current status and future perspective of flattening filter free photon beams [J].
Georg, Dietmar ;
Knoos, Tommy ;
McClean, Brendan .
MEDICAL PHYSICS, 2011, 38 (03) :1280-1293
[6]   Photon beam quality variations of a flattening filter free linear accelerator [J].
Georg, Dietmar ;
Kragl, Gabriele ;
Af Wetterstedt, Sacha ;
McCavana, Patrick ;
McClean, Brendan ;
Knoos, Tommy .
MEDICAL PHYSICS, 2010, 37 (01) :49-53
[7]   COMMISSIONING OF PHOTON BEAMS OF A FLATTENING FILTER-FREE LINEAR ACCELERATOR AND THE ACCURACY OF BEAM MODELING USING AN ANISOTROPIC ANALYTICAL ALGORITHM [J].
Hrbacek, Jan ;
Lang, Stephanie ;
Kloeck, Stephan .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 80 (04) :1228-1237
[8]  
IEC, 1989, INT EL COMM PUBL
[9]  
IEC, 1989, INT EL COMM PUBL
[10]  
International Atomic Energy Agency, 2000, IAEA TRS-398