Construction of adaptive spectral analyzers on the basis of acousto-optic spectrometers

被引:1
作者
Fadeyev, A. V. [1 ]
Pozhar, V. E. [1 ]
机构
[1] Russian Acad Sci, Sci Technol Ctr Unique Instrumentat, Moscow, Russia
关键词
SEPARATION; ERROR;
D O I
10.1364/JOT.80.000444
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This paper discusses the problem of selecting the working spectral points in the method of selective spectral recording, used to detect contaminants in air by means of differential optical absorption spectroscopy on acousto-optic spectrometers with random spectral addressing. Reducing the number of recorded points of the spectrum, which allows the measurement time to be reduced by an order of magnitude, can simultaneously decrease the selectivity of the analysis, and this presents a problem if the composition of the impurity substances is unknown. A partial solution of the problem is proposed that consists of a preliminary study of the variations of the spectra of the gas mixture, and this makes it possible to statistically distinguish the main variable components in the composition of the gaseous mixture and to identify them with definite gases. Versions are proposed for the construction of adaptive systems based on acousto-optic spectrometers and using the technique developed here. (C) 2013 Optical Society of America.
引用
收藏
页码:444 / 449
页数:6
相关论文
共 22 条
[21]  
Pustovoit V. I., 2001, P SOC PHOTO-OPT INS, V4574, P174
[22]   Numerical analysis and estimation of the statistical error of differential optical absorption spectroscopy measurements with least-squares methods [J].
Stutz, J ;
Platt, U .
APPLIED OPTICS, 1996, 35 (30) :6041-6053