Comparison of exposure standards in the mammography x-ray region

被引:13
作者
Coletti, JG
Pearson, DW
DeWerd, LA
OBrien, CM
Lamperti, PJ
机构
[1] UNIV WISCONSIN,DEPT MED PHYS,MADISON,WI 53706
[2] NATL INST STAND & TECHNOL,GAITHERSBURG,MD 20899
关键词
free-air ionization chamber; calibration; mammography; dosimetry;
D O I
10.1118/1.598148
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Direct comparisons of the National Institute of Standards and Technology's (NIST) Ritz 20 kV to 100 kV standard free-air ionization chamber and a portable variable-length free-air ionization chamber designed by the University of Wisconsin - Madison Accredited Dosimetry Calibration Laboratory (UW-ADCL) were made on NISTs low-energy tungsten x-ray range. As a result of this direct comparison, NIST has established a UW-ADCL designed chamber, the Attix chamber, as the national standard chamber for the mammography energy x-ray range. The Ritz standard chamber and the Attix standard chamber have been extensively compared using the new molybdenum and rhodium beam qualities. The results indicate that exposure measurements in the mammography energy x-ray region with the two free-air chambers can be made with a discrepancy of less than 0.35%. (C) 1997 American Association of Physicists in Medicine.
引用
收藏
页码:1263 / 1267
页数:5
相关论文
共 8 条
[1]  
ATTIX FH, 1961, 5646 NAV RES LAB
[2]   MAMMOGRAPHY EXPOSURE STANDARD - DESIGN AND CHARACTERIZATION OF FREE-AIR IONIZATION-CHAMBER [J].
COLETTI, JG ;
PEARSON, DW ;
DEWERD, LA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (03) :2574-2577
[3]  
LAITANO RF, 1984, 83 RTPROT CRE CAS LA
[4]  
Lamperti Paul J., 1988, NBS SPECIAL PUBLICAT, P250
[5]  
MARSH ARS, 1980, 53 NPL RS
[6]   DESIGN OF FREE-AIR IONIZATION CHAMBERS FOR THE SOFT X-RAY REGION (20-100 KV) [J].
RITZ, VH .
RADIOLOGY, 1959, 73 (06) :911-922
[7]  
RITZ VH, 1960, J RES NBS C ENG INST, V64, P49
[8]  
WYCKOFF HO, 1956, NBS HDB, P64