Simulated One-Pass List-Mode: A Highly Flexible Method of Image Reconstruction for PET

被引:0
作者
Gillam, J. E. [1 ]
Solevi, P. [1 ]
Oliver, J. E. [1 ]
Rafecas, M. [1 ]
机构
[1] Univ Valencia, IFIC CSIC, Edificio Inst Invest,Apartado Correos 22085, E-46071 Valencia, Spain
来源
2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2011年
关键词
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Simulated One Pass List-mode is a means of image-reconstruction that distributes on-the-fly calculations of the full system matrix across multiple iterations during Maximum Likelihood-Expectation Maximization image reconstruction. On the fly list-mode processing of coincidence data provides a very flexible approach to image reconstruction that can be applied to many novel systems for Positron Emission Tomography. However, in addition to standard on-the-fly calculation and list-mode methods, Simulated One Pass List-mode provides for extra possibilities during image reconstruction. Two possibilities are investigated here: system modeling accuracy can be altered at each iteration and enhanced modeling can be conduced only for specific Regions of Interest. In this investigation simulated data from a simple two dimensional PET system is used to demonstrate these two advantages of Simulated One Pass List-mode image reconstruction for Positron Emission Tomography.
引用
收藏
页码:4206 / 4210
页数:5
相关论文
共 4 条
[1]  
Gillam J., 2011, FULL 3D C P
[2]   Characterization of a PET detector head based on continuous LYSO crystals and monolithic, 64-pixel silicon photomultiplier matrices [J].
Llosa, G. ;
Barrio, J. ;
Lacasta, C. ;
Bisogni, M. G. ;
Del Guerra, A. ;
Marcatili, S. ;
Barrillon, P. ;
Bondil-Blin, S. ;
de la Taille, C. ;
Piemonte, C. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (23) :7299-7315
[3]   One-pass list-mode EM algorithm for high-resolution 3-D PET image reconstruction into large arrays [J].
Reader, AJ ;
Ally, S ;
Bakatselos, F ;
Manavaki, R ;
Walledge, RJ ;
Jeavons, AP ;
Julyan, PJ ;
Zhao, S ;
Hastings, DL ;
Zweit, J .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) :693-699
[4]   FAST CALCULATION OF THE EXACT RADIOLOGICAL PATH FOR A 3-DIMENSIONAL CT ARRAY [J].
SIDDON, RL .
MEDICAL PHYSICS, 1985, 12 (02) :252-255