Accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients

被引:33
作者
Hirschl, MM [1 ]
Binder, M [1 ]
Herkner, H [1 ]
Bur, A [1 ]
Brunner, M [1 ]
Seidler, D [1 ]
Stuhlinger, HG [1 ]
Laggner, AN [1 ]
机构
[1] UNIV VIENNA,DEPT DERMATOL,VIENNA,AUSTRIA
关键词
arterial pressure; noninvasive monitoring; blood pressure measurement; critical illness; emergency measurement;
D O I
10.1097/00003246-199610000-00014
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To evaluate the accuracy and reliability of noninvasive continuous finger blood pressure measurement in critically ill patients. Design: Prospective data collection. Setting: Emergency department in a 2,000-bed hospital. Patients: Thirty-nine patients admitted to the emergency department requiring invasive arterial blood pressure monitoring were enrolled to the study protocol. Interventions: Continuous noninvasive blood pressure measurement was performed on the middle phalanx of the second and third finger, using a test instrument which provides continuous arterial waveform display with the use of a finger cuff. Invasive mean arterial blood pressure measurement was done by cannula tion of the radial artery and direct transduction of the systemic arterial pressure waveform. Measurements and Main Results: Three thousand one hundred eighteen pairs of simultaneous finger cuff and intra-arterial blood pressure measurements were collected in 1-min intervals from 39 patients over a total of 51.8 hrs. The overall discrepancy between both measurements was 0.10 mm Hg. The standard deviation of the differences was +/-5.02 mm Hg. The mean bias in patients treated with catecholamines was 0.01 mm Hg and was not different from the bias observed in patients without catecholamines (mean bias: 0.23 mm Hg; p > .22). Whereas 95% of all comparisons between finger cuff and intra arterial measurement had a discrepancy less than or equal to+/-10 mm Hg, 4.7% had a discrepancy between +/-10.1 to 15 mm Hg and 0.3% exhibited a discrepancy >+/-15 mm Hg. In 29 (74%) patients, the duration of errors >10 mm Hg was less than or equal to 1 min. In seven (18%) patients, the duration of errors >10 mm Hg was between 2 to 3 mins and in three (8%) patients, the errors lasted for >3 mins. Conclusions: Our data provide a guide to the accuracy and reliability of noninvasive finger blood pressure measurements in critically ill patients. Although most test instrument measurements were reliable, in 8% of all patients large discrepancies (>10 mm Hg) between both measurements with a duration of >3 mins were noted. Concerning the considerable risk for arterial cannulation, our preliminary data demonstrate that the test instrument (PORTAPRES(R) TNO Biomedical Instrumentation Research Unit; The Netherlands) is an advance in noninvasive monitoring of critically ill patients and may be useful in most emergency clinical settings.
引用
收藏
页码:1684 / 1689
页数:6
相关论文
共 25 条
[1]  
[Anonymous], 1990, BLOOD PRESSURE MEASU
[2]  
Armitage P, 1987, Statistical methods in medical research, V2nd
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   OSLER NODES, PSEUDOANEURYSM FORMATION, AND SEPSIS COMPLICATING PERCUTANEOUS RADIAL ARTERY CANNULATION [J].
COHEN, A ;
REYES, R ;
KIRK, M ;
FULKS, RM .
CRITICAL CARE MEDICINE, 1984, 12 (12) :1078-1079
[5]   ARTERIAL-OCCLUSION AFTER CANNULATION [J].
EVANS, PJD ;
KERR, JH .
BRITISH MEDICAL JOURNAL, 1975, 3 (5977) :197-199
[6]   DIRECT BLOOD-PRESSURE MEASUREMENT - DYNAMIC-RESPONSE REQUIREMENTS [J].
GARDNER, RM .
ANESTHESIOLOGY, 1981, 54 (03) :227-236
[7]   THE ACCURACY OF FINAPRES NONINVASIVE MEAN ARTERIAL-PRESSURE MEASUREMENTS IN ANESTHETIZED PATIENTS [J].
GIBBS, NM ;
LARACH, DR ;
DERR, JA .
ANESTHESIOLOGY, 1991, 74 (04) :647-652
[8]  
IMHOLZ BPM, 1992, J HYPERTENS, V10, P979
[9]   NONINVASIVE CONTINUOUS FINGER BLOOD-PRESSURE MEASUREMENT DURING ORTHOSTATIC STRESS COMPARED TO INTRAARTERIAL PRESSURE [J].
IMHOLZ, BPM ;
SETTELS, JJ ;
VANDERMEIRACKER, AH ;
WESSELING, KH ;
WIELING, W .
CARDIOVASCULAR RESEARCH, 1990, 24 (03) :214-221
[10]   FEASIBILITY OF AMBULATORY, CONTINUOUS 24-HOUR FINGER ARTERIAL-PRESSURE RECORDING [J].
IMHOLZ, BPM ;
LANGEWOUTERS, GJ ;
VANMONTFRANS, GA ;
PARATI, G ;
VANGOUDOEVER, J ;
WESSELING, KH ;
WIELING, W ;
MANCIA, G .
HYPERTENSION, 1993, 21 (01) :65-73