Monaco and film dosimetry of 3D CRT, IMRT and VMAT cases in a realistic pelvic prosthetic phantom

被引:5
作者
Ade, Nicholas [1 ]
du Plessis, F. C. P. [1 ]
机构
[1] Univ Free State, Med Phys Dept, POB 339, ZA-9300 Bloemfontein, South Africa
基金
英国医学研究理事会;
关键词
Nylon phantom; Ti implant; Multi-field photon-beams; EBT3; film; Monaco TPS; GAFCHROMIC EBT FILM; HIP-PROSTHESIS; DOSE PERTURBATION; PHOTON; BEAM; RADIOTHERAPY; IMPLANTS;
D O I
10.1016/j.radphyschem.2017.11.013
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The dosimetry of patients with metallic hip implants during irradiation of pelvic lesions is challenging due to dose distortions caused by implants. This work presents a dosimetric comparison of various multi-field photon beam dose distributions in the presence of unilateral hip titanium prosthesis (UHTiP) embedded in a unique pelvic phantom made out of water-equivalent nylon slices. The impact of the UHTiP on the accuracy of dose calculations from a Monaco TPS (treatment planning system) using the X-ray voxel Monte Carlo (XVMC) algorithm was benchmarked against measured dose data using Gafchromic EBT3 film. Multi-field beam arrangements including a 4-field box, 5-field 3DCRT (three-dimensional conformal radiation therapy), 6-field IMRT (intensity modulated radiation therapy) and a single-arc VMAT (volumetric modulated arc therapy) plan were set up for 6 MV and 15 MV beams. These plans were generated for the pelvic phantom that contains the prosthesis with film inserted. Compared to Monaco TPS dose calculations, film measurements showed enhanced dose in the prosthesis which was not predicted by Monaco due to its limitation in relative density assignment. The enhanced prosthesis dose increased with increase in beam energy and decreased with the complexity of the treatment plans, with VMAT giving the least escalated dose. The dose increased between 5% and 19% for 6 MV and between 6% and 21% for 15 MV. A gamma index analysis showed that 70-92% of dose points (excluding the prosthesis) were within 3% discrepancy. Increasing the number of treatment fields increases target dose coverage and improves the agreement between film and Monaco. When the relative electron density (RED) in the prosthesis was varied between 3.72 and 15 the dose discrepancy between film and Monaco increased from 30% to 57% for 6 MV and from 30% to 50% for 15 MV. The study indicates that beam weights for fields that pass through the prosthesis should be minimised and its RED must be correct for accurate dose calculation on Monaco.
引用
收藏
页码:50 / 57
页数:8
相关论文
共 29 条
[1]   Dose comparison between Gafchromic film, XiO, and Monaco treatment planning systems in a novel pelvic phantom that contains a titanium hip prosthesis [J].
Ade, Nicholas ;
du Plessis, F. C. P. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (05) :162-173
[2]   Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies [J].
Arjomandy, Bijan ;
Tailor, Ramesh ;
Anand, Aman ;
Sahoo, Narayan ;
Gillin, Michael ;
Prado, Karl ;
Vicic, Milos .
MEDICAL PHYSICS, 2010, 37 (05) :1942-1947
[3]   EFFECT OF A FEMORAL-HEAD PROSTHESIS ON MEGAVOLTAGE BEAM RADIOTHERAPY [J].
BIGGS, PJ ;
RUSSELL, MD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 14 (03) :581-586
[4]   Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification [J].
Borca, Valeria Casanova ;
Pasquino, Massimo ;
Russo, Giuliana ;
Grosso, Pierangelo ;
Cante, Domenico ;
Sciacero, Piera ;
Girelli, Giuseppe ;
La Porta, Maria Rosa ;
Tofani, Santi .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02) :158-171
[5]   INTENSITY-MODULATED RADIATION THERAPY WITH NONCOPLANAR BEAMS FOR TREATMENT OF PROSTATE CANCER IN PATIENTS WITH BILATERAL HIP PROSTHESIS-A CASE STUDY [J].
Brooks, Chris ;
Cheung, Rex Min ;
Kudchadker, Rajat J. .
MEDICAL DOSIMETRY, 2010, 35 (02) :87-91
[6]  
Buffard E, 2006, NUCL INSTRUM METH B, V251, P9, DOI 10.1016/j.nimb.2006.05.031
[7]  
Carolan M, 2000, Australas Radiol, V44, P290, DOI 10.1046/j.1440-1673.2000.00816.x
[8]   Dose perturbation in the radiotherapy of breast cancer patients implanted with the Magna-Site: a Monte Carlo study [J].
Chatzigiannis, Christos ;
Lymperopoulou, Georgia ;
Sandilos, Panagiotis ;
Dardoufas, Constantinos ;
Yakoumakis, Emmanouil ;
Georgiou, Evaggelos ;
Karaiskos, Pantelis .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (02) :58-70
[9]   Dose enhancement close to platinum implants for the 4, 6, and 10 MV stereotactic radiosurgery [J].
Cheung, JYC ;
Ng, BKP ;
Yu, KN .
MEDICAL PHYSICS, 2004, 31 (10) :2787-2791
[10]   BACKSCATTER DOSE PERTURBATION AT HIGH ATOMIC-NUMBER INTERFACES IN MEGAVOLTAGE PHOTON BEAMS [J].
DAS, IJ ;
KAHN, FM .
MEDICAL PHYSICS, 1989, 16 (03) :367-375